NA 03.010.079Recalls & reported problems
Synthes GmbH · DRILL STOP FOR 4.5MM/6.5MM STEPPED DRILL BIT · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982067494), received between Oct 31, 2018 and Jun 19, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility4
- Break2
- Crack1
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2025Break; Crack; Break; Crack
IT WAS REPORTED THAT INSTRUMENTS WERE UNABLE TO BE STERILIZES/ASSEMBLES. CHIPPING AND CRACKING. UNABLE TO USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22289769 ↗MalfunctionMar 29, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN NEW (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE STEPPED DRILL BIT AND DRILL STOP WERE SLIPPING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONSEQUENCE. THERE IS NO FURTHER INFORMATION AVAILABLE. CONCOMITANT DEVICES REPORTED: UNKNOWN HIP SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN), UNKNOWN LFN NAIL (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) DRILL STOP FOR 4.5MM/6.5MM STEPPED DRILL BIT. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11582994 ↗InjuryFeb 4, 2021Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT ON (B)(6) 2021, THE DRILL STOP FOR 4.5MM /6.5MM STEPPED DRILL BIT IN THE FEMORAL RECON NAIL TRAY BROKE OFF IN THE FEMORAL HEAD OF THE PATIENT INTRAOPERATIVELY. TRIED TO RETRIEVE THE BROKEN BIT WITH THE PITUITARY. THERE WERE A FEW EXTRA X-RAYS TAKEN WHEN ATTEMPTING TO RETRIEVE THE BROKEN PIECE. THE BROKEN PIECE WAS LEFT IN THE FEMORAL HEAD. THERE WERE 10 MINUTES OF SURGICAL DELAY. THE PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) DRILL STOP FOR 4.5MM/6.5MM STEPPED DRILL BIT. THIS IS REPORT 1 OF 1 (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 11281682 ↗MalfunctionJan 28, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT FOR AN INTRAMEDULLARY NAILING OF FEMUR SURGERY. DURING THE SURGERY, THE DRILL STOP WAS SLIDE TWICE IN THE SAME PROCEDURE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT DELAY REPORTED. THERE WERE NO PATIENT CONSEQUENCES. CONCOMITANT DEVICE REPORTED: DRILL BIT (PART#: 03.010.078, LOT#: UNKNOWN, QUANTITY: 1), UNKNOWN SCREW (PART#: 03.010.078, LOT#: UNKNOWN, QUANTITY: 2). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT INVOLVES ONE (1)DRILL STOP FOR 4.5MM/6.5MM STEPPED DRILL BIT. THIS IS REPORT 1 OF 1 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11247381 ↗MalfunctionJan 29, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT OPEN REDUCTION AND INTERNAL FIXATION (ORIF) OF RIGHT FEMUR. DURING THE PROCEDURE, THE DRILL STOP AND THE STEPPED CANNULATED DRILL BIT WERE NOT LOCKING INTO PLACE. THE PROCEDURE SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. PATIENT STATUS IS UNKNOWN. THIS REPORT IS FOR A DRILL STOP FOR 4.5MM/6.5MM STEPPED DRILL BIT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8286857 ↗MalfunctionOct 31, 2018Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A FEMORAL RECON NAIL PROCEDURE ON (B)(6) 2018, THE DRILL STOP SLIPPED ON DRILL BIT WHILE DRILLING FOR UNKNOWN RECON SCREWS. THE DRILL STOP WOULD NO LONGER LOCKED INTO THE DRILL BIT. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO PATIENT CONSEQUENCE REPORTED. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICES REPORTED: UNKNOWN RECON SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY # UNKNOWN), UNKNOWN FEMORAL RECON NAIL (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) DRILL STOP FOR 4.5MM/6.5MM STEPPED DRILL BIT. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8026212 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page