HEALIX ADVANCE DYNATAPE DYNACORD 280723Recalls & reported problems
Medos International Sàrl · HEALIX ADVANCE BR Anchor w/ DYNATAPE Suture TCP/PLGA Absorbable Anchor with (1) 2.5mm White/Black DYNATAPE Braided Flat Suture, 36inch (91 cm) and (1) Blue, Size 2 (5 metric) DYNACORD Braided Suture, 36inch (91cm), (1) Utility Shuttle Suture WHITE 4.5mm · Class 2 · Implant
All models of HEALIX ADVANCE DYNATAPE DYNACORD (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886705031023), received between Feb 25, 2022 and Sep 5, 2025 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Migration2
- Adverse Event Without Identified Device or Use Problem1
- Degraded1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Unspecified Tissue Injury1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2025Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT AFTER CREATING A BONE HOLE WITH AN AWL, THE ANCHOR ON THE 4.5 HLX ADV BR 2.5 DYNAT WHT DEVICE WAS INSERTED, BUT THE ANCHOR BROKE. THE REPORTER INDICATED THAT IT WAS POSSIBLE THAT THE PATIENT'S BONE WAS TOO HARD, BUT SINCE THE SUBSEQUENT ANCHOR DID NOT BREAK, A PRODUCT ISSUE WAS SUSPECTED. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22983252 ↗MalfunctionDec 2, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE ON THE RIGHT SHOULDER FOR A SHOULDER ROTATOR CUFF TEAR, IT WAS DISCOVERED THAT THE ANCHOR ON THE 4.5 HLX ADV BR 2.5 DYNAT WHT DEVICE BROKE. IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO INSERT THE DEVICE AFTER CREATING THE BONE HOLE BUT DISCONTINUED ITS USE WHEN THE ANCHOR PORTION BROKE IN A SPIRAL SHAPE. NO FRAGMENTS OF THE DAMAGED IMPLANT WERE LEFT IN THE BODY. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20824788 ↗MalfunctionNov 7, 2023Break; Break
IT WAS REPORTED BY SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE 4.5 MM HEALIX ADVANCE TM BR ANCHOR WITH (1) 2.5 MM DYNATAPE TM IN WHITE/ BLACK AND (1) #2 DYNACORD TM IN BLUE, WITH UTILITY SHUTTLE SUTURE NO NEEDLES, BROKE MULTIPLE TIMES. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18084278 ↗MalfunctionMay 22, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THROUGH CLINICAL RESEARCH THAT DURING A ROTATOR CUFF REPAIR PROCEDURE OF THE LEFT SHOULDER ON (B)(6) 2022, IT WAS OBSERVED THAT THE HEALIX ADVANCE BR ANCHOR W/ DYNATAPE SUTURE TCP/PLGA ABSORBABLE ANCHOR WITH (1) 2.5MM WHITE/BLACK DY DEVICE PULLED THROUGH THE ANCHOR. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16979270 ↗MalfunctionDec 30, 2022Break; Break
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE 4.5 MM HEALIX ADVANCE TM BR ANCHOR WITH (1) 2.5 MM DYNATAPE TM IN WHITE/ BLACK AND (1) #2 DYNACORD TM IN BLUE, WITH UTILITY SHUTTLE SUTURE, NO NEEDLES BROKE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16080573 ↗MalfunctionDec 15, 2022Degraded; Migration; Degraded; Migration
THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY IN-HOUSE CLINICAL RESEARCH TEAM THAT DURING A ROTATOR CUFF REPAIR/SLAP LESION REPAIR OF THE RIGHT SHOULDER ON (B)(6) 2022, IT WAS OBSERVED THAT THE 4.5 MM HEALIX ADVANCE TM BR ANCHOR WITH (1) 2.5 MM DYNATAPE TM IN WHITE/ BLACK AND (1) #2 DYNACORD TM IN BLUE, WITH UTILITY SHUTTLE SUTURE, NO NEEDLES AND THE 5.5MM HEALX ADV SP BIOCOMPOSITE ANCHOR DEVICES WERE INVOLVED IN A DEVICE MIGRATION. IT WAS FURTHER REPORTED THAT THERE WAS INADEQUATE CUTTING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15990346 ↗MalfunctionAug 1, 2022Migration; Migration
THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR WITH BICEPS TENODESIS PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE 4.5 HLX ADV BR 2.5 DYNAT WHT DEVICE PULLED OUT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15140903 ↗MalfunctionJun 27, 2022Break; Break
IT WAS REPORTED BY SALES REP THAT DURING A SHOULDER SCOPE ON (B)(6) 2022, IT WAS OBSERVED THAT WHILE USING A 4.5 MM HEALIX ADVANCE TM BR ANCHOR WITH (1) 2.5 MM DYNATAPE TM IN WHITE/ BLACK AND (1) #2 DYNACORD TM IN BLUE, WITH UTILITY SHUTTLE SUTURE NO NEEDLES DEVICE, THE ANCHOR BODY BROKE WHEN SURGEON FIRST TRIED TO ADVANCE DISTAL TIP OF ANCHOR INTO PRE PUNCHED PILOT HOLE. THEY WERE ABLE TO OPEN NEW ANCHOR AND INSERT AFTER USING TAP. SURGEON STATED HE FELT THE ANCHOR WAS FLAWED. THERE WERE NO ADVERSE PATIENT CONSEQUENCE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14834791 ↗InjuryApr 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THROUGH CLINICAL RESEARCH REPORT FOR CLINICAL EVIDENCE ON SAFETY AND PERFORMANCE OF THE HEALIX ADVANCE BR WITH DYNATAPE DEVICE THAT POSTOPERATIVELY TO A ROTATOR CUFF REPAIR PROCEDURE PERFORMED ON (B)(6) 2021, THE PATIENT HAD A REVISION PROCEDURE ON (B)(6) 2021 DUE TO A RECURRENT TEAR LIKELY SECONDARY TO SMOKING AND NONCOMPLIANCE. ACCORDING TO THE REPORT, THE PATIENT HAD A POSTOPERATIVE PAIN AND RE-TEAR ON (B)(6) 2021. THE STATUS OF THE PATIENT POSTOPERATIVELY WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) 4.5 HLX ADV BR 2.5 DYNAT WHT DEVICE. THIS COMPLAINT INVOLVES TWO (2) DEVICES.
Patient problems: Unspecified Tissue Injury
MAUDE report 14252883 ↗MalfunctionFeb 25, 2022Break; Break
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THROUGH A CLINICAL TRIAL THAT DURING A MEDIAL COLLATERAL LIGAMENT (MCL) REPAIR SURGERY OF THE RIGHT KNEE ON (B)(6) 2021, IT WAS OBSERVED THAT THE ANCHOR ON THE 4.5 HLX ADV BR 2.5 DYNAT WHT DEVICE BROKE WHILE BEING INSERTED INTO THE BONE. ACCORDING TO THE REPORT, 3/4 OF THE ANCHOR WAS IN BONE WITH EXCELLENT FIXATION; THEREFORE, A NEW ANCHOR WAS NOT USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13598933 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page