Skip to content

VA-LCP 02.117.504Recalls & reported problems

Synthes GmbH · 2.7MM/3.5MM VA-LCP MEDL DSTL HUMERUS PL 4H/LT/108MM-LONG · Class 2 · Implant

All models of VA-LCP (256)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982036780), received between Jan 8, 2019 and Mar 13, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

1 other recall cover devices cleared under the same FDA submission (K120070); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • Nonunion/Delayed-union Bone Fracture1
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PLATES WERE IMPLANTED IN THE PATIENT, THE LOCKING SCREWS DID NOT LOCK IN 1-2 HOLES OF THE PL AND MEDIAL PLATES SO THOSE PLATE HOLES WERE NOT FILLED. THE SURGEON INITIALLY USED A VA GUIDE BUT LATER USED A NON LOCKING DRILL GUIDE. UNSURE IF IT¿S BECAUSE HE USED A METAPHYSEAL SCREW FIRST TO SUCK THE PLATE DOWN, AND THEN CHANGED FOR A VAL. WE¿VE HAD ISSUES WITH THIS PLATE NOT LOCKING, SURGEON DID NOT WANT PATIENT TO BE CHARGED. WHILE INSERTING CORTICAL SCREWS AND A 7.3 CANNULATED FOR THE OLECRANON OSTEOTOMY HE ALSO COMPLAINED OF THE SCREWDRIVER HEADS BEING STRIPPED OR WARPED. THERE WERE NO PATIENT OUTCOMES REPORTED. THIS REPORT IS FOR ONE (1) 2.7/3.5 VA MEDL DSTL HUM PL 4H/LT/108-LG. THIS IS REPORT 1 OF 7 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18897644 ↗
  • InjuryApr 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2019, THE PATIENT WAS INVOLVED IN AN INCIDENT AT WORK THAT RESULTED IN MULTIPLE INJURIES, ONE OF WHICH WAS A FRACTURE TO HIS ARM. TO REPAIR THE FRACTURE, THE SURGEON PLACED AN IMPLANT IN THE PATIENT'S ARM. ON (B)(6) 2019, THE IMPLANT BROKE CAUSING ANOTHER FRACTURE TO THE PATIENT'S ARM. ON (B)(6) 2019, THE PATIENT UNDERWENT A REVISION PROCEDURE TO REMOVE THE BROKEN IMPLANT. THIS IS REPORT 15 OF 17 FOR (B)(4).

    Patient problems: Unspecified Infection

    MAUDE report 14091257 ↗
  • InjuryJan 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED ON UNKNOWN DATE, A PATIENT UNDERWENT REVISION WITH A 3.5 MM DISTAL HUMERUS EXTRA ARTICULAR PLATE 8 HOLES AND 8 SCREWS DUE TO NON-UNION. THE PATIENT HAD HARDWARE REMOVAL OF THE FOLLOWING DEVICES: TWO (2) VARIABLE ANGLE ELBOW PLATE, ONE (1) UNKNOWN 3.5 LOCKING SCREW, FIVE (5) UNKNOWN 3.5 CORTEX SCREW, TWO (2) UNKNOWN 2.7 MM METAPHYSEAL SCREW , FOUR (4) UNKNOWN 2.7 VA LOCKING SCREW, AND ONE (1) UNKNOWN 2.7 CORTEX SCREW. IT WAS UNKNOWN IF THERE WAS SURGICAL DELAY. THE REMOVAL AND PROCEDURE WENT WELL. PATIENT OUTCOME WAS FINE. THIS COMPLAINT INVOLVES SEVEN(7) DEVICES. THIS REPORT IS FOR ONE (1) 2.7MM/3.5MM VA-LCP MEDL DSTL. THIS REPORT IS 1 OF 7 FOR (B)(4).

    Patient problems: No Code Available; Nonunion/Delayed-union Bone Fracture

    MAUDE report 8226406 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120070. One submission often covers many models, so check the product and lot details in each record.

  • Z-0124-2016 · Sep 28, 2015 · Terminated
    Synthes 2.7MM/3.5MM VA-LCP Medial Distal Humerus Plates 10H/RT/186mm- Sterile and Synthes 2.7MM/3.5MM VA-LCP Extended Medial Distal Humerus Plates 10H/RT/189mm-Sterile The Synthes

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates