NA 04.005.522SRecalls & reported problems
Synthes GmbH · 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 32MM F/IM NAIL-STER · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982085917), received between Sep 21, 2018 and Jan 23, 2026 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Migration2
- Device-Device Incompatibility1
- Difficult to Advance1
- Failure to Cut1
Patient problems
- Failure of Implant2
- Insufficient Information2
- No Code Available2
- Pain2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Device-Device Incompatibility
IT WAS NOT POSSIBLE TO UNLOCK THE SYSTEM TO REMOVE IT.
Patient problems: Insufficient Information
MAUDE report 24168165 ↗InjuryNov 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT THE SURGERY FOR CERVICAL BASE FRACTURE WITH THE PRODUCTS IN QUESTION. THE PATIENT HAD BEEN ABLE TO UNDERGO AN EXTRAMEDULLARY REPAIR, BUT WAS SHIFTED TO INTRAMEDULLARY AFTER SURGERY, AND IS NOW ON THE VERGE OF AN ENDOPROSTHETIC CUTOUT. AN ARTIFICIAL HEAD REPLACEMENT WAS PERFORMED ON (B)(6) 2022. THE CAUSE IS UNKNOWN BECAUSE NO POSTOPERATIVE CT WAS TAKEN. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 32MM F/IM NAIL-STER. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 15715797 ↗InjuryAug 26, 2022Migration; Migration
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT THE ORIF SURGERY FOR TROCHANTERIC FRACTURE OF FEMUR WITH THE TFNA LONG IMPLANTS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. A REVISION SURGERY WAS PLANNED FOR (B)(6) 2022, THE TFNA LONG IMPLANTS WILL BE REMOVED AND REPLACED WITH BHA. THE DETAILS OF THE DEFECT ARE UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 32MM F/IM NAIL-STER. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 15302527 ↗InjuryAug 26, 2022Migration; Migration
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT THE ORIF SURGERY FOR TROCHANTERIC FRACTURE OF FEMUR WITH THE TFNA LONG IMPLANTS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. A REVISION SURGERY WAS PLANNED FOR (B)(6) 2022, THE TFNA LONG IMPLANTS WILL BE REMOVED AND REPLACED WITH BHA. THE DETAILS OF THE DEFECT ARE UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 32MM F/IM NAIL-STER. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 15302497 ↗MalfunctionNov 5, 2018Failure to Cut; Difficult to Advance; Failure to Cut; Difficult to Advance
IT WAS REPORTED THAT DURING A TIBIAL NAIL CASE PROCEDURE ON (B)(6) 2018, THE FULLY THREADED NON-SELF TAPPING SCREW DID NOT GO INTO THE NAIL. THE SURGEON TRIED SEVERAL TIMES TO GET THE TITANIUM (TI) LOCKING SCREW TO GO IN, THEN PULLED IT OUT AND NOTICED THAT IT WAS NOT A SELF-TAPPING SCREW WHICH THEY NORMALLY USE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SURGICAL DELAY OF 5 MINUTES. PATIENT STATUS WAS GOOD. THIS REPORT IS FOR A 5.0MM TI LOCKING SCREW 32MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8039372 ↗InjurySep 21, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS FURTHER REPORTED THAT DURING THE REMOVAL PROCEDURE, SCREWS WERE REMOVED AND NO ADDITIONAL DEVICES WERE IMPLANTED.
Patient problems: No Code Available; Pain
MAUDE report 7898225 ↗InjurySep 21, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS FURTHER REPORTED THAT DURING THE REMOVAL PROCEDURE, SCREWS WERE REMOVED AND NO ADDITIONAL DEVICES WERE IMPLANTED.
Patient problems: No Code Available; Pain
MAUDE report 7898184 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.