HEALIX ADVANCE BR ANCHOR w/DYNACORD 222004Recalls & reported problems
Medos International Sàrl · HEALIX ADVANCE BR ANCHOR w/DYNACORD TCP/PLGA Absorbable Anchor (1) Blue (1) White/Blue/Green Striped, Size 2 (5 Metric) DYNACORD Suture, 36" (91cm) 5.5mm · Class 2 · Implant
All models of HEALIX ADVANCE BR ANCHOR w/DYNACORD (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886705029464), received between Jun 27, 2018 and Jul 23, 2025 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Adverse Event Without Identified Device or Use Problem2
- Crack1
- Detachment of Device or Device Component1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient2
- Nerve Damage1
- No Consequences Or Impact To Patient1
- Perforation of Vessels1
- Unspecified Tissue Injury1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJul 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
EVENT DESCRIPTION: RE-TEARS OR NEW TEARS [POST-OPERATIVE]. OPERATIVE SIDE: RIGHT. WAS THE FMS VUE II SYSTEM USED FOR THIS PROCEDURE?: YES, WAS THE VAPR SYSTEM USED FOR THIS PROCEDURE?: YES, WAS THE SPEEDTRAP GRAFT PREPARATION SYSTEM USED FOR THIS PROCEDURE?: NO. IF THE START DATE IS EQUAL TO THE DATE OF PROCEDURE, WHEN DID THE EVENT OCCUR?: BLANK. SEVERITY: MODERATE. DID THIS EVENT RESULT IN A RE-OPERATION/REVISION? IF YES, COMPLETE THE RE-OPERATION/REVISION FORM.: YES, HAS IMPLANT CAUSED OR CONTRIBUTED TO DEATH OR SERIOUS INJURY TO THE PATIENT?: NO.
Patient problems: Unspecified Tissue Injury
MAUDE report 22575309 ↗MalfunctionOct 11, 2023Crack; Crack
IT WAS REPORTED BY SALES REP VIA EMAIL THAT DURING ROTATOR CUFF REPAIR BODY OF THE 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) CRACKED AFTER THE FIRST TWIST, NOT WORKING ON HARD BONE. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17912937 ↗MalfunctionJun 12, 2023Migration; Migration
THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) DEVICE HAD NO FIXATION AND PULLED OUT OF THE PATIENT'S BONE IMMEDIATELY. ACCORDING TO THE REPORT, THE PHYSICIAN UPSIZED TO A 6.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) DEVICE BUT PULLED AS WELL. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A FIVE MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17110047 ↗InjuryFeb 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THROUGH A CLINICAL TRIAL USING 5.5HLX ADV BR ANC-DYNA STR/BL DEVICE THAT A PATIENT HAD NEUROVASCULAR INJURY RELATED TO A BICEPS TENODESIS PROCEDURE OF THE LEFT SHOULDER THAT WAS PERFORMED ON (B)(6) 2021. ACCORDING TO THE REPORT, THE PATIENT CONSEQUENCE WAS MILD; AND THEREFORE, DID NOT REQUIRE A REVISION/REOPERATION PROCEDURE. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Nerve Damage; Perforation of Vessels
MAUDE report 13598878 ↗InjurySep 9, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE THREADS ON THE 5.5HLX ADV BR ANC-DYNA STR/BL DEVICE PULLED OFF ONCE IMPACTED. THE SURGEON COULD NOT RECOVERED THE ANCHOR AND HAD TO USE A NEW ANCHOR TO FINISH THE PROCEDURE. THE STATUS OF THE PATIENT POST-SURGERY WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 12446444 ↗MalfunctionSep 9, 2021Break; Break
IT WAS REPORTED BY THE SALES REP THAT DURING A BICEP TENODESIS PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE ANCHOR ON THE 4.5HLX ADV BR ANC-DYNA STR/BL DEVICE BROKE UPON INSERTION AS IT WAS TAP INTO THE BONE HOLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12446413 ↗InjuryNov 1, 2019Break; Break
IT WAS REPORTED VIA PHONE THAT THE SUTURE BRIDGE OF THE 5.5MM HEALIX ADVANCE¿ BR ANCHOR WITH DYNACORD¿ SUTURE (BLUE AND WHITE/BLUE/GREEN) BROKE. THIS DEVICE IS NOT AVAILABLE TO BE RETURNED FOR ANALYSIS. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE STATED THE REPORTED FAILURE OCCURRED INTRA-OP AND A SURGICAL DELAY DID NOT OCCUR. THE AFFILIATE ALSO REPORTED THAT FRAGMENTS WERE NOT GENERATED BUT A PART OF THE DEVICE REMAINED IN THE PATIENT. IT WAS ALSO REPORTED THE PROCEDURE WAS COMPLETED WITH A READILY AVAILABLE ANCHOR AND SURGICAL INTERVENTION IS NOT PLANNED. THERE WERE NO REPORTED CONSEQUENCES TO THE USER OR PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 9268001 ↗MalfunctionJun 27, 2018Break; Break
IT WAS REPORTED VIA EMAIL THAT A DOCTOR EXPERIENCED AN ISSUE A DYNACORD BREAK. AFTER FOLLOWING UP WITH THE SALES REP VIA PHONE, HE STATED THAT DURING A ROTATOR CUFF PROCEDURE WHILE THE DOCTOR WAS TYING THE HALF HITCH KNOT, THE DYNACORD BROKE. HE DID NOT THINK IT NEEDED TO BE REPORTED SINCE THE BREAK WAS WHERE THE DOCTOR WAS GOING TO CUT THE SUTURE IN THE END. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7642471 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page