VA-LCP 04.115.850SRecalls & reported problems
Synthes GmbH · 2.4MM TI VA-LCP VOLAR RIM DSTL RAD PL/7H HD/5H SHFT/RT-STER · Class 2 · Implant
All models of VA-LCP (256)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982105172), received between Jun 27, 2025 and Jun 27, 2025 — classified as 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Adhesion(s)1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT UNDERWENT AN UNKNOWN SURGERY FOR RADIUS WITH THE PLATE AND THE SCREWS. AS THE BONE ADHESION RESULTED, THE REMOVAL SURGERY WAS PERFORMED ON (B)(6) 2025. AT THAT TIME, IT WAS DISCOVERED THAT THE MOST DISTAL MOST ULNAR SCREW HAD BROKEN AT THE NECK. THE BROKEN SCREW SHAFT WAS SLIGHTLY PROTRUDING, SO IT WAS FILED DOWN WITH A HALO REAMER AND REMOVED WITH AN EXTRACTION BOLT. THE SURGEON CONFIRMED THAT THERE WAS NO RESIDUE BASED ON THE FLUOROSCOPY AND POSTOPERATIVE X-RAYS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30MINS OF DELAY. ON (B)(6) 2025, THE SALES REP LOOKED AT THE X-RAYS AND CT SCANS WITH THE SURGEON, BUT IT IS UNCLEAR WHEN THE SCREW BROKE. NO FURTHER INFORMATION IS AVAILABLE. NOTE: TWO SCREW PART NUMBERS ARE LISTED, BUT IT IS UNCLEAR WHICH ONE BROKE. LOOKING AT THE HISTORY AT THE TIME, ONLY THE ABOVE TWO-PART NUMBERS WERE USED.
Patient problems: Adhesion(s)
MAUDE report 22351177 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page