GRYPHON DYNACORD 210121Recalls & reported problems
Medos International Sàrl · GRYPHON P BR DS ANCHOR w/DYNACORD TCP/PLGA Absorbable Anchor (1) Blue (1) White/Blue/Green Striped, Size 2 (5 Metric) DYNACORD Suture, 36 inches (91cm) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886705029884), received between Jun 10, 2019 and Mar 20, 2025 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Migration3
- Crack2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Injury2
- No Code Available2
- Not Applicable2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2025Migration; Migration
IT WAS REPORTED THAT DURING A SHOULDER REPAIR SURGICAL PROCEDURE THE TWO (X2) GRYPHON P BR ANC-DYNA STR/BL DEVICES WERE REPORTED TO BE NOT AS SHARP AS THEY SHOULD BE. IT WAS REPORTED THAT THE ANCHORS WERE INSERTED BUT PULLED BACK OUT AS THE HOLE CREATED WAS TOO BIG. IT WAS NOT REPORTED WHETHER ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21651066 ↗MalfunctionMar 20, 2025Migration; Migration
IT WAS REPORTED THAT DURING A SHOULDER REPAIR SURGICAL PROCEDURE THE TWO (X2) GRYPHON P BR ANC-DYNA STR/BL DEVICES WERE REPORTED TO BE NOT AS SHARP AS THEY SHOULD BE. IT WAS REPORTED THAT THE ANCHORS WERE INSERTED BUT PULLED BACK OUT AS THE HOLE CREATED WAS TOO BIG. IT WAS NOT REPORTED WHETHER ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21651061 ↗MalfunctionOct 1, 2024Migration; Migration
IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE GRYPHON® P BR DS ANCHOR WITH DYNACORD¿ DEVICE PULLED OUT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20342393 ↗MalfunctionJun 1, 2021Break; Break
IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT A GRYPHON P BR DS ANCHOR WITH DYNACORD DEVICE BROKE DURING INSERTION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11913864 ↗MalfunctionMay 26, 2021Break; Break
IT WAS REPORTED BY SALES REP THAT DURING A SHOULDER LABRUM PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE GRYPHON® P BR DS ANCHOR WITH DYNACORD¿ WAS BROKE UPON OPENING ITS PACKAGE BEFORE IT WAS ATTEMPTED TO IMPLANT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11886371 ↗InjuryJun 10, 2019Break; Crack; Break; Crack
IT WAS REPORTED BY THE SALES REP VIA EMAIL THAT DURING A BANKART REPAIR PROCEDURE TWO GRYPHON DOUBLE SUTURE ANCHORS WITH DYNACORD CRACKED IN HALF DURING IMPLANTATION. TWO OTHER GRYPHON ANCHORS WITH DYNACORD ALSO SPLIT INTO PIECES DURING THE CASE. NO FRAGMENTS WERE LEFT IN THE PATIENT. THE CASE WAS DELAYED THIRTY TO FORTY MINUTES DUE TO THE ANCHOR MALFUNCTIONS. ADDITIONAL SURGICAL INTERVENTION WAS REQUIRED, A NEW BONE HOLE HAD TO BE CREATED. THE ANCHORS WERE DISCARDED INTO THE SHARPS CONTAINER BY THE CUSTOMER. THE PROCEDURE WAS COMPLETED WITH SPARE DEVICES OF THE SAME PRODUCT.
Patient problems: Injury; No Code Available; Not Applicable
MAUDE report 8683904 ↗InjuryJun 10, 2019Break; Crack; Break; Crack
IT WAS REPORTED BY THE SALES REP VIA EMAIL THAT DURING A BANKART REPAIR PROCEDURE TWO GRYPHON DOUBLE SUTURE ANCHORS WITH DYNACORD CRACKED IN HALF DURING IMPLANTATION. TWO OTHER GRYPHON ANCHORS WITH DYNACORD ALSO SPLIT INTO PIECES DURING THE CASE. NO FRAGMENTS WERE LEFT IN THE PATIENT. THE CASE WAS DELAYED THIRTY TO FORTY MINUTES DUE TO THE ANCHOR MALFUNCTIONS. ADDITIONAL SURGICAL INTERVENTION WAS REQUIRED, A NEW BONE HOLE HAD TO BE CREATED. THE ANCHORS WERE DISCARDED INTO THE SHARPS CONTAINER BY THE CUSTOMER. THE PROCEDURE WAS COMPLETED WITH SPARE DEVICES OF THE SAME PRODUCT.
Patient problems: Injury; No Code Available; Not Applicable
MAUDE report 8683743 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page