Olympus WM-DP2Recalls & reported problems

Keymed (Medical & Industrial Equipment) Limited · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15019778004655), received between Jul 19, 2021 and Sep 9, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Break1

Patient problems

  • Electric Shock2
  • No Clinical Signs, Symptoms or Conditions1
  • Numbness1
  • Swelling/ Edema1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2024Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE, PRESENTED THE STEM CHANNEL FOR THE CASTOR DAMAGED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20170420 ↗
  • InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTS WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A WM-DP2 WORKSTATION AND A MAJ-860 ACTIVE CORD (ALONG WITH A VALLEYLAB GENERATOR AND BOVIE), THE ENDOSCOPY TECHNICIAN EXPERIENCED AN ELECTRICAL SHOCK WHILE CONNECTING AN ELECTROSURGERY ACCESSORY (BOSTON SCIENTIFIC SINGLE USE POLYPECTOMY SNARE). IT IS REPORTED THE TECHNICIAN'S GLOVES WERE WET AND THE GENERATOR WAS POWERED ON AT THE TIME HE ATTEMPTED TO CONNECT THE SNARE. THE TECHNICIAN DECLINED AN EMERGENCY ROOM VISIT FOR EVALUATION AT THE TIME AND CONTINUED TO WORK THE REMAINDER OF THE DAY. THE TECHNICIAN'S CURRENT CONDITION IS DESCRIBED AS UNABLE TO WORK DUE TO "HIS ARM AND HAND ARE STILL SWOLLEN AND NUMB." THE DEVICES WERE USED FOR THE REMAINED OF THE DAY WITH NO FURTHER ISSUES REPORTED. BIOMED CHECKED UNITS AND FOUND NO ISSUES. THE MAJ-860 WAS INSPECTED WITH NO VISIBLE DAMAGE NOTED. BIOMED HAS REMOVED THE NON-OLYMPUS GENERATOR FROM USE AND REPLACED WITH ANOTHER UNIT ON A SEPARATE CART/ ROLL STAND. THE OLYMPUS CORD (MAJ-860) AND BOVIE WERE CHECKED OUT BY BIOMED, AND NO MALFUNCTIONS WERE IDENTIFIED. THE OLYMPUS MAJ-860 HAS BEEN REPLACED WITH A BOSTON SCIENTIFIC ACTIVE CORD. NO FURTHER ISSUES REPORTED. CASE WIT

    Patient problems: Electric Shock

    MAUDE report 12189891 ↗
  • InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTS WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A WM-DP2 WORKSTATION AND A MAJ-860 ACTIVE CORD (ALONG WITH A VALLEYLAB GENERATOR AND BOVIE), THE ENDOSCOPY TECHNICIAN EXPERIENCED AN ELECTRICAL SHOCK WHILE CONNECTING AN ELECTROSURGERY ACCESSORY (BOSTON SCIENTIFIC SINGLE USE POLYPECTOMY SNARE). IT IS REPORTED THE TECHNICIAN'S GLOVES WERE WET AND THE GENERATOR WAS POWERED ON AT THE TIME HE ATTEMPTED TO CONNECT THE SNARE. THE TECHNICIAN DECLINED AN EMERGENCY ROOM VISIT FOR EVALUATION AT THE TIME AND CONTINUED TO WORK THE REMAINDER OF THE DAY. THE TECHNICIAN'S CURRENT CONDITION IS DESCRIBED AS UNABLE TO WORK DUE TO "HIS ARM AND HAND ARE STILL SWOLLEN AND NUMB." THE DEVICES WERE USED FOR THE REMAINED OF THE DAY WITH NO FURTHER ISSUES REPORTED. BIOMED CHECKED UNITS AND FOUND NO ISSUES. THE MAJ-860 WAS INSPECTED WITH NO VISIBLE DAMAGE NOTED. BIOMED HAS REMOVED THE NON-OLYMPUS GENERATOR FROM USE AND REPLACED WITH ANOTHER UNIT ON A SEPARATE CART/ ROLL STAND. THE OLYMPUS CORD (MAJ-860) AND BOVIE WERE CHECKED OUT BY BIOMED, AND NO MALFUNCTIONS WERE IDENTIFIED. THE OLYMPUS MAJ-860 HAS BEEN REPLACED WITH A BOSTON SCIENTIFIC ACTIVE CORD. NO FURTHER ISSUES REPORTED. CASE WIT

    Patient problems: Electric Shock; Numbness; Swelling/ Edema

    MAUDE report 12189888 ↗