Skip to content

NA 02.130.355Recalls & reported problems

Synthes GmbH · 2.0MM VAL CONDYLAR PLATE 2 HOLES HD-6 HOLES SHAFT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982045416), received between Apr 19, 2022 and Sep 20, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Break1

Patient problems

  • Deformity/ Disfigurement1
  • Discomfort1
  • Malunion of Bone1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 20, 2024Break; Break

    IT WAS REPORTED THAT DURING SURGERY ON (B)(6) 2024, THE PLATE BROKE WHILE BENDING. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. FRAGMENTS WERE GENERATED AND WERE REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20268060 ↗
  • InjuryApr 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HAND PLATES. IN COLLABORATION WITH DEPUY SYNTHES, TOGETHER WITH CLEVELAND CLINIC FOUNDATION (CCF) DATABASE DESIGNERS, PROGRAM MANAGERS, RESEARCH ASSISTANTS, AND BIOSTATISTICIANS, HAS DEVELOPED A REPORT THAT SUMMARIZES SAFETY AND EFFECTIVENESS DATA ON 67 PATIENTS WHO UNDERWENT HAND SURGERY AT CCF BETWEEN 2008 AND 2023 UTILIZING DEPUY SYNTHES (DPS) HAND PLATES. IMPLANTS INVOLVED WERE LCP COMPACT HAND PLATES OR VA LOCKING HAND SYSTEM. 3 PATIENTS WERE NOTED TO HAVE MORE THAN ONE PLATE IMPLANTED. DPS SCREWS WERE INVOLVED. PATIENT #(B)(6) 58-YEAR-OLD MALE WAS IMPLANTED WITH A VA LOCKING HAND PLATES (02.130.366) WITH 6 SCREWS AND VA LOCKING HAND SYSTEM (02.130.355) WITH 7 SCREWS. POST OP COMPLICATION WAS REPORTED FOR MALROTATION AND MALUNION OF LEFT 4TH METACARPAL SHAFT FRACTURE. THE COMPLICATION IS NOTED TO BE RELATED TO THE PROCEDURE OR IMPLANTS. THERE IS NO DEVICE FAILURE NOTED. TREATMENT: 1. REMOVAL OF DEEP HARDWARE, LEFT HAND. 2. CORRECTIVE OSTEOTOMY, LEFT 4TH METACARPAL SHAFT. THIS REPORT IS FOR A 2.0MM VAL CONDYLAR PLATE 2 HOLES HD-6 HOLES SHAFT. THIS IS REPORT 1 OF 8 FOR (B)(4).

    Patient problems: Deformity/ Disfigurement; Malunion of Bone

    MAUDE report 19088390 ↗
  • InjuryApr 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT HARDWARE REMOVAL PROCEDURE DUE TO DISCOMFORT. THE DEVICES EXPLANTED WERE ONE (1) 2.7MM/3.5MM VA-LCP PROXIMAL OLECRANON PLATE, ONE (1) 2.0MM VAL CONDYLAR PLATE, TWO (2) 3.5MM CORTEX SCREWS, FOUR (4) 2.7MM VA LOCKING SCREWS AND FIVE (5) 2.0MM VA LOCKING SCREWS. ALL IMPLANTS WERE REMOVED INTACT AND BONE HAD SUCCESSFULLY HEALED. THIS REPORT IS FOR ONE (1) 2.0MM VAL CONDYLAR PLATE 2 HOLES HD-6 HOLES SHAFT. THIS IS REPORT 2 OF 13 FOR COMPLAINT (B)(4).

    Patient problems: Discomfort

    MAUDE report 14142061 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page