HEALIX ADVANCE BR ANCHOR w/DYNACORD 222004Recalls & reported problems

Medos International Sàrl · HEALIX ADVANCE BR ANCHOR w/DYNACORD TCP/PLGA Absorbable Anchor (1) Blue (1) White/Blue/Green Striped, Size 2 (5 Metric) DYNACORD Suture, 36" (91cm) 5.5mm · Class 2 · Implant

All models of HEALIX ADVANCE BR ANCHOR w/DYNACORD (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886705029464), received between Jun 27, 2018 and Jul 23, 2025 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Adverse Event Without Identified Device or Use Problem2
  • Crack1
  • Detachment of Device or Device Component1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient2
  • Nerve Damage1
  • No Consequences Or Impact To Patient1
  • Perforation of Vessels1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    EVENT DESCRIPTION: RE-TEARS OR NEW TEARS [POST-OPERATIVE]. OPERATIVE SIDE: RIGHT. WAS THE FMS VUE II SYSTEM USED FOR THIS PROCEDURE?: YES, WAS THE VAPR SYSTEM USED FOR THIS PROCEDURE?: YES, WAS THE SPEEDTRAP GRAFT PREPARATION SYSTEM USED FOR THIS PROCEDURE?: NO. IF THE START DATE IS EQUAL TO THE DATE OF PROCEDURE, WHEN DID THE EVENT OCCUR?: BLANK. SEVERITY: MODERATE. DID THIS EVENT RESULT IN A RE-OPERATION/REVISION? IF YES, COMPLETE THE RE-OPERATION/REVISION FORM.: YES, HAS IMPLANT CAUSED OR CONTRIBUTED TO DEATH OR SERIOUS INJURY TO THE PATIENT?: NO.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 22575309 ↗
  • MalfunctionOct 11, 2023Crack; Crack

    IT WAS REPORTED BY SALES REP VIA EMAIL THAT DURING ROTATOR CUFF REPAIR BODY OF THE 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) CRACKED AFTER THE FIRST TWIST, NOT WORKING ON HARD BONE. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17912937 ↗
  • MalfunctionJun 12, 2023Migration; Migration

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) DEVICE HAD NO FIXATION AND PULLED OUT OF THE PATIENT'S BONE IMMEDIATELY. ACCORDING TO THE REPORT, THE PHYSICIAN UPSIZED TO A 6.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) DEVICE BUT PULLED AS WELL. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A FIVE MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17110047 ↗
  • InjuryFeb 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THROUGH A CLINICAL TRIAL USING 5.5HLX ADV BR ANC-DYNA STR/BL DEVICE THAT A PATIENT HAD NEUROVASCULAR INJURY RELATED TO A BICEPS TENODESIS PROCEDURE OF THE LEFT SHOULDER THAT WAS PERFORMED ON (B)(6) 2021. ACCORDING TO THE REPORT, THE PATIENT CONSEQUENCE WAS MILD; AND THEREFORE, DID NOT REQUIRE A REVISION/REOPERATION PROCEDURE. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Nerve Damage; Perforation of Vessels

    MAUDE report 13598878 ↗
  • InjurySep 9, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE THREADS ON THE 5.5HLX ADV BR ANC-DYNA STR/BL DEVICE PULLED OFF ONCE IMPACTED. THE SURGEON COULD NOT RECOVERED THE ANCHOR AND HAD TO USE A NEW ANCHOR TO FINISH THE PROCEDURE. THE STATUS OF THE PATIENT POST-SURGERY WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 12446444 ↗
  • MalfunctionSep 9, 2021Break; Break

    IT WAS REPORTED BY THE SALES REP THAT DURING A BICEP TENODESIS PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE ANCHOR ON THE 4.5HLX ADV BR ANC-DYNA STR/BL DEVICE BROKE UPON INSERTION AS IT WAS TAP INTO THE BONE HOLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12446413 ↗
  • InjuryNov 1, 2019Break; Break

    IT WAS REPORTED VIA PHONE THAT THE SUTURE BRIDGE OF THE 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) BROKE. THIS DEVICE IS NOT AVAILABLE TO BE RETURNED FOR ANALYSIS. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE STATED THE REPORTED FAILURE OCCURRED INTRA-OP AND A SURGICAL DELAY DID NOT OCCUR. THE AFFILIATE ALSO REPORTED THAT FRAGMENTS WERE NOT GENERATED BUT A PART OF THE DEVICE REMAINED IN THE PATIENT. IT WAS ALSO REPORTED THE PROCEDURE WAS COMPLETED WITH A READILY AVAILABLE ANCHOR AND SURGICAL INTERVENTION IS NOT PLANNED. THERE WERE NO REPORTED CONSEQUENCES TO THE USER OR PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 9268001 ↗
  • MalfunctionJun 27, 2018Break; Break

    IT WAS REPORTED VIA EMAIL THAT A DOCTOR EXPERIENCED AN ISSUE A DYNACORD BREAK. AFTER FOLLOWING UP WITH THE SALES REP VIA PHONE, HE STATED THAT DURING A ROTATOR CUFF PROCEDURE WHILE THE DOCTOR WAS TYING THE HALF HITCH KNOT, THE DYNACORD BROKE. HE DID NOT THINK IT NEEDED TO BE REPORTED SINCE THE BREAK WAS WHERE THE DOCTOR WAS GOING TO CUT THE SUTURE IN THE END. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7642471 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →