NA 04.005.538SRecalls & reported problems
Synthes GmbH · 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 48MM F/IM NAIL-STER · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982086075), received between Apr 18, 2018 and Aug 3, 2020 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break2
- No Apparent Adverse Event2
Patient problems
- Nonunion/Delayed-union Bone Fracture4
- No Code Available2
- No Consequences Or Impact To Patient2
- Impaired Healing1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 3, 2020Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
THE INITIAL COMPLAINT WAS REVIEWED AND FOUND NOT REPORTABLE. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE SCREWS WERE NOT BROKEN AS PREVIOUSLY REPORTED. THERE IS NO ALLEGATION OF A MALFUNCTION AGAINST THE SCREWS.
Patient problems: No Code Available; No Consequences Or Impact To Patient; Nonunion/Delayed-union Bone Fracture
MAUDE report 10359044 ↗InjuryAug 3, 2020Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
THE INITIAL COMPLAINT WAS REVIEWED AND FOUND NOT REPORTABLE. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE SCREWS WERE NOT BROKEN AS PREVIOUSLY REPORTED. THERE IS NO ALLEGATION OF A MALFUNCTION AGAINST THE SCREWS.
Patient problems: No Code Available; No Consequences Or Impact To Patient; Nonunion/Delayed-union Bone Fracture
MAUDE report 10358993 ↗InjuryJul 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED IN (B)(6) 2017, THE PATIENT UNDERWENT THE SURGERY BY USING A COMPETITORS FEMORAL IMPLANT SYSTEM IN ORDER TO PREVENT BONE FRACTURES DUE TO THE BREAST CANCER BEING SPREAD TO THE FEMUR. ON (B)(6) 2020, A REVISION SURGERY WAS PERFORMED TO REPLACE THE FEMORAL IMPLANTS DUE TO DEVICE BREAKAGE WITH SYNTHES TROCHANTERIC FIXATION NAIL ANTIROTATION (TFNA) LONG IMPLANTS. ON (B)(6) 2020, A REVISION SURGERY WAS PERFORMED DUE TO THE BREAKAGE OF THE TFNA LONG IMPLANTS. THE PATIENT WAS REVISED TO ANOTHER SET OF IMPLANTS. THERE WAS A ONE (1) HOUR SURGICAL DELAY DUE TO EXTENDED TIME IN REMOVING THE BROKEN DEVICE. IT WAS FURTHER REPORTED THE PATIENT ALSO HAD A NON-UNION WHICH MAY HAVE CONTRIBUTED TO THE DEVICE BREAKAGE. THIS IS REPORT 3 OF 5 FOR (B)(4).
Patient problems: Impaired Healing; Nonunion/Delayed-union Bone Fracture
MAUDE report 10343611 ↗InjuryApr 18, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT OPEN REDUCTION INTERNAL FIXATION (ORIF) OF A PROXIMAL FEMUR FRACTURE USING A TROCHANTERIC FIXATION NAIL - ADVANCE (TFNA) ON (B)(6) 2017. POSTOPERATIVELY, A NON-UNION WAS IDENTIFIED ON A ROUTINE FOLLOW-UP X-RAY. PATIENT WAS ASYMPTOMATIC (NO COMPLAINTS OF PAIN OR DISCOMFORT). THE PATIENT DID NOT HEAL DUE TO LATERAL CORTEX PORTION OF THE FEMUR BONE WAS MISSING BECAUSE OF THE COMPLEX FRACTURE PATTERN. THE PATIENT UNDERWENT A REVISION ON (B)(6) 2018 AND RECEIVED A RIGHT TOTAL HIP REPLACEMENT. ALL THE REMOVED IMPLANTS WERE INTACT. THERE WERE NO ISSUES DURING THE REVISION AND NO SURGICAL TIME DELAY. THE PATIENT STATUS WAS STABLE. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 7440858 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page