TRUESPAN ORTHOCORD 228151Recalls & reported problems

Medos International Sàrl · TRUESPAN Meniscal Repair System PEEK 12 DEGREES ORTHOCORD Violet braided composite suture Size 2-0, (2) PEEK Backstops, and (1) Applier with 12 DEGREES needle Depth Stop: 10mm-20mm Plus or Minus 1mm; Needle: 10mm Plus or Minus .13mm · Class 2 · Implant

All models of TRUESPAN ORTHOCORD (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is the TrueSpan OrthoCord meniscal repair system, a suture kit used in arthroscopic knee surgery. The 1,664 reports often describe parts breaking during surgery, two plates deploying at once, and devices not fitting with other parts. Some describe the device not releasing. The records include no deaths, 58 injuries and 1,606 malfunctions, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,686
Deaths reported
0
Injuries reported
58
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,686 adverse event reports that list this exact device (DI 10886705026012), received between Dec 28, 2017 and Sep 30, 2026 — classified as 58 injuries, 1,628 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (703 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 26 reportsJan 2024: 17 reportsFeb 2024: 11 reportsMar 2024: 17 reportsApr 2024: 22 reportsMay 2024: 17 reportsJun 2024: 27 reportsJul 2024: 8 reportsAug 2024: 19 reportsSep 2024: 23 reportsOct 2024: 17 reportsNov 2024: 23 reportsDec 2024: 20 reportsJan 2025: 20 reportsFeb 2025: 26 reportsMar 2025: 7 reportsApr 2025: 10 reportsMay 2025: 14 reportsJun 2025: 16 reportsJul 2025: 12 reportsAug 2025: 9 reportsSep 2025: 5 reportsOct 2025: 1 reportsNov 2025: 23 reportsDec 2025: 18 reportsJan 2026: 14 reportsFeb 2026: 21 reportsMar 2026: 32 reportsApr 2026: 16 reportsMay 2026: 20 reportsJun 2026: 12 reportsJul 2026: 2 reportsAug 2026: 2 reportsSep 2026: 32 reportsOct 2026: 0 reports20242025202632
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility703
  • Break326
  • Premature Activation121
  • Migration96
  • Positioning Failure95
  • Mechanical Jam82
  • Difficult to Advance78
  • Detachment of Device or Device Component75

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,195
  • No Consequences Or Impact To Patient313
  • Not Applicable142
  • Unspecified Tissue Injury23
  • Foreign Body In Patient21
  • No Code Available19
  • Tissue Damage14
  • No Known Impact Or Consequence To Patient13

Recent adverse event reports

The 25 most recent of 1,686 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Premature Activation

    It was reported that when the first shot of the TRUESPAN ORTHOCORD was fired, the device activated and released both shots. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943821 ↗
  • MalfunctionSep 29, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938839 ↗
  • MalfunctionSep 29, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938838 ↗
  • MalfunctionSep 29, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938835 ↗
  • MalfunctionSep 29, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938833 ↗
  • MalfunctionSep 29, 2026Device-Device Incompatibility; Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935263 ↗
  • MalfunctionSep 29, 2026Device-Device Incompatibility; Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935254 ↗
  • MalfunctionSep 29, 2026Break; Unintended Movement; Physical Resistance/Sticking

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934167 ↗
  • MalfunctionSep 29, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933673 ↗
  • MalfunctionSep 29, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933494 ↗
  • MalfunctionSep 29, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933488 ↗
  • MalfunctionSep 25, 2026Device-Device Incompatibility

    It was reported that during a meniscal repair surgical procedure, when using the TRUESPAN 12 DEGREE PEEK device, it was observed that the device could not fire. It was reported that the surgeon attempted to use 2X additional TRUESPAN 12 DEGREE PEEK devices to continue the surgery, but the same problem happened again with both devices. It was reported that another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information could be provided. This is report 1 of 3 for PC-(b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899806 ↗
  • MalfunctionSep 25, 2026Device-Device Incompatibility

    It was reported that during a meniscal repair surgical procedure, when using the TRUESPAN 12 DEGREE PEEK device, it was observed that the device could not fire. It was reported that the surgeon attempted to use 2X additional TRUESPAN 12 DEGREE PEEK devices to continue the surgery, but the same problem happened again with both devices. It was reported that another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information could be provided.This is report 2 of 3 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899803 ↗
  • MalfunctionSep 25, 2026Device-Device Incompatibility

    It was reported that during a meniscal repair surgical procedure, when using the TRUESPAN 12 DEGREE PEEK device, it was observed that the device could not fire. It was reported that the surgeon attempted to use 2X additional TRUESPAN 12 DEGREE PEEK devices to continue the surgery, but the same problem happened again with both devices. It was reported that another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information could be provided.This is report 3 of 3 for (b)(4)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899802 ↗
  • MalfunctionSep 21, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846147 ↗
  • MalfunctionSep 16, 2026Difficult to Advance

    It was reported that during a Meniscal Repair surgery, the second plate of the TRUESPAN ORTHOCORD did not fix the meniscus after deployment. Changed to another TRUESPAN ORTHOCORD to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 2 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50791086 ↗
  • MalfunctionSep 16, 2026Difficult to Advance

    It was reported that during a Meniscal Repair surgery, the second plate of the TRUESPAN ORTHOCORD did not fix the meniscus after deployment. Changed to another TRUESPAN ORTHOCORD to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 1 of 2 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50791085 ↗
  • MalfunctionSep 16, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50790109 ↗
  • MalfunctionSep 16, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50790087 ↗
  • MalfunctionSep 16, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787768 ↗
  • MalfunctionSep 16, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787700 ↗
  • MalfunctionSep 16, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787244 ↗
  • MalfunctionSep 16, 2026Device-Device Incompatibility

    It was reported that during a Meniscal Repair surgery, a TRUESPAN ORTHOCORD did not fire. Changed to another two TRUESPAN ORTHOCORDs to continue the surgery but the same problem happened again. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.This is report 2 of 3 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787242 ↗
  • MalfunctionSep 16, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787240 ↗
  • MalfunctionSep 16, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787036 ↗