This INTRAFIX Advance tibial fastener is a small biodegradable screw and sheath used to secure soft tissue during knee ligament repair, including arthroscopy. The 50 reports mostly describe the screw or its sheath breaking during surgery, with fragments then removed and another device used to finish the operation. No deaths and no injuries were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 52
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 52 adverse event reports that list this exact device (DI 10886705026739), received between Sep 6, 2023 and Sep 23, 2026 — classified as 52 malfunctions.
The most frequently reported specific device problem is “Break” (51 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break51
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions52
Recent adverse event reports
The 25 most recent of 52 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873664 ↗MalfunctionSep 23, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873662 ↗MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829138 ↗MalfunctionSep 2, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50572862 ↗MalfunctionJun 12, 2026Break
IT WAS REPORTED THAT DURING A ANTERIOR CRUCIATE LIGAMENT REPAIR WHEN INSERTING THE SCREW, THE SCREW WAS BROKEN, AND ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25484813 ↗MalfunctionJun 12, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPY, THE INTRAFX ADVBR SCW8X30 W/SMALL SHEATH WAS BROKEN AND ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 3 MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25482057 ↗MalfunctionJun 11, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE INTRAFX ADVBR SCW8X30 W/SMSHTH ANCHOR DEVICE BROKE. ALL BROKEN PARTS WERE REMOVED, ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25477193 ↗MalfunctionJun 8, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, WHEN THE INTRAFIX SCREW WAS INSERTED, THE SCREW BROKE. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25425254 ↗MalfunctionMay 28, 2026Break
THIS IS REPORT 2 OF 2 OF (B)(4). IT WAS REPORTED THAT DURING AN ARTHROSCOPY SURGERY, THE INTRAFIX DEVICE SCREW WAS BROKEN WHEN INSERTED. ALL BROKEN PARTS WERE REMOVED FROM THE PATIENT. CHANGED TO ANOTHER SAME PRODUCT TO CONTINUE THE SURGERY. IT WAS NOTED THAT THE RIGIDLOOP DEVICE WAS FRAYING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25317558 ↗MalfunctionMay 8, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SHEATH OF THE SCREW BROKE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25127697 ↗MalfunctionApr 27, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE WAS BROKEN OFF. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24997061 ↗MalfunctionApr 23, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SHEATH BROKE DURING IMPLANTATION. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24971424 ↗MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT DURING KNEE ARTHROSCOPY PROCEDURE WITH AN INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE, WHEN INSERT THE SCREW, THE SCREW WAS BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT OR DELAYS. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24944832 ↗MalfunctionApr 11, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE BROKE. REMOVED ALL THE BROKEN PARTS. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 5 MIN. DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24852488 ↗MalfunctionApr 7, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION, WHEN INSERTING THE INTRAFIX SCREW, THE SCREW WAS BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24798995 ↗MalfunctionMar 31, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE BROKE. ALL PIECES WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 5 MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746422 ↗MalfunctionMar 23, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPY SURGICAL PROCEDURE, WHEN INSERTING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE IT WAS OBSERVED THAT THE SCREW WAS BROKEN. IT WAS REPORTED THAT ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24668583 ↗MalfunctionMar 10, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY, WHEN THE INTRAFX ADVBR SCW8X30 W/SMSHTH SCREW WAS INSERTED, THE SCREW WAS BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24560256 ↗MalfunctionFeb 20, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE INTRAFX ADVBR DEVICE SCREW BROKE OFF. ALL PIECES WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A FIVE MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24400590 ↗MalfunctionFeb 3, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE, IT WAS DISCOVERED THAT THE SHEATH OF THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE BROKE. IT WAS REPORTED THAT THERE WAS A FIVE MINUTE DELAY IN SURGERY. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24251295 ↗MalfunctionDec 29, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SCREW BROKE WHEN TRYING TO INSERT IT. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A FIVE MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23909631 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT DURING THE LIGAMENT RECONSTRUCTION SURGERY, IT WAS OBSERVED THAT WHEN INSERTING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE, THE SCREW WAS BROKEN. IT WAS REPORTED THAT ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23702958 ↗MalfunctionNov 25, 2025Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE AFTER PLACING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE IT WAS OBSERVED THAT THE SCREW BROKE. THE SCREW WAS PARTIALLY REMOVED AND A MILAGRO ADVANCE SCREW 7X23MM DEVICE WAS ADVANCED AND PLACED AND WAS ALSO BELIEVED TO HAVE PARTLY BROKE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23649263 ↗MalfunctionNov 18, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE, WHEN INSERTING THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE, IT WAS OBSERVED THAT THE SCREW WAS BROKEN. IT WAS REPORTED THAT ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23586976 ↗MalfunctionOct 30, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT PROCEDURE, THE INTRAFX ADVBR SCW8X30 W/SMSHTH DEVICE SCREW SHEATH SHATTERED DURING IMPLANTATION. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426841 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page