Skip to content

NA 242316Recalls & reported problems

Medos International Sàrl · HD Coupler, C-Mount 20mm · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886705026272), received between Feb 26, 2020 and May 28, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Device-Device Incompatibility3
  • Display or Visual Feedback Problem3
  • Difficult to Advance2
  • Material Deformation2
  • Crack1
  • Device Contamination with Chemical or Other Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient1
  • Not Applicable1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED THAT (2X) HD COUPLER, C-MOUNT, 20 MM DEVICES OPENING/CLOSER LEVER/KNOB WAS BROKEN AND THERE WAS PHYSICAL DAMAGE WHICH DID NOT ALLOW FOR ATTACHMENT TO THE SCOPE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE DEVICE PRONG WAS BROKEN FROM ITS LOCKING SYSTEM. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22097691 ↗
  • MalfunctionMay 28, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED THAT (2X) HD COUPLER, C-MOUNT, 20MM DEVICES OPENING/CLOSER LEVER/KNOB WAS BROKEN AND THERE WAS PHYSICAL DAMAGE WHICH DID NOT ALLOW FOR ATTACHMENT TO THE SCOPE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE APERTURE LEVER WAS BROKEN. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22097688 ↗
  • MalfunctionSep 25, 2024Crack; Display or Visual Feedback Problem; Crack; Display or Visual Feedback Problem

    IT WAS REPORTED BY AN AFFILIATE THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON 8/23/2024 THE 2X SHE_2RSTCK_5.9,30,167CW_ MITEK DEVICES WERE BLURRY / CRACKED WHILE USED WITH THE 3X HD COUPLER, C-MOUNT, 20MM DEVICES. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. THE DATE OF EVENT OCCURRED WAS UNKNOWN BUT IT WAS KNOWN TO HAVE OCCURRED EITHER ON 8/23/2024 OR 8/24/2024. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 4 OF 4 OF THE SAME EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20305282 ↗
  • MalfunctionNov 21, 2023Break; Break

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT TWO HD COUPLER, C-MOUNT, 20MM DEVICE HAD BROKEN RELEASE. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18184882 ↗
  • MalfunctionNov 21, 2023Break; Break

    THIS IS REPORT 2 OF 2 FOR: (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT TWO HD COUPLER, C-MOUNT, 20MM DEVICE HAD BROKEN RELEASE. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18184869 ↗
  • MalfunctionJan 24, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED BY THE CUSTOMER THAT THE RELEASE FOR THE COUPLER, C-MOUNT, 20MM BROKE; AND THAT A SCOPE WAS STILL ATTACHED BUT NOT BROKEN. IT WAS REPORTED THAT THE DEVICE WAS NOT FOR HUMAN USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16235682 ↗
  • MalfunctionAug 2, 2022Display or Visual Feedback Problem; Difficult to Advance; Material Deformation; Display or Visual Feedback Problem; Difficult to Advance; Material Deformation

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN UNITED KINGDOM THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE COUPLER, C-MOUNT, 20MM DEVICE DID NOT APPEAR TO BE SITTING CORRECTLY AGAINST THE SCOPE. ACCORDING TO THE REPORT, IT WAS GIVING A BLURRED IMAGE; AND WAS ALSO SLIGHTLY BENT ON THE HANDLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15147847 ↗
  • MalfunctionAug 2, 2022Display or Visual Feedback Problem; Difficult to Advance; Material Deformation; Display or Visual Feedback Problem; Difficult to Advance; Material Deformation

    THIS IS REPORT 1 OF 2 FOR (B)(4).. IT WAS REPORTED BY THE SALES REP IN UNITED KINGDOM THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE COUPLER, C-MOUNT, 20MM DEVICE DID NOT APPEAR TO BE SITTING CORRECTLY AGAINST THE SCOPE. ACCORDING TO THE REPORT, IT WAS GIVING A BLURRED IMAGE; AND WAS ALSO SLIGHTLY BENT ON THE HANDLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15147843 ↗
  • MalfunctionMar 14, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED BY THE SALES REP VIA EMAIL THAT DURING A SHOULDER SCOPE PROCEDURE THE COUPLER, C-MOUNT, 20 MM IS CLOUDY. THE PROCEDURE WAS COMPLETED WITH ANOTHER COUPLER WITH NO PATIENT HARM BUT THERE WAS A MINUTE SURGICAL DELAY TO THE CASE TO CHANGE OUT THE COUPLER. THE DEVICE WILL BE RETURNING FOR EVALUATION.

    Patient problems: Not Applicable

    MAUDE report 9832572 ↗
  • MalfunctionFeb 26, 2020Break; Break

    IT WAS REPORTED BY SALES REP VIA EMAIL, THAT WHILE CLEANING A COUPLER, C-MOUNT, 20 MM AND A CAMERA HEAD AC, C-MOUNT, THE COUPLER BROKE OFF IN THE CAMERA HEAD. NO SURGICAL DELAY OR PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION PROVIDED BY THE AFFILIATE REPORTED THE EVENT DID NOT OCCUR DURING STERILE PROCESSING, FILED INSPECTION, OR INTERNAL SERVICE ACTIVITIES. THE AFFILIATE WAS ALSO ABLE TO CONFIRM SURGICAL INTERVENTION IS NOT PLANNED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9758330 ↗

Other ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES) devices

All ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES) devices →

Sources and updates