Skip to content

Ambu aView™ 2 Advance 405011000Recalls & reported problems

Ambu A/S · Ambu aView 2 Advance · Class 1

All models of Ambu aView 2 Advance (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05707480145089), received between Apr 4, 2024 and Jun 12, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 5, 2023Open, Classified

    Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

    Z-0714-2024 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 21, 2023Open, Classified

    Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames

    Z-2628-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Battery Problem2
  • Failure to Power Up2
  • Insufficient Device Problem Information2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2024Battery Problem; Insufficient Device Problem Information; Battery Problem; Insufficient Device Problem Information

    ACCORDING TO THE INFORMATION FROM THE CUSTOMER, THE HOUSING OF THE MONITOR HAS BULGED. THEY REPORTED THAT THE BATTERY HAS INFLATED AND BURST THE MONITOR. PICTURES WERE DELIVERED FOR THE INVESTIGATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522066 ↗
  • MalfunctionJun 12, 2024Battery Problem; Insufficient Device Problem Information; Battery Problem; Insufficient Device Problem Information

    ACCORDING TO THE INFORMATION FROM THE CUSTOMER, THE HOUSING OF THE MONITOR HAS BULGED. THEY REPORTED THAT THE BATTERY HAS INFLATED AND BURST THE MONITOR. PICTURES WERE DELIVERED FOR THE INVESTIGATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19518071 ↗
  • MalfunctionApr 4, 2024Failure to Power Up; Failure to Power Up

    THE USER REPORTED THAT THE MONITOR WAS NOT RESPONDING AND HAVING ISSUE BOOTING UP WHILST TRYING TO INTUBATE A PATIENT. ANOTHER MONITOR WAS USED AND THE PATIENT WAS INTUBATED SUCCESSFULLY. THE USER HAS REPORTED THAT THE MONITOR HAS WORKED AGAIN AFTER A HARD RESET HOWEVER THEY HAVE NOTICED BOOTING AND SHUTTING DOWN ISSUES WITH THIS MONITOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19048038 ↗
  • MalfunctionApr 4, 2024Failure to Power Up; Failure to Power Up

    THE USER REPORTED THAT THE MONITOR WAS NOT RESPONDING AND HAVING ISSUE BOOTING UP WHILST TRYING TO INTUBATE A PATIENT. ANOTHER MONITOR WAS USED AND THE PATIENT WAS INTUBATED SUCCESSFULLY. THE USER HAS REPORTED THAT THE MONITOR HAS WORKED AGAIN AFTER A HARD RESET HOWEVER THEY HAVE NOTICED BOOTING AND SHUTTING DOWN ISSUES WITH THIS MONITOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19039919 ↗

Other ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES) devices

All ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES) devices →

Sources and updates