NA 02.210.088Recalls & reported problems
Synthes GmbH · 1.8MM VA LOCKING BUTTRESS PIN WITH T8 STARDRIVE RECESS/18MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982052728), received between Feb 26, 2019 and Jun 6, 2022 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).
4 other recalls cover devices cleared under the same FDA submission (K092556, K102694); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility5
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 6, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2022, DURING AN ORTHOPEDIC PROCEDURE, THE SCREWS WOULD NOT LOCK IN 2 VARIABLE ANGLE (VA) DISTAL RADIUS PLATES. MULTIPLE LOCKING BUTTRESS PINS AND LOCKING SCREWS WERE TRIED ON 2 DIFFERENT PLATES. THEY COULD NOT GET THE LOCKING SCREWS TO ENGAGE THE PLATE. A SCREW IN DRILL GUIDE WOULD LOCK IN CORRECTLY HOWEVER, EVENTUALLY A 6 HOLE LCP VOLAR COLUMN PLATE WAS USED THAT WORKED FINE. ADDITIONAL X-RAYS WERE TAKEN. THERE WAS A SURGICAL DELAY OF MORE THAN 30 MINUTES. THE PROCEDURE COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS IS REPORT 11 OF 11 FOR (B)(4). THIS REPORT IS FOR A BUTTRESS PIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14602033 ↗MalfunctionJun 6, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2022, DURING AN ORTHOPEDIC PROCEDURE, THE SCREWS WOULD NOT LOCK IN 2 VARIABLE ANGLE (VA) DISTAL RADIUS PLATES. MULTIPLE LOCKING BUTTRESS PINS AND LOCKING SCREWS WERE TRIED ON 2 DIFFERENT PLATES. THEY COULD NOT GET THE LOCKING SCREWS TO ENGAGE THE PLATE. A SCREW IN DRILL GUIDE WOULD LOCK IN CORRECTLY HOWEVER, EVENTUALLY A 6 HOLE LCP VOLAR COLUMN PLATE WAS USED THAT WORKED FINE. ADDITIONAL X-RAYS WERE TAKEN. THERE WAS A SURGICAL DELAY OF MORE THAN 30 MINUTES. THE PROCEDURE COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS IS REPORT 10 OF 11 FOR (B)(4). THIS REPORT IS FOR A BUTTRESS PIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14601930 ↗MalfunctionJun 6, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2022, DURING AN ORTHOPEDIC PROCEDURE, THE SCREWS WOULD NOT LOCK IN 2 VARIABLE ANGLE (VA) DISTAL RADIUS PLATES. MULTIPLE LOCKING BUTTRESS PINS AND LOCKING SCREWS WERE TRIED ON 2 DIFFERENT PLATES. THEY COULD NOT GET THE LOCKING SCREWS TO ENGAGE THE PLATE. A SCREW IN DRILL GUIDE WOULD LOCK IN CORRECTLY HOWEVER, EVENTUALLY A 6 HOLE LCP VOLAR COLUMN PLATE WAS USED THAT WORKED FINE. ADDITIONAL X-RAYS WERE TAKEN. THERE WAS A SURGICAL DELAY OF MORE THAN 30 MINUTES. THE PROCEDURE COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS IS REPORT 9 OF 11 FOR (B)(4). THIS REPORT IS FOR A BUTTRESS PIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14601787 ↗MalfunctionJun 6, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2022, DURING AN ORTHOPEDIC PROCEDURE, THE SCREWS WOULD NOT LOCK IN 2 VARIABLE ANGLE (VA) DISTAL RADIUS PLATES. MULTIPLE LOCKING BUTTRESS PINS AND LOCKING SCREWS WERE TRIED ON 2 DIFFERENT PLATES. THEY COULD NOT GET THE LOCKING SCREWS TO ENGAGE THE PLATE. A SCREW IN DRILL GUIDE WOULD LOCK IN CORRECTLY HOWEVER, EVENTUALLY A 6 HOLE LCP VOLAR COLUMN PLATE WAS USED THAT WORKED FINE. ADDITIONAL X-RAYS WERE TAKEN. THERE WAS A SURGICAL DELAY OF MORE THAN 30 MINUTES. THE PROCEDURE COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS IS REPORT 4 OF 11 FOR (B)(4). THIS REPORT IS FOR A BUTTRESS PIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14601219 ↗MalfunctionFeb 26, 2019Device-Device Incompatibility; Device-Device Incompatibility
CONCOMITANT DEVICES REPORTED: 1.8MM DRILL BIT (PART # 310.509 , LOT UNKNOWN, QUANTITY 1), VARIABLE ANGLE GUIDE (PART UNKNOWN, LOT UNKNOWN, QUANTITY 1).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8371387 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092556, K102694. One submission often covers many models, so check the product and lot details in each record.
- Z-2389-2016 · Jun 28, 2016 · TerminatedSynthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone,
- Z-0323-2016 · Nov 2, 2015 · Terminated2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone frag
- Z-2057-2013 · May 31, 2013 · TerminatedSynthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and
- Z-0956-2013 · Jan 11, 2013 · TerminatedVariable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Cor
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page