Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982038807), received between Oct 17, 2019 and Nov 13, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility3
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED, THAT ON (B)(6) 2024. ORIF ANKLE WAS PERFORMED AT HOSPITAL. AS THE PLATE FAILED TO LOCK WITH THE SCREWS IN THE ANKLE SURGERY. STAFF TRIED MULTIPLE DIFFERENT SCREWS, DRILLED ON ANGLE & VA, TRIED A DIFFERENT PLATE, YET THE LOCKING SCREWS WOULD NOT LOCK INTO THE DISTAL CLUSTER. DUE TO THE PLATE NOT LOCKING, IT CAUSED, UNNECESSARY DRILLING OF BONE AND ELONGATED TIME UNDER ANESTHESIA. THERE WAS A SURGICAL DELAY OF 1 HOUR. SURGEON SWAPPED THE PLATE OUT FOR FUNCTIONAL HARDWARE. THERE WAS NO PATIENT STATUS OUTCOME CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20675276 ↗
  • MalfunctionJul 28, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THAT TWO VARIABLE ANGLE LOCKING SCREWS FAILED TO LOCK INTO TWO DISTAL HOLES IN A VARIABLE ANGLE LATERAL CLAVICLE PLATE. THE PLATE HAD TO BE REMOVED AND REPLACED WITH ANOTHER WHICH WORKED AS REQUIRED. USER FOLLOWED IFU. OPERATING TIME WAS EXTENDED. SURGERY WAS DELAYED FOR 20 MINUTES. FRAGMENTS WERE NOT GENERATED . THERE WAS PATIENT OUTCOME/ CONSEQUENCES BUT THERE WAS NO CLINICAL OUTCOME EXPERIENCED BY THE PATIENT.(INFECTION, INFLAMMATION..). THIS REPORT IS FOR ONE (1) VA-LCP LAT DIST FIBULA PL 2.7. THIS IS REPORT 1 OF 3 FOR COMPLAINT: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15121122 ↗
  • MalfunctionDec 29, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2020, THE VARIABLE ANGLE LOCKING COMPRESSION PLATE HOLES WERE STRIPPED, A VARIABLE ANGLE CONE WAS USED IN THE VA DISTAL FIBULA PLATES, AND A LOCKING DRILL TOWER IN THE LARGE FRAGMENT PLATE. THERE WAS A FIVE (5) MINUTE SURGICAL DELAY. THE PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: 4.5MM BROAD LCP® PLATE 7 HOLES/134MM (PART NUMBER 226.571, LOT UNKNOWN, QUANTITY 1), 2.7MM VA-LCP LATERAL DISTAL FIBULA PLATE/5 HOLES/LEFT (PART NUMBER 02.118.405, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES ONE (1) 2.7MM VA-LCP LATERAL DISTAL FIBULA PLATE/3 HOLES/LEFT. THIS IS REPORT 2 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11088383 ↗
  • MalfunctionOct 17, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2019, TWO (2) OF THE VARIABLE ANGLE (VA) DISTAL LOCKING SCREWS DID NOT LOCK INTO THE VARIABLE ANGLE - LOCKING COMPRESSION PLATE (VA-LCP) DURING AN OPEN REDUCTION AND INTERNAL FIXATION (ORIF) OF THE ANKLE. THE SCREWS WERE LEFT IN PLACE AND THERE IS NO PLAN FOR REVISION. THERE WAS NO SURGICAL DELAY REPORTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT OUTCOME. THIS REPORT IS FOR ONE (1) 2.7MM VA-LCP LATERAL DISTAL FIBULA PLATE/3 HOLES/LEFT. THIS IS REPORT 1 OF 3 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient