MITEK CRUCIATE+ 219003Recalls & reported problems
Medos International Sàrl · MITEK CRUCIATE+ Beath Pin Bullet 2.4mm 15-85mm Plus or Minus 8 Percent · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886705025855), received between Jan 30, 2019 and Feb 5, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Difficult to Advance1
- Material Deformation1
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2024Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation
IT WAS REPORTED BY SALES REP THAT DURING SOFT TISSUE ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT WHILE TRYING TO DRILL TIBIAL TUNNEL, THE CRUCIATE+ BEATH PIN BULLET DEVICE WOULD NOT ENGAGE THE CRUCIATE GUIDE CARRIAGE PROPERLY. IT WAS FURTHER REPORTED THAT THE DEVICE MIGHT HAVE BEEN BENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18639669 ↗MalfunctionJul 2, 2019Break; Break
IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT THE FAILURE OF THE CRUCIATE+ BEATH PIN BULLET WAS IDENTIFIED DURING ITS CORRECT USE IN A CADAVEROUS COURSE, THE BEATH PIN 14'', WHICH WAS A PART OF THE DISPOSABLE KIT GOT TRAPPED AND LOCKED INSIDE THE BULLET. WHEN TRYING TO RELEASE IT, THE BULLET WAS BROKEN AND WAS REMAINING IN 2 PARTS IMAGES ATTACHED. BUT STILL THE DRILL DID NOT COME OUT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8756231 ↗MalfunctionJan 30, 2019Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED DURING AN UNKNOWN PROCEDURE THE ACR DRILLPN W/EYE AND THE CRUCIATE + BEATH PIN BULLET BECAME STUCK TOGETHER. THE DEVICES BECAME STUCK WHILE DRILLING THE TIBIAL TUNNEL. THE PROCEDURE WAS COMPLETED USING A BACK-UP TRAY. THERE WAS 39-SECOND DELAY IN SURGERY AND NO PATIENT CONSEQUENCE. THIS REPORT IS 2 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8290072 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026