Skip to content

NA 03.211.007Recalls & reported problems

Synthes GmbH · MESH PLATE CUTTERS · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10886982076052), received between Aug 3, 2018 and May 2, 2024 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Break” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break12
  • Crack1
  • Device-Device Incompatibility1
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Patient Involvement2
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 2, 2024Break; Crack; Device-Device Incompatibility; Break; Crack; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DEVICES WERE UNABLE TO STERILIZE/ASSEMBLE. IT WAS ALSO REPORTED THAT THE DEVICES HAD CHIPPING/CRACKING AND UNABLE TO BE USE. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) MESH PLATE CUTTERS. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19234952 ↗
  • MalfunctionOct 6, 2023Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2023 THE INSTRUMENT WAS BROKEN. THE PLATE CUTTER WAS FOUND BROKEN IN SET .THE PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT STATUS/ OUTCOME: PATIENT FINE. THIS REPORT IS FOR ONE (1) MESH PLATE CUTTERS THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17887718 ↗
  • MalfunctionSep 22, 2023Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2023, UPON INSPECTION OF THE DEVICE, IT WAS FOUND TO HAVE A BROKEN TOP HINGE OF THE PLATE CUTTER. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE MESH PLATE CUTTERS. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17799780 ↗
  • MalfunctionOct 14, 2022Break; Break

    IT WAS REPORTED ON (B)(6)2022 ,SURGEON USING MESH PLATE CUTTER. ONE OF THE POSTS ON CUTTER BROKE OFF. THE SURGERY WAS THEN COMPLETED SUCCESSFULLY. THERE WAS NO SURGICAL DELAY. PATIENT OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) MESH PLATE CUTTERS THIS IS REPORT 1 OF 1 FOR COMPLAINT PC-(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15607415 ↗
  • MalfunctionOct 4, 2022Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE ONE OF THE ACCOUNTS INFORMED US THAT THE MESH PLATE CUTTER FROM THEIR MIDFOOT/FOREFOOT SET WAS BROKEN. THE EVENT OCCURRED POST-OP AND THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE DEVICE. THIS REPORT IS FOR ONE MESH PLATE CUTTERS. THIS REPORT IS 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15540633 ↗
  • MalfunctionMar 23, 2022Break; Break

    IT WAS REPORTED THAT A MESH PLATE CUTTER WAS DISMANTLED. PATIENT AND PROCEDURE INVOLVEMENT ARE UNKNOWN. NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) MESH PLATE CUTTERS THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13865235 ↗
  • MalfunctionFeb 28, 2022Break; Break

    THE MESH PLATE CUTTERS (03.211.007) BROKE AFTER SINGLE USE ON THE 2.4 MM / 2.7 MM VA-LCP MESH PLATE (02.211.224) DURING A FOOT SURGERY ON (B)(6) 2022. THE CUSTOMER JUST PURCHASED THE MESH PLATE CUTTERS AND IS FRUSTRATED THEY BROKE AFTER SINGLE USE. THE NUB THAT IS INSERTED INTO THE LOCKING HOLE SNAPPED OFF WHEN APPLYING TENSION. THIS REPORT IS FOR ONE (1) MESH PLATE CUTTERS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13618880 ↗
  • MalfunctionFeb 16, 2022Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PRODUCTS BROKE DUE TO USAGE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) MESH PLATE CUTTERS. THIS IS REPORT 1 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13543674 ↗
  • InjurySep 16, 2021Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2021, DURING A PROCEDURE PLATE CUTTER WAS BROKE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. FRAGMENTS WERE GENERATED AND NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICE REPORTED: UNK - PLATES: MESH IMPLANT (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS REPORT IS FOR ONE (1) MESH PLATE CUTTERS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 12483035 ↗
  • MalfunctionFeb 24, 2020Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2020 THAT THE METAL PORTION OF THE MESH PLATE CUTTERS HAS BROKEN DURING THE STERILE PROCESSING AFTER A CASE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE MESH PLATE CUTTERS. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 9744877 ↗
  • MalfunctionApr 10, 2019Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY ON (B)(6) 2019, THE MESH PLATE CUTTER WAS DISCOVERED BROKEN FROM THE VARIABLE ANGLE LOCKING COMPRESSION PLATE FOREFOOT MIDFOOT SYSTEM SET. REPORTEDLY, THIS DID NOT AFFECT THE SURGERY AND THERE WAS NO PATIENT IMPACT. THIS REPORT IS FOR ONE (1) MESH PLATE CUTTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8500623 ↗
  • MalfunctionAug 3, 2018Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING INSPECTION IN THE STERILE PROCESSING DEPARTMENT (SPD), THE TIP OF THE STARDRIVE SCREWDRIVER SHAFT WAS OBSERVED TO BE TWISTED AND ALSO A MESH PLATE CUTTERS WAS DISCOVERED TO BE BROKEN WHILE THE SALES CONSULTANT WAS SETTING UP THE TRAY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) MESH PLATE CUTTERS. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 7748953 ↗

Other INSTRUMENT, CUTTING, ORTHOPEDIC devices

All INSTRUMENT, CUTTING, ORTHOPEDIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page