VA-LCP 02.117.303SRecalls & reported problems
Synthes GmbH · 2.7MM/3.5MM VA-LCP POSTLAT DHP 3H/LT/75MM-SHORT-STER · Class 2 · Implant
All models of VA-LCP (256)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982036339), received between Oct 21, 2024 and Oct 21, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2024, DURING A CASE THE SURGEON WAS TRYING TO INSERT LOCKING SCREWS IN THE DISTAL HEAD OF THE PLATE WHEN 2 OF THE SCREWS DID NOT LOCK IN. THE SURGEON TRIED MULTIPLE TIMES AND USED DIFFERENT SCREWS, BUT IT WOULD NOT LOCK. WE HAD TO USE METAPHYSEAL SCREWS INSTEAD. THERE WAS NO SURGICAL DELAY. THE SURGEON TRIED DIFFERENT LOCKING SCREWS BUT IT DID NOT WORK. USED A TYPE OF NON-LOCKING SCREW INSTEAD. PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT OUTCOME WAS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 20495001 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page