INTRAFIX 254807Recalls & reported problems

Medos International Sàrl · INTRAFIX Advance Tibial Fastener System BIOCRYL RAPIDE Sheath and Screw Small 8 x 30mm BR · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This INTRAFIX Advance tibial fastener is a small biodegradable screw and sheath used to secure soft tissue during knee ligament repair, including arthroscopy. The 50 reports mostly describe the screw or its sheath breaking during surgery, with fragments then removed and another device used to finish the operation. No deaths and no injuries were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
52
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 52 adverse event reports that list this exact device (DI 10886705026739), received between Sep 6, 2023 and Sep 23, 2026 — classified as 52 malfunctions.

The most frequently reported specific device problem is “Break” (51 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 3 reportsApr 2026: 5 reportsMay 2026: 2 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break51
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions52

Recent adverse event reports

The 25 most recent of 52 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873664 ↗
  • MalfunctionSep 23, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873662 ↗
  • MalfunctionSep 18, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829138 ↗
  • MalfunctionSep 2, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50572862 ↗
  • MalfunctionJun 12, 2026Break

    IT WAS REPORTED THAT DURING A ANTERIOR CRUCIATE LIGAMENT REPAIR WHEN INSERTING THE SCREW, THE SCREW WAS BROKEN, AND ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25484813 ↗
  • MalfunctionJun 12, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPY, THE INTRAFX ADVBR SCW8X30 W/SMALL SHEATH WAS BROKEN AND ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 3 MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25482057 ↗
  • MalfunctionJun 11, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE INTRAFX ADVBR SCW8X30 W/SMSHTH ANCHOR DEVICE BROKE. ALL BROKEN PARTS WERE REMOVED, ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477193 ↗
  • MalfunctionJun 8, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, WHEN THE INTRAFIX SCREW WAS INSERTED, THE SCREW BROKE. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25425254 ↗
  • MalfunctionMay 28, 2026Break

    THIS IS REPORT 2 OF 2 OF (B)(4). IT WAS REPORTED THAT DURING AN ARTHROSCOPY SURGERY, THE INTRAFIX DEVICE SCREW WAS BROKEN WHEN INSERTED. ALL BROKEN PARTS WERE REMOVED FROM THE PATIENT. CHANGED TO ANOTHER SAME PRODUCT TO CONTINUE THE SURGERY. IT WAS NOTED THAT THE RIGIDLOOP DEVICE WAS FRAYING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25317558 ↗
  • MalfunctionMay 8, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SHEATH OF THE SCREW BROKE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25127697 ↗
  • MalfunctionApr 27, 2026Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE WAS BROKEN OFF. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24997061 ↗
  • MalfunctionApr 23, 2026Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SHEATH BROKE DURING IMPLANTATION. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24971424 ↗
  • MalfunctionApr 21, 2026Break

    IT WAS REPORTED THAT DURING KNEE ARTHROSCOPY PROCEDURE WITH AN INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE, WHEN INSERT THE SCREW, THE SCREW WAS BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT OR DELAYS. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24944832 ↗
  • MalfunctionApr 11, 2026Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE BROKE. REMOVED ALL THE BROKEN PARTS. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 5 MIN. DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24852488 ↗
  • MalfunctionApr 7, 2026Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION, WHEN INSERTING THE INTRAFIX SCREW, THE SCREW WAS BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24798995 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE BROKE. ALL PIECES WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 5 MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24746422 ↗
  • MalfunctionMar 23, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPY SURGICAL PROCEDURE, WHEN INSERTING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE IT WAS OBSERVED THAT THE SCREW WAS BROKEN. IT WAS REPORTED THAT ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24668583 ↗
  • MalfunctionMar 10, 2026Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY, WHEN THE INTRAFX ADVBR SCW8X30 W/SMSHTH SCREW WAS INSERTED, THE SCREW WAS BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24560256 ↗
  • MalfunctionFeb 20, 2026Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE INTRAFX ADVBR DEVICE SCREW BROKE OFF. ALL PIECES WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A FIVE MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24400590 ↗
  • MalfunctionFeb 3, 2026Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE, IT WAS DISCOVERED THAT THE SHEATH OF THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE BROKE. IT WAS REPORTED THAT THERE WAS A FIVE MINUTE DELAY IN SURGERY. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251295 ↗
  • MalfunctionDec 29, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SCREW BROKE WHEN TRYING TO INSERT IT. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A FIVE MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23909631 ↗
  • MalfunctionDec 3, 2025Break; Break

    IT WAS REPORTED THAT DURING THE LIGAMENT RECONSTRUCTION SURGERY, IT WAS OBSERVED THAT WHEN INSERTING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE, THE SCREW WAS BROKEN. IT WAS REPORTED THAT ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23702958 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE AFTER PLACING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE IT WAS OBSERVED THAT THE SCREW BROKE. THE SCREW WAS PARTIALLY REMOVED AND A MILAGRO ADVANCE SCREW 7X23MM DEVICE WAS ADVANCED AND PLACED AND WAS ALSO BELIEVED TO HAVE PARTLY BROKE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23649263 ↗
  • MalfunctionNov 18, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE, WHEN INSERTING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE, IT WAS OBSERVED THAT THE SCREW WAS BROKEN. IT WAS REPORTED THAT ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586976 ↗
  • MalfunctionOct 30, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT PROCEDURE, THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SCREW SHEATH SHATTERED DURING IMPLANTATION. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23426841 ↗

Other Fastener, fixation, biodegradable, soft tissue devices