VA-LCP 02.111.620Recalls & reported problems
Synthes GmbH · 2.4MM VA-LCP 2-CLMN VLR DSTL RADIUS PL 6H HD/2H SHAFT/RIGHT · Class 2 · Implant
All models of VA-LCP (256)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982031952), received between Dec 24, 2019 and Mar 19, 2021 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K083694); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Twisted/Bent2
- Device-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE THREADS ON EITHER THE VARIABLE ANGLE LOCKING COMPRESSION PLATE (VA-LCP) TWO COLUMN VOLAR DISTAL RADIUS PLATE OR THE FOUR (4) VARIABLE ANGLE (VA) LOCKING SCREWS STARDRIVE WERE STRIPPED PRIOR TO THE CASE. EACH SCREW INSERTED INTO THE TWO RADIAL STYLOID HOLES WOULD SPIN AND NOT LOCK IN. SURGEON HAD TO REMOVE THE PLATE ALL TOGETHER AND START OVER. ANOTHER SET WAS OPENED TO GET A NEW PLATE. THE SURGERY WAS DELAYED FOR THIRTY (30) MINUTES DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) 2.4MM VA-LCP 2-CLMN VLR DSTL RADIUS PL 6H HD/2H SHAFT/RIGHT. THIS IS REPORT 1 OF 5 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11530649 ↗InjuryMar 11, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE THAT THE 1.25MM K-WIRE BROKE OFF AND GOT STUCK IN THE DISTAL HOLE OF THE 6-HOLE VOLAR PLATE. UNABLE TO REMOVE AND HAD TO MOVE TO A 7-HOLE PLATE. THE SURGERY WAS COMPLETED WITHOUT DELAY. THE PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT ONE (2) DEVICE. THIS IS 1 OF 2 FOR REPORT (B)(4).
Patient problems: No Code Available
MAUDE report 9820552 ↗MalfunctionDec 24, 2019Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON (B)(6) 2019 DURING A DISTAL RADIUS FRACTURE SURGERY FOR A BROKEN WRIST WHEN THE SURGEON USED A SCREW AND GUIDE TO SCREW IN THE MOST DISTAL SCREW INTO THE PLATE, THE SCREW WENT THROUGH THE PLATE. THE PLATE AND SCREW WERE REMOVED AND A NEW PLATE WAS USED. THERE WAS A SURGICAL DELAY OF FIVE (5) MINUTES. THE SURGERY WAS COMPLETED WITH NO ADVERSE CONSEQUENCE TO THE PATIENT. CONCOMITANT DEVICES REPORTED: UNKNOWN SCREW (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# 1), UNKNOWN DRILL BIT (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# 1), UNKNOWN GUIDE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# 1). THIS REPORT IS FOR ONE 2.4MM VA-LCP 2-CLMN VLR DSTL RADIUS PL 6H HD/2H SHAFT/RIGHT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9516502 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083694. One submission often covers many models, so check the product and lot details in each record.
- Z-2087-2013 · Jul 30, 2013 · TerminatedSynthes 2.4 VA-LCP 2 CLMN VLR DSTL Radius PL 7H HD/3H Shaft/Right Product Usage: Intended for fixation of complex intra-and extra-articular fractures and osteotomies of the dista
- Z-2057-2013 · May 31, 2013 · TerminatedSynthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page