Skip to content

LCP 02.114.506Recalls & reported problems

Synthes GmbH · 1.5MM LCP T-PLATE/3H HEAD 8 HOLE SHAFT WITH GUIDES · Class 2 · Implant

All models of LCP (367)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982035028), received between Dec 7, 2017 and Dec 7, 2017 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 7, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORT THAT ON (B)(6) 2017, PATIENT UNDERWENT IMPLANT SURGERY FOR TREATMENT OF A (ORIF) OPEN REDUCTION INTERNAL FIXATION TO THE RIGHT FOURTH METACARPAL BONE DUE TO A BONE FRACTURE WHERE THE SURGEON WAS USING THE 1.5MM (LCP)LOCKING COMPRESSION SYSTEM. DURING THE PRODUCE, SURGEON CHOSE TO USE ONE (1) 1.5MM LCP T-PLATE AND THREE (3) 1.5MM CORTEX SELF TAPPING SCREWS. SURGEON ORIGINALLY WANTED TO USE A CONDYLAR PLATE FOR THE FRACTURE REDUCTION, BUT THIS PLATE WAS UNAVAILABLE IN THE LCP SET. (DUE TO THIS PLATE WAS USED DURING A SURGERY THE DAY BEFORE.) SURGEON CUT AND CONTOURED THE T-PLATE TO THE APPROPRIATE LENGTH AND ATTEMPTED TO INSERT THREE (3) CORTEX SCREWS. SURGEON STRUGGLED WITH THREE (3) DIFFERENT SCREWDRIVERS WHEN ATTEMPTING TO IMPLANT EACH OF THE THREE (3) CORTEX SCREWS. ALL THREE SCREWDRIVERS BECAME STRIPPED AND WOULD NO LONGER RETAIN THE SCREWS. ADDITIONAL SCREWDRIVERS WERE AVAILABLE IN THE OPERATING ROOM TO CONTINUE THE SURGERY. AT THIS POINT IN THE PROCEDURE, SURGEON REMOVED THE PREVIOUSLY IMPLANTED THREE CORTEX SCREWS AND PLATE. PATIENT WAS ULTIMATELY IMPLANTED WITH TWO (2) 1.5MM CORTEX SCREWS AND FIBER WIRE SUTURES TO COMPRESS THE BONE FRACTURE. SURGERY WAS COMPLET

    Patient problems: No Code Available

    MAUDE report 7096546 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page