NA 02.130.208Recalls & reported problems
Synthes GmbH · 1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 8MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982044624), received between Apr 11, 2018 and Apr 17, 2023 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Failure To Adhere Or Bond” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure To Adhere Or Bond2
- Device-Device Incompatibility1
- Material Integrity Problem1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Code Available2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2023 THE DEVICE WAS NOT ENGAGING. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE (1) 1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 8MM THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16756255 ↗MalfunctionMar 17, 2022Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN HAND PROCEDURE, THE STARDRIVE SCREWS WERE DISCOVERED AS DAMAGED. NO SURGICAL DELAY OR PATIENT CONSEQUENCE REPORTED. THIS REPORT INVOLVES ONE (1) SCREW. THIS IS REPORT 5 OF 6 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13792146 ↗MalfunctionJan 14, 2022Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6), 2021, DURING AN UNKNOWN HAND PROCEDURE, THE STARDRIVE SCREWS WERE DISCOVERED AS DAMAGED. NO SURGICAL DELAY OR PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) 1.5MM VA-LCKNG SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 8MM THIS IS REPORT 1 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13257904 ↗InjuryApr 11, 2018Failure To Adhere Or Bond; Failure To Adhere Or Bond
IT WAS REPORTED THAT TWO (2) LOCKING SCREWS WOULD NOT LOCK INTO AN UNKNOWN PLATE DURING A HAND FRACTURE REPAIR PROCEDURE ON (B)(6) 2018. THE SURGEON WAS MANUALLY PLACING SCREWS THROUGH THE SCREW HOLE OF THE PLATE WHEN THE TWO SCREWS MALFUNCTIONED. THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT USING TWO CORTEX SCREWS. NO PATIENT HARM WAS REPORTED. THERE WAS A 45 SECOND SURGICAL DELAY RELATED TO SWITCHING THE SCREWS. THE PATIENT OUTCOME WAS REPORTED STABLE. PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Code Available
MAUDE report 7421354 ↗InjuryApr 11, 2018Failure To Adhere Or Bond; Failure To Adhere Or Bond
IT WAS REPORTED THAT TWO (2) LOCKING SCREWS WOULD NOT LOCK INTO AN UNKNOWN PLATE DURING A HAND FRACTURE REPAIR PROCEDURE ON (B)(6) 2018. THE SURGEON WAS MANUALLY PLACING SCREWS THROUGH THE SCREW HOLE OF THE PLATE WHEN THE TWO SCREWS MALFUNCTIONED. THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT USING TWO CORTEX SCREWS. NO PATIENT HARM WAS REPORTED. THERE WAS A 45 SECOND SURGICAL DELAY RELATED TO SWITCHING THE SCREWS. THE PATIENT OUTCOME WAS REPORTED STABLE. PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Code Available
MAUDE report 7421352 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page