Skip to content

VA-LCP 02.117.604Recalls & reported problems

Synthes GmbH · 2.7MM/3.5MM VA-LCP EXT MEDL DSTL HUM PL 4H/RT/111MM-LONG · Class 2 · Implant

All models of VA-LCP (256)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982037008), received between Feb 12, 2018 and Dec 10, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K120070); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1

Patient problems

  • Failure of Implant1
  • Nonunion/Delayed-union Bone Fracture1
  • Post Operative Wound Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT A HARDWARE REMOVAL DUE TO INFECTION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THE PATIENT STATUS IS UNKNOWN. THIS IS REPORT 3 OF 6 FOR (B)(4). THIS (B)(4) CAPTURES THE INFECTION POSTOPERATIVELY WHILE THIS (B)(4) CAPTURES THE SCREWS THAT BROKE DURING REMOVAL INTRAOPERATIVELY.

    Patient problems: Post Operative Wound Infection

    MAUDE report 10984425 ↗
  • InjuryFeb 12, 2018Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT OPEN REDUCTION INTERNAL FIXATION (ORIF) RIGHT ELBOW, ON AN UNKNOWN DATE. THE PATIENT PRESENTED WITH A FRACTURE NONUNION IN THE HUMERUS AND A BROKEN PLATE ON (B)(6) 2018. THE PATIENT WAS REVISED TO A TOTAL ELBOW AFTER HAVING THE FOLLOWING HARDWARE REMOVED: TWO (2) PLATES, ONE (1) OLECRANON NAIL, ONE (1) OLECRANON END CAP, AND FOURTEEN (14) SCREWS. ALL HARDWARE WAS REMOVED INTACT EXCEPT FOR THE MEDIAL DISTAL HUMERUS PLATE WHICH WAS BROKEN. ALL HARDWARE WAS REMOVED WITH NO DELAYS OR COMPLICATIONS. THERE ARE NO X-RAYS. CONCOMITANT DEVICE REPORT: 2.7 MM/3.5 MM VA-LCP POSTLAT DSTL HUM PL/4H/RT/88 MM-MEDIUM (PART 02.117.204, LOT UNKNOWN, QUANTITY 1), OLECRANON OSTEOTOMY NAIL (PART 02.007.001, LOT UNKNOWN, QUANTITY 1), OLECRANON OSTEOTOMY END CAP (PART 02.007.002, LOT UNKNOWN, QUANTITY 1) AND UNKNOWN SCREWS (PART UNKNOWN, LOT UNKNOWN, QUANTITY 14). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 7263367 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120070. One submission often covers many models, so check the product and lot details in each record.

  • Z-0124-2016 · Sep 28, 2015 · Terminated
    Synthes 2.7MM/3.5MM VA-LCP Medial Distal Humerus Plates 10H/RT/186mm- Sterile and Synthes 2.7MM/3.5MM VA-LCP Extended Medial Distal Humerus Plates 10H/RT/189mm-Sterile The Synthes

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page