Synthes GmbH · 5.0MM VARIABLE ANGLE LOCKNG SCREW/SLF-TPNG/STRDRV/80MM · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10886982062239), received between Mar 9, 2018 and Jun 10, 2022 — classified as 11 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K083025, K110354); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem10
  • Material Twisted/Bent2
  • Migration1

Patient problems

  • Nonunion/Delayed-union Bone Fracture9
  • No Code Available3
  • Unspecified Infection3
  • No Clinical Signs, Symptoms or Conditions2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022 THERE WAS A REMOVAL OF THE VARIABLE ANGLE (VA) CONDYLE PLATE AND CORRESPONDING SCREWS DUE TO NONUNION/MALUNION. THE SURGEON REMOVED THE PLATE AND SCREWS, REALIGNED THE FEMUR, INSERTED IM NAIL, AND INSERTED BONE GRAFT OBTAINED VIA RADIOIMMUNOASSAY (RIA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/80. THIS IS REPORT 12 OF 12 FOR COMPLAINT (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 14654905 ↗
  • InjuryJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022 THERE WAS A REMOVAL OF THE VARIABLE ANGLE (VA) CONDYLE PLATE AND CORRESPONDING SCREWS DUE TO NONUNION/MALUNION. THE SURGEON REMOVED THE PLATE AND SCREWS, REALIGNED THE FEMUR, INSERTED IM NAIL, AND INSERTED BONE GRAFT OBTAINED VIA RADIOIMMUNOASSAY (RIA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/80. THIS IS REPORT 11 OF 12 FOR COMPLAINT (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 14654759 ↗
  • InjuryJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022 THERE WAS A REMOVAL OF THE VARIABLE ANGLE (VA) CONDYLE PLATE AND CORRESPONDING SCREWS DUE TO NONUNION/MALUNION. THE SURGEON REMOVED THE PLATE AND SCREWS, REALIGNED THE FEMUR, INSERTED IM NAIL, AND INSERTED BONE GRAFT OBTAINED VIA RADIOIMMUNOASSAY (RIA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/80. THIS IS REPORT 10 OF 12 FOR COMPLAINT (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 14654685 ↗
  • InjuryJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022 THERE WAS A REMOVAL OF THE VARIABLE ANGLE (VA) CONDYLE PLATE AND CORRESPONDING SCREWS DUE TO NONUNION/MALUNION. THE SURGEON REMOVED THE PLATE AND SCREWS, REALIGNED THE FEMUR, INSERTED IM NAIL, AND INSERTED BONE GRAFT OBTAINED VIA RADIOIMMUNOASSAY (RIA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/80. THIS IS REPORT 7 OF 12 FOR COMPLAINT (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 14654623 ↗
  • InjuryJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022 THERE WAS A REMOVAL OF THE VARIABLE ANGLE (VA) CONDYLE PLATE AND CORRESPONDING SCREWS DUE TO NONUNION/MALUNION. THE SURGEON REMOVED THE PLATE AND SCREWS, REALIGNED THE FEMUR, INSERTED IM NAIL, AND INSERTED BONE GRAFT OBTAINED VIA RADIOIMMUNOASSAY (RIA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/80. THIS IS REPORT 8 OF 12 FOR COMPLAINT (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 14654616 ↗
  • InjuryJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022 THERE WAS A REMOVAL OF THE VARIABLE ANGLE (VA) CONDYLE PLATE AND CORRESPONDING SCREWS DUE TO NONUNION/MALUNION. THE SURGEON REMOVED THE PLATE AND SCREWS, REALIGNED THE FEMUR, INSERTED IM NAIL, AND INSERTED BONE GRAFT OBTAINED VIA RADIOIMMUNOASSAY (RIA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/80. THIS IS REPORT 8 OF 12 FOR COMPLAINT (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 14654603 ↗
  • InjuryJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022 THERE WAS A REMOVAL OF THE VARIABLE ANGLE (VA) CONDYLE PLATE AND CORRESPONDING SCREWS DUE TO NONUNION/MALUNION. THE SURGEON REMOVED THE PLATE AND SCREWS, REALIGNED THE FEMUR, INSERTED IM NAIL, AND INSERTED BONE GRAFT OBTAINED VIA RADIOIMMUNOASSAY (RIA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/80. THIS IS REPORT 3 OF 12 FOR COMPLAINT (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 14654211 ↗
  • MalfunctionJun 7, 2022Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2022, 5.0 VA LOCKING SCREW WOULD NOT THREAD INTO THE PLATE IN TWO OF THE DISTAL INTERLOCKING HOLES. THE CORRECT DRILL GUIDE AND DRILL BIT WERE USED TO PREPARE FOR INSERTION BUT SCREWS STRIPPED OUT IN BOTH HOLES. ANOTHER PLATE WAS PLACED WITH SAME SCREWS WITH NO ISSUE. THERE WAS NO REPORTED SURGICAL DELAY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) 5.0MM VARIABLE ANGLE LOCKNG SCREW/SLF-TPNG/STRDRV/80MM. THIS IS REPORT 3 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14622307 ↗
  • MalfunctionJun 7, 2022Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2022, 5.0 VA LOCKING SCREW WOULD NOT THREAD INTO THE PLATE IN TWO OF THE DISTAL INTERLOCKING HOLES. THE CORRECT DRILL GUIDE AND DRILL BIT WERE USED TO PREPARE FOR INSERTION BUT SCREWS STRIPPED OUT IN BOTH HOLES. ANOTHER PLATE WAS PLACED WITH SAME SCREWS WITH NO ISSUE. THERE WAS NO REPORTED SURGICAL DELAY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) 5.0MM VARIABLE ANGLE LOCKNG SCREW/SLF-TPNG/STRDRV/80MM. THIS IS REPORT 2 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14622246 ↗
  • InjuryApr 14, 2022Migration; Migration

    IT WAS REPORTED THAT DISTAL FEMUR OPEN REDUCTION INTERNAL FIXATION REVISION OCCURRED ON (B)(6) 2022 DUE TO MULTIPLE 5.0MM VA LOCKING SCREWS BACKING OUT OF PLATE. IT IS UNKNOWN WHEN THE ORIGINAL SURGERY OCCURRED. DURING THE REVISION THE SURGEON ATTEMPTED TO RE-TIGHTEN THE SAME SCREW BACK INTO THE PLATE WITH A SCREWDRIVER AND APPROPRIATE 6NM TOQUE LIMITING T HANDLE. THE RE-TIGHTENING OF THE SCREWS DID NOT WORK AS THE SCREW HEADS WERE NOT LOCKING INTO THE PLATE. UPON REMOVAL AND INSPECTION OF THE ORIGINAL 12H VARIABLE ANGLE CURVED CONDYLAR PLATE, IT WAS NOTED BY SURGEON THAT THE SCREW HOLES OF THE PLATE AND THE LOCKING THREADS ON THE 5.0MM VARIABLE ANGLE LOCKING SCREWS WERE COMPROMISED. HE HAD TO PLACE A NEW PLATE AND SCREWS IN THE PATIENT AS A CORRECTION FOR THE ORIGINAL SURGERY. NO SURGERY DELAYED AND THE PROCEDURE SUCCESSFULLY COMPLETED. THIS IS REPORT 6 OF 6 FOR (B)(4). THIS REPORT IS FOR A 5.0 VA LOCKING SCREW.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 14112237 ↗
  • InjuryMar 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER QUALITY UNIT RECEIVED A MAUDE REPORT FORWARDED FROM DEPARTMENT OF HUMAN SERVICES; THIS REPORT IS AGAINST USER FACILITY REPORT MW5074624. ONLY ADDITIONAL AND/OR CORRECTED INFORMATION WILL BE CONTAINED IN THIS REPORT. A COPY OF THE MAUDE EVENT REPORT IS ATTACHED. IT WAS REPORTED THAT ON AN UNKNOWN DATE APPROXIMATELY THREE YEARS AGO, PATIENT WAS IMPLANTED WITH ONE (1) UNKNOWN TOTAL KNEE ARTHROPLASTY (TKA) CONSTRUCT (MANUFACTURER UNKNOWN). ON (B)(6) 2016, PATIENT UNDERWENT IMPLANT PROCEDURE TO TREAT LOSS STRESS FEMURE FRACTURE. PATIENT WAS IMPLANTED WITH ONE 4.5 MM (1) VARIABLE ANGLE (VA) CURVED CONDYLAR PLATE, EIGHT (8) 5.0 MM VA LOCKING SCREWS, ONE (1) 5.0 MM CANNULATED VA LOCKING SCREW, ONE (1) 4.5 MM CORTEX SCREW. SUBSEQUENTLY, ON AN UNKNOWN DATE, JUST FIFTY-ONE (51) DAYS AFTER THE PATIENT BEING RELEASED TO A FULL WEIGHT BEARING STATUS, THE PLATE BROKE AND THE PATIENT DEVELOPED A NONUNION AND THE BONE HAD RE FRACTURED. ON (B)(6) 2017, PATIENT UNDERWENT REVISION SURGERY. THE SURGEON WENT INTO SURGERY WITH THE PLAN TO NOT REMOVE THE IMPLANTS, BUT SUBSEQUENTLY, ENDED UP REMOVING ALL IMPLANTS. THE SURGEON CUT OUT THE TISSUE SURROUNDING THE PLATE AND THE PLATE WAS STILL STUCK.

    Patient problems: No Code Available; Nonunion/Delayed-union Bone Fracture; Unspecified Infection

    MAUDE report 7327942 ↗
  • InjuryMar 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER QUALITY UNIT RECEIVED A MAUDE REPORT FORWARDED FROM DEPARTMENT OF HUMAN SERVICES; THIS REPORT IS AGAINST USER FACILITY REPORT MW5074624. ONLY ADDITIONAL AND/OR CORRECTED INFORMATION WILL BE CONTAINED IN THIS REPORT. A COPY OF THE MAUDE EVENT REPORT IS ATTACHED. IT WAS REPORTED THAT ON AN UNKNOWN DATE APPROXIMATELY THREE YEARS AGO, PATIENT WAS IMPLANTED WITH ONE (1) UNKNOWN TOTAL KNEE ARTHROPLASTY (TKA) CONSTRUCT (MANUFACTURER UNKNOWN). ON AUG 26, 2016, PATIENT UNDERWENT IMPLANT PROCEDURE TO TREAT LOSS STRESS FEMUR FRACTURE. PATIENT WAS IMPLANTED WITH ONE 4.5 MM (1) VARIABLE ANGLE (VA) CURVED CONDYLAR PLATE, EIGHT (8) 5.0 MM VA LOCKING SCREWS, ONE (1) 5.0 MM CANNULATED VA LOCKING SCREW, ONE (1) 4.5 MM CORTEX SCREW. SUBSEQUENTLY, ON AN UNKNOWN DATE, JUST FIFTY-ONE (51) DAYS AFTER THE PATIENT BEING RELEASED TO A FULL WEIGHT BEARING STATUS, THE PLATE BROKE AND THE PATIENT DEVELOPED A NONUNION AND THE BONE HAD RE FRACTURED. ON JAN 08, 2017, PATIENT UNDERWENT REVISION SURGERY. THE SURGEON WENT INTO SURGERY WITH THE PLAN TO NOT REMOVE THE IMPLANTS, BUT SUBSEQUENTLY, ENDED UP REMOVING ALL IMPLANTS. THE SURGEON CUT OUT THE TISSUE SURROUNDING THE PLATE AND THE PLATE WAS STILL STUCK.

    Patient problems: No Code Available; Nonunion/Delayed-union Bone Fracture; Unspecified Infection

    MAUDE report 7327934 ↗
  • InjuryMar 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER QUALITY UNIT RECEIVED A MAUDE REPORT FORWARDED FROM DEPARTMENT OF HUMAN SERVICES; THIS REPORT IS AGAINST USER FACILITY REPORT MW5074624. ONLY ADDITIONAL AND/OR CORRECTED INFORMATION WILL BE CONTAINED IN THIS REPORT. A COPY OF THE MAUDE EVENT REPORT IS ATTACHED. IT WAS REPORTED THAT ON AN UNKNOWN DATE APPROXIMATELY THREE YEARS AGO, PATIENT WAS IMPLANTED WITH ONE (1) UNKNOWN TOTAL KNEE ARTHROPLASTY (TKA) CONSTRUCT (MANUFACTURER UNKNOWN). ON (B)(6) 2016, PATIENT UNDERWENT IMPLANT PROCEDURE TO TREAT LOSS STRESS FEMURE FRACTURE. PATIENT WAS IMPLANTED WITH ONE 4.5 MM (1) VARIABLE ANGLE (VA) CURVED CONDYLAR PLATE, EIGHT (8) 5.0 MM VA LOCKING SCREWS, ONE (1) 5.0 MM CANNULATED VA LOCKING SCREW, ONE (1) 4.5 MM CORTEX SCREW. SUBSEQUENTLY, ON AN UNKNOWN DATE, JUST FIFTY-ONE (51) DAYS AFTER THE PATIENT BEING RELEASED TO A FULL WEIGHT BEARING STATUS, THE PLATE BROKE AND THE PATIENT DEVELOPED A NONUNION AND THE BONE HAD RE FRACTURED. ON (B)(6) 2017, PATIENT UNDERWENT REVISION SURGERY. THE SURGEON WENT INTO SURGERY WITH THE PLAN TO NOT REMOVE THE IMPLANTS, BUT SUBSEQUENTLY, ENDED UP REMOVING ALL IMPLANTS. THE SURGEON CUT OUT THE TISSUE SURROUNDING THE PLATE AND THE PLATE WAS STILL STUCK.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 7327508 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083025, K110354. One submission often covers many models, so check the product and lot details in each record.

  • Z-1125-2014 · Feb 5, 2014 · Terminated
    Synthes 5.0mm Variable Locking Screw This device is part of the Variable Angle Locking Compression Plate family and this technology included variable angle Combi holes in the plat
  • Z-0401-2014 · May 29, 2013 · Terminated
    Synthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER Synthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER is used for buttressing multi fragmentary diatal femur fra

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page