Olympus WM-DP2Recalls & reported problems
Keymed (Medical & Industrial Equipment) Limited · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15019778004655), received between Jul 19, 2021 and Sep 9, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break1
Patient problems
- Electric Shock2
- No Clinical Signs, Symptoms or Conditions1
- Numbness1
- Swelling/ Edema1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2024Break; Break
IT WAS REPORTED THE SUBJECT DEVICE, PRESENTED THE STEM CHANNEL FOR THE CASTOR DAMAGED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20170420 ↗InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTS WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A WM-DP2 WORKSTATION AND A MAJ-860 ACTIVE CORD (ALONG WITH A VALLEYLAB GENERATOR AND BOVIE), THE ENDOSCOPY TECHNICIAN EXPERIENCED AN ELECTRICAL SHOCK WHILE CONNECTING AN ELECTROSURGERY ACCESSORY (BOSTON SCIENTIFIC SINGLE USE POLYPECTOMY SNARE). IT IS REPORTED THE TECHNICIAN'S GLOVES WERE WET AND THE GENERATOR WAS POWERED ON AT THE TIME HE ATTEMPTED TO CONNECT THE SNARE. THE TECHNICIAN DECLINED AN EMERGENCY ROOM VISIT FOR EVALUATION AT THE TIME AND CONTINUED TO WORK THE REMAINDER OF THE DAY. THE TECHNICIAN'S CURRENT CONDITION IS DESCRIBED AS UNABLE TO WORK DUE TO "HIS ARM AND HAND ARE STILL SWOLLEN AND NUMB." THE DEVICES WERE USED FOR THE REMAINED OF THE DAY WITH NO FURTHER ISSUES REPORTED. BIOMED CHECKED UNITS AND FOUND NO ISSUES. THE MAJ-860 WAS INSPECTED WITH NO VISIBLE DAMAGE NOTED. BIOMED HAS REMOVED THE NON-OLYMPUS GENERATOR FROM USE AND REPLACED WITH ANOTHER UNIT ON A SEPARATE CART/ ROLL STAND. THE OLYMPUS CORD (MAJ-860) AND BOVIE WERE CHECKED OUT BY BIOMED, AND NO MALFUNCTIONS WERE IDENTIFIED. THE OLYMPUS MAJ-860 HAS BEEN REPLACED WITH A BOSTON SCIENTIFIC ACTIVE CORD. NO FURTHER ISSUES REPORTED. CASE WIT
Patient problems: Electric Shock
MAUDE report 12189891 ↗InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTS WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A WM-DP2 WORKSTATION AND A MAJ-860 ACTIVE CORD (ALONG WITH A VALLEYLAB GENERATOR AND BOVIE), THE ENDOSCOPY TECHNICIAN EXPERIENCED AN ELECTRICAL SHOCK WHILE CONNECTING AN ELECTROSURGERY ACCESSORY (BOSTON SCIENTIFIC SINGLE USE POLYPECTOMY SNARE). IT IS REPORTED THE TECHNICIAN'S GLOVES WERE WET AND THE GENERATOR WAS POWERED ON AT THE TIME HE ATTEMPTED TO CONNECT THE SNARE. THE TECHNICIAN DECLINED AN EMERGENCY ROOM VISIT FOR EVALUATION AT THE TIME AND CONTINUED TO WORK THE REMAINDER OF THE DAY. THE TECHNICIAN'S CURRENT CONDITION IS DESCRIBED AS UNABLE TO WORK DUE TO "HIS ARM AND HAND ARE STILL SWOLLEN AND NUMB." THE DEVICES WERE USED FOR THE REMAINED OF THE DAY WITH NO FURTHER ISSUES REPORTED. BIOMED CHECKED UNITS AND FOUND NO ISSUES. THE MAJ-860 WAS INSPECTED WITH NO VISIBLE DAMAGE NOTED. BIOMED HAS REMOVED THE NON-OLYMPUS GENERATOR FROM USE AND REPLACED WITH ANOTHER UNIT ON A SEPARATE CART/ ROLL STAND. THE OLYMPUS CORD (MAJ-860) AND BOVIE WERE CHECKED OUT BY BIOMED, AND NO MALFUNCTIONS WERE IDENTIFIED. THE OLYMPUS MAJ-860 HAS BEEN REPLACED WITH A BOSTON SCIENTIFIC ACTIVE CORD. NO FURTHER ISSUES REPORTED. CASE WIT
Patient problems: Electric Shock; Numbness; Swelling/ Edema
MAUDE report 12189888 ↗
Other Accessories, Photographic, For Endoscope (Exclude Light Sources) devices
- NA 242437Medos International Sàrl · ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)191 reports
- Ambu aView™ 2 Advance 405011000Ambu A/S · ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)2 recalls4 reports
- NA 242436Medos International Sàrl · ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)67 reports
- NA 242435Medos International Sàrl · ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)37 reports
- NA 242316Medos International Sàrl · ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)10 reports
- Olympus WM-NP2Keymed (Medical & Industrial Equipment) Limited · Accessories, Photographic, For Endoscope (Exclude Light Sources)4 reports
- Olympus TC-L Compact Trolley Set 1Keymed (Medical & Industrial Equipment) Limited · Accessories, Photographic, For Endoscope (Exclude Light Sources)2 reports
- NA 242312Medos International Sàrl · ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.