NA 03.037.140Recalls & reported problems
Synthes GmbH · LOCKING DEVICE FOR SCREW ONLY AIMING ARM · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982070593), received between Jun 25, 2019 and Feb 17, 2021 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break4
- Device-Device Incompatibility2
- Material Deformation1
Patient problems
- No Patient Involvement3
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 17, 2021Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT ON (B)(6) 2021, THE SCREW INSERTER WAS DEFORMED. UPON INSERTION OF THE LAG SCREW USING THE SCREW INSERTER THE INSERTION T HANDLE DEVICE BECAME UNATTACHED FROM THE LAG SCREW AND PREVENTED FURTHER INSERTION. THE T HANDLE INSERTER AND COUPLING SCREW WERE INSPECTED AND SEEMED TO BE DISFIGURED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE OUTCOME WAS UNKNOWN. NO PATIENT CONSEQUENCES. CONCOMITANT DEVICE REPORTED: HELICAL BLADE/SCREW COUPLING SCREW (PART: 03.037.026; LOT:3L89171; QUANTITY: 1). SCREW INSERTER (PART NUMBER 03.037.025, LOT UNKNOWN, QUANTITY 1). SCREWS: LAG (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES ONE (1) LOCKING DEVICE FOR SCREW ONLY AIMING ARM. THIS IS REPORT 2 OF 4 FOR (B)(4)..
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11337326 ↗MalfunctionAug 13, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, ONE LOCKING DEVICE FOR SCREW ONLY AIMING ARM WAS BROKEN DUE TO WEAR AND TEAR. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) LOCKING DEVICE FOR SCREW ONLY AIMING ARM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10404381 ↗MalfunctionJun 24, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, A DRIVING CAP AND TWO (2) LOCKING DEVICE FOR SCREW ONLY AIMING ARM WERE FOUND BROKEN AT THE STERILE PROCESSING DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) LOCKING DEVICE FOR SCREW ONLY AIMING ARM. THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10192312 ↗MalfunctionJun 24, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, A DRIVING CAP AND TWO (2) LOCKING DEVICE FOR SCREW ONLY AIMING ARM WERE FOUND BROKEN AT THE STERILE PROCESSING DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) LOCKING DEVICE FOR SCREW ONLY AIMING ARM. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10192304 ↗MalfunctionNov 21, 2019Break; Break
UPDATED CONCOMITANT DEVICE REPORTED: UNKNOWN HANDLE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9355407 ↗InjuryJun 25, 2019Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN (B)(6) IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT A FEMORAL NAILING SURGERY WITH THE RIGHT TROCHANTERIC FEMORAL NAILING SYSTEM ADVANCED (TFNA) INTRAMEDULLARY NAIL FOR A SUB-TROCHANTERIC FEMORAL FRACTURE ON (B)(6) 2019. THE FRACTURE WAS DIFFICULT TO REDUCE. BEFORE IMPLANTATION, THE NURSING STAFF AND THE SURGEON TESTED TO INSERT THE LAG SCREW INTO THE NAIL AND HAD NO ISSUES WITH THE JIG. DURING IMPLANTATION, THE NAIL AND LAG SCREW WAS IMPLANTED WITH SOME DIFFICULTY IN THE INSERTION OF THE NAIL. DISTAL CROSS BOLTS WERE THEN IMPLANTED AND THE JIG WAS REMOVED. IN THE FINAL SHOTS, THE SCREW APPEARED TO HAVE MISSED THE NAIL. THE SURGEON REMOVED THE LAG SCREW AND THEN WITH THE ASSISTANCE OF ANOTHER SURGEON REINSERTED THE LAG SCREW AFTER CONNECTING THE JIG INTO THE NAIL IN-SITU. X-RAYS WERE THEN TAKEN AND WAS CONFIRMED THAT THE LAG SCREW WAS SUCCESSFULLY RE-POSITIONED TO THE NAIL. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 45 MINUTES. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICES REPORTED: UNKNOWN TFNA NAIL (PART # 04.037.028S, LOT # H309035, QUANTITY# 1). LOCKING SCREW (PART# 04.005.530S, LOT# 2L43586, QU
Patient problems: No Code Available
MAUDE report 8730621 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page