LCP 02.104.008Recalls & reported problems
Synthes GmbH · 3.5MM LCP EXTRA-ARTICLR DISTAL HUMERUS PL 8H/RT 194MM LONG · Class 2 · Implant
All models of LCP (367)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982029027), received between Dec 22, 2017 and May 13, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
Patient problems
- No Code Available1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 13, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT REVISION SURGERY DUE TO PATIENTS HUMERUS HAD A NONUNION. REMOVAL OF THE PLATE AND SCREWS WAS CONDUCTED. SURGEON THEN INSERTED A BONE GRAFT FROM THE PATIENTS TIBIA AND APPLIED A LARGE FRAGMENT PLATE. THREE OF THE CORTEX SCREWS WERE BROKEN AT THE HEAD-SHAFT JUNCTION. ADDITIONAL X-RAYS FOR THE REMOVAL OF BROKEN SCREW SHAFTS. THERE WAS A SURGICAL DELAY OF APPROXIMATELY 15 MINUTES TO SURGICAL TIME. THE PROCEDURE SUCCESSFULLY COMPLETED. THE PATIENT WAS REVISED WITH A NEW DEVICE. CONCOMITANT DEVICES REPORTED: UNKNOWN 3.5MM CORTEX SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 3). THIS COMPLAINT INVOLVES 5 DEVICES. THIS REPORT IS FOR (1) 3.5 LCP X-ARTC DSTL HUM PL 8H/RT 194 LG. THIS REPORT IS 1 OF 5 (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 11822319 ↗InjuryDec 22, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) TO REPAIR A DISTAL HUMERUS NONUNION ON (B)(6) 2017. ORIGINALLY, ON AN UNKNOWN DATE SIX MONTHS PRIOR, THE FRACTURE WAS TREATED NON-SURGICALLY WITH A SLING. SUBSEQUENTLY, A NONUNION WAS IDENTIFIED AND DISUSE OSTEOPENIA RELATED TO THE SLING WAS DIAGNOSED. THE PATIENT WAS BROUGHT TO THE OPERATING ROOM ON (B)(6) 2017. THE FRACTURE WAS REDUCED AND AN EIGHT-HOLE PLATE, SCREWS AND BONE GRAFT WERE IMPLANTED DURING AN UNEVENTFUL SURGERY. POSTOPERATIVE X-RAYS THAT SAME DAY SHOWED THAT THE PLATE WAS TOO SHORT, AND THE SURGEON DECIDED TO RETURN THE PATIENT TO THE OPERATING ROOM FOR A HARDWARE REMOVAL AND REVISION ON (B)(6) 2017. THE SURGEON EXPLANTED AND REPLACED THE EIGHT-HOLE PLATE AND PROXIMAL SCREWS WITHOUT INCIDENT. HE RE-USED THE DISTAL SCREWS. THE REVISION CONSTRUCT INCLUDED: ONE 12-HOLE PLATE, ¿90/90¿ PLATED WITH THE 12-HOLE PLATE AND SCREWS AND BONE GRAFT. THERE WERE NO REPORTED SURGICAL DELAYS, NO FRAGMENTS INVOLVED AND NO ADDITIONAL X-RAYS OR ADDITIONAL MEDICAL INTERVENTION REQUIRED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE PATIENT IN STABLE CONDITION. CONCOMITANT DEVICES: LOCKING SCREW (PART
Patient problems: No Code Available
MAUDE report 7142935 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page