Skip to content

NA 04.118.530Recalls & reported problems

Synthes GmbH · 2.7MM TI METAPHYSEAL SCR/SLF- TPNG W/T8 STRDRV RECESS/30MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982108944), received between Oct 2, 2023 and Oct 2, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K120070, K120717, K120854); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Entrapment of Device1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2023Break; Entrapment of Device; Break; Entrapment of Device

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6), 2023, THE PATIENT UNDERWENT AN UNKNOWN SURGERY. THE SCREW WAS USED TO PULL THE BONE FRAGMENT NEAR WHEN APPLYING AN OLECRANON 4-HOLE PLATE. THE PATIENT¿S BONE QUALITY WAS GOOD, AND IT TOOK TIME TO INSERT THE SCREW, BUT THE SURGEON SUCCEEDED IN INSERTING THE SCREW FULLY. AFTER THAT, THE HEAD PART OF THE SCREW BROKE OFF WHEN THE SURGEON TIGHTENED THE SCREW AT THE END. THE SURGEON DETERMINED THAT IT WOULD BE DIFFICULT TO REMOVE, AND THE SCREW WAS TO REMAIN IN THE PATIENT¿S BODY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. IT WAS CONFIRMED BY X-RAYS THAT OTHER THAN THE SHAFT OF THE SCREW REMAINED IN THE BONE, NO BROKEN FRAGMENTS REMAINED IN THE PATIENT¿S BODY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 2.7MM TI METAPHYSEAL SCR/SLF- TPNG W/T8 STRDRV RECESS/30MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 17855206 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120070, K120717, K120854. One submission often covers many models, so check the product and lot details in each record.

  • Z-0124-2016 · Sep 28, 2015 · Terminated
    Synthes 2.7MM/3.5MM VA-LCP Medial Distal Humerus Plates 10H/RT/186mm- Sterile and Synthes 2.7MM/3.5MM VA-LCP Extended Medial Distal Humerus Plates 10H/RT/189mm-Sterile The Synthes

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page