The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982035882), received between May 3, 2021 and Apr 27, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 27, 2026Material Twisted/Bent

    IT WAS REPORTED THAT DRIVERS ARE STRIPPED AND NOT FITTING INTO SCREWS. PLATE'S LOCKING HOLES DID NOT ALLOW THE LOCKING SCREWS TO THREAD INTO THE PLATE, CAUSING THE SCREWS TO SPIN WITHOUT ENGAGING. THIS RESULTED IN A SURGICAL DELAY OF 30 MINUTES. THERE WAS NO REPORTED PATIENT OR USER HARM OR CLINICAL CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25008756 ↗
  • MalfunctionAug 20, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021, WHILE PERFORMING A DISTAL HUMERUS, THE SURGEON HAD TO RE-ADJUST LOCKING SCREWS MULTIPLE TIMES ON A VARIABLE ANGLE-LOCKING COMPRESSION PLATE (VA-LCP) POSTEROLATERAL DISTAL HUMERUS PLATE WHILE TRYING TO ACHIEVE REDUCTION OF THE SMALL FRAGMENTS. ULTIMATELY ONE OR TWO OF THE 2.7 VARIABLE ANGLE (VA) LOCKING SCREWS DID NOT ENGAGE IN A PLATE. SURGEON DECIDED TO LEAVE THESE HOLES EMPTY. HE DETERMINED THAT HE HAD LIKELY DAMAGED THE HOLES RE-INSERTING THE SCREWS MULTIPLE TIMES. ALSO WHILE REDUCING, AN ADJUSTMENT NUT FELL OFF OF THE REDUCTION FORCEPS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF TEN (10) MINUTES. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) 2.7MM/3.5MM VA-LCP POSTLAT DHP-LAT SUPT 3H/LT/75MM-SHORT. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12350626 ↗
  • MalfunctionMay 3, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE 2.7MM VARIABLE ANGLE LOCKING SCREWS WOULD NOT LOCK INTO THE FOURTH MOST PROXIMAL HOLE OF THE 2.7MM/3.5MM VARIABLE ANGLE LCP POSTEROLATERAL DISTAL HUMERUS PLATES WITH LATERAL SUPPORT. BOTH IMPLANTS REMAINED IN THE PATIENT. THE 4.0MM CANNULATED SCREW SHORT AND 4.0MM CANNULATED SCREW SHORT THREAD 64MM WERE STRIPPED OR THE CANNULATED HEX SCREWDRIVER WAS STRIPPED. A SECOND SET FOR ANOTHER SCREWDRIVER WAS OPENED AND THE HANDLE BROKE. BOTH SCREWS REMAINED IN THE PATIENT. THERE WAS TEN (10) MINUTES OF SURGICAL DELAY. THE PROCEDURE OUTCOME IS UNKNOWN. THE PATIENT OUTCOME WAS GOOD. THIS REPORT INVOLVES ONE (1) 2.7MM/3.5MM VA-LCP POSTLAT DHP-LAT SUPT 3H/LT/75MM-SHORT. THIS IS REPORT 5 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11760500 ↗