NA 02.214.115Recalls & reported problems
Synthes GmbH · 1.5MM CORTEX SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 15MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982055491), received between May 14, 2020 and May 14, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- Foreign Body In Patient3
- No Code Available3
- Tissue Damage3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 14, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, DURING OPEN REDUCTION AND INTERNAL FIXATION (ORIF) OF THE THIRD PROXIMAL PHALANX, THE 3 1.5MM CORTEX SCREW SELF TAPPING WITH STARDRIVE RECESS 15MM HEADS BROKE OFF WHILE INSERTING OR REMOVING THE SCREWS. SURGEON ATTEMPTED TO RETRIEVE THE SCREWS AND THE SCREW SHAFTS BUT WAS UNSUCCESSFUL. THERE WAS A LESS THAN OPTIMAL FIXATION DUE TO BROKEN SCREW HEADS AND LONG SCREW WHICH MAY CAUSE IRRITATION WAS UNABLE TO BE RETRIEVED. SURGEON WAS UNABLE TO REMOVE SCREW SHAFTS WITHOUT EXCESSIVE DISSECTION. SCREW SHAFTS WERE NOT REMOVED. THERE WERE FRAGMENTS GENERATED. THERE WAS A SURGICAL DELAY OF 20 MINUTES. PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES 3 DEVICES. THIS REPORT IS FOR 1 1.5MM CORTEX SCREW SLF-TPNG WITH STARDRIVE RECESS 15MM. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: Foreign Body In Patient; No Code Available; Tissue Damage
MAUDE report 10058973 ↗InjuryMay 14, 2020Break; Break
CONCOMITANT DEVICE REPORTED: SCREWDRIVER (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1).
Patient problems: Foreign Body In Patient; No Code Available; Tissue Damage
MAUDE report 10058964 ↗InjuryMay 14, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, DURING OPEN REDUCTION AND INTERNAL FIXATION (ORIF) OF THE THIRD PROXIMAL PHALANX, THE THREE (3) 1.5MM CORTEX SCREW SELF TAPPING WITH STARDRIVE RECESS 15MM HEADS BROKE OFF WHILE INSERTING OR REMOVING THE SCREWS. SURGEON ATTEMPTED TO RETRIEVE THE SCREWS AND THE SCREW SHAFTS BUT WAS UNSUCCESSFUL. THERE WAS A LESS THAN OPTIMAL FIXATION DUE TO BROKEN SCREW HEADS AND LONG SCREW WHICH MAY CAUSE IRRITATION WAS UNABLE TO BE RETRIEVED. SURGEON WAS UNABLE TO REMOVE SCREW SHAFTS WITHOUT EXCESSIVE DISSECTION. SCREW SHAFTS WERE NOT REMOVED. THERE WERE FRAGMENTS GENERATED. THERE WAS A SURGICAL DELAY OF 20 MINUTES. PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES 3 DEVICES. THIS REPORT IS 1 1.5MM CORTEX SCREW SLF-TPNG WITH STARDRIVE RECESS 15MM. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: Foreign Body In Patient; No Code Available; Tissue Damage
MAUDE report 10058963 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page