Skip to content

HEALIX TI 3 SUTURE ANCHOR w/DYNACORD 222056Recalls & reported problems

Medos International Sàrl · HEALIX TI 3 SUTURE ANCHOR w/DYNACORD Titanium Anchor (1) Blue (1) White/Blue/Green Striped (1) White/Black Striped, Size 2 (5 Metric), DYNACORD Suture, 36" (91 cm) 5.5mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886705029952), received between Feb 25, 2025 and Feb 25, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • Unspecified Tissue Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2025Break; Break

    IT WAS REPORTED THAT DURING ROTATOR CUFF REPAIR PROCEDURE, WHILE THE 5.5 HEALIX TI ANCHOR WITH DYNACORD 3-SUT DEVICE WAS INSERTED IT WAS OBSERVED THAT THE SHAFT BROKE, LEAVING THE IMPLANT 50% PROUD OF THE CORTICAL LAYER OF BONE. ACCORDING TO THE REPORTER, THE DOCTOR ATTEMPTED TO REMOVE THE IMPLANT USING THE BROKEN SHAFT AND A SCREW REMOVAL KIT BUT WAS UNABLE. IT WAS REPORTED THAT REMOVAL WAS SUCCESSFUL BY USING A FREER AROUND THE IMPLANT, RESULTING IN BONE LOSS OF THE PATIENT. A DIFFERENT HEALIX TI SIZE WAS USED TO COMPLETE THE REPAIR. THERE WAS A FIFTEEN MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 21456826 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates