EXPert Nail 04.003.001Recalls & reported problems
Synthes GmbH · 12MM TI END CAP T40 STARDRIVE 5MM EXT FOR FEMORAL NAILS-EX · Class 2 · Implant
All models of EXPert Nail (151)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982081315), received between Dec 12, 2016 and Jun 1, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K033618, K040336, K172157); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Device-Device Incompatibility1
- Fitting Problem1
Patient problems
- No Code Available2
- Failure of Implant1
- Not Applicable1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 1, 2020Device-Device Incompatibility; Device-Device Incompatibility
CONCOMITANT DEVICES REPORTED: NAILS: RAFN (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1), NAIL INSERTION HANDLES (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1), INSERTION INSTRUMENTS: CONNECTING/COUPLING SCREW: TRAUMA (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1), EXPERT END CAP Ø12 CANN EXTENS. 10 F/LFN (PART NUMBER 04.003.002, LOT 4L91497, QUANTITY 1), EXPERT END CAP Ø12 CANN EXTENS. 15 F/LFN (PART NUMBER 04.003.003, LOT 22P9591, QUANTITY 1).
Patient problems: Failure of Implant; No Code Available
MAUDE report 10105788 ↗InjuryApr 23, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH ONE (1) RECON-NAIL AND TWO (2) RECON-SCREWS DURING A HIP SURGERY ON (B)(6) 2018. IT WAS NOTICED WITH THE HELP OF POST-OPERATIVE X-RAYS, THAT THE RECON SCREWS SEEMED TOO LONG, SO THE SURGEON BROUGHT THE PATIENT BACK TO THE SURGERY ON (B)(6) 2018 TO BACK THE SCREWS OUT A COUPLE OF TURNS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY DELAY. THE PATIENT STATUS WAS REPORTED TO BE FINE AT THE END OF THE PROCEDURE. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS 1 OF 2 FOR (B)(4).
Patient problems: Not Applicable
MAUDE report 7452551 ↗InjuryDec 12, 2016Fitting Problem; Fitting Problem
IT WAS REPORTED THAT DURING STAGE TWO REMOVAL OF EXTERNAL FIXATOR AND IMPLANTATION OF RETROGRADE FEMORAL NAIL PROCEDURE ON (B)(6) 2016, SURGEON SELECTED 5 MM END CAP AND IT WOULD NOT ENGAGE / DID NOT THREAD INTO THE NAIL. THEN SURGEON SELECTED 0 MM END CAP HOWEVER SECOND 5 MM END CAP WAS TESTED ON THE BACK TABLE WITH THE TRIAL RETROGRADE NAIL (EX RAFN NAIL) AND IT WAS ENGAGED AS PLANNED AND THEREFORE, 0 MM END CAP WAS REMOVED AND SECOND 5 MM END CAP WAS INSERTED WITHOUT ANY ISSUE. THERE WAS 5 MINUTES DELAY IN PROCEDURE AND THE SURGERY WAS COMPLETED SUCCESSFULLY AND THE PATIENT OUTCOME WAS REPORTED TO BE AS PLANNED. THE FIRST MALFUNCTIONED 5 MM END CAP WAS TESTED ON THE BACK TABLE WITH TRIAL EX RAFN NAIL AND IT WOULD NOT ENGAGE THE 5 MM END CAP WAS MANUALLY REMOVED WITH NEEDLE HOLDER. INTRA OPERATIVE UNPLANNED X-RAYS TAKEN TO RESOLVE END CAP PROBLEM AND WERE NOT AVAILABLE FOR REVIEW. THIS COMPLAINT INVOLVES ONE DEVICE. CONCOMITANT DEVICES REPORTED: RETROGRADE FEMORAL NAIL (PART # 04.013.352S, LOT # 7316088, QUANTITY # 1), TRIAL EX RAFN NAIL (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY #1). THIS REPORT IS 1 OF 1 FOR (B)(4)
Patient problems: No Code Available
MAUDE report 6167349 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033618, K040336, K172157. One submission often covers many models, so check the product and lot details in each record.
- Z-1915-2015 · May 27, 2015 · TerminatedDePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm, 75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part of the Expert Lateral Femur Nail System and indicated to stabilize Sub
- Z-2058-2013 · Nov 2, 2012 · TerminatedSynthes Trauma Nail System. The devices are indication for bone fixation.
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page