NA 04.130.210Recalls & reported problems

Synthes GmbH · 1.5MM TI VA-LCKNG SCR SLF-TPNG WITH T4 STARDRIVE RECESS 10MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982112750), received between Feb 12, 2018 and Dec 12, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Entrapment of Device2
  • Failure To Adhere Or Bond1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • No Code Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2024Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT UNDERWENT AN UNKNOWN SURGERY FOR METACARPAL ON MIDDLE FINGER WITH VA-HAND AND THE PRODUCTS IN QUESTION. IN THE SURGERY, THE HEADS OF THE SCREWS IN QUESTION WERE DAMAGED WHEN THE LAST FASTENING WAS ATTEMPTED. THE DRILL TIP WAS THEN DAMAGED WHEN DRILLING THE NEXT HOLE. THE DRILL TIP WAS ALSO BROKEN IN THE NEXT HOLE. THE SCREWS WITH A BROKEN HEADS AND TWO BROKEN DRILL TIPS HAVE BEEN IMPLANTED IN THE BODY AT THE SURGEON'S DISCRETION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH 30 MINUTES SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 20913917 ↗
  • MalfunctionSep 26, 2023Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT ORIF SURGERY FOR THIRD AND FOURTH METACARPAL FRACTURE WITH SCREWDRIVER SHAFT, DRILL BIT, VA LOCKING STRAIGHT PLATE AND LOCKING SCREWS. WHILE THE SURGEON WAS DRILLING TO FIX THE VA LOCKING STRAIGHT PLATE, THE DRILL BIT BROKE. FORTUNATELY, THE DRILL BIT COULD BE REMOVED FROM THE BODY, AND THE SURGERY PROCEEDED. NEXT, THE SCREWDRIVER SHAFT BROKE WHEN INSERTING THE LOCKING SCREW. BUT THIS ONE WAS ALSO REMOVED FROM THE BODY. THE SURGEON PROCEEDED WITH THE SURGERY USING A SCREWDRIVER OF THE SAME STANDARD. AFTER THAT, WHEN INSERTING ANOTHER LOCKING SCREW, TWO SCREW HEADS BROKE. THE SURGEON REMOVED ONE LOCKING SCREW, BUT THE OTHER LOCKING SCREW WAS INSERTED IN A GOOD POSITION FOR FIXATION, SO THE SURGEON LEFT IT IN PLACE AND REMOVED THE IMPLANT. THE SURGEON THEN CHANGED TO 2.0MM PLATE AND SCREW AND FIXED IT, COMPLETING THE SURGERY. ALL THAT REMAINS IN THE PATIENT¿S BODY IS A SINGLE SCREW WITH A BROKEN SCREW HEAD. THE SURGEON THOUGHT THE CAUSE OF THE FRACTURE WAS THAT THE PATIENT WAS 19 YEARS OLD AND HAD GOOD BONE QUALITY. THE SURGERY WAS COMPLETED SUCCESSFULLY MORE TH

    Patient problems: Insufficient Information

    MAUDE report 17820410 ↗
  • MalfunctionSep 26, 2023Break; Entrapment of Device; Break; Entrapment of Device

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6), 2023, THE PATIENT UNDERWENT ORIF SURGERY FOR THIRD AND FOURTH METACARPAL FRACTURE WITH SCREWDRIVER SHAFT, DRILL BIT, VA LOCKING STRAIGHT PLATE AND LOCKING SCREWS. WHILE THE SURGEON WAS DRILLING TO FIX THE VA LOCKING STRAIGHT PLATE, THE DRILL BIT BROKE. FORTUNATELY, THE DRILL BIT COULD BE REMOVED FROM THE BODY, AND THE SURGERY PROCEEDED. NEXT, THE SCREWDRIVER SHAFT BROKE WHEN INSERTING THE LOCKING SCREW. BUT THIS ONE WAS ALSO REMOVED FROM THE BODY. THE SURGEON PROCEEDED WITH THE SURGERY USING A SCREWDRIVER OF THE SAME STANDARD. AFTER THAT, WHEN INSERTING ANOTHER LOCKING SCREW, TWO SCREW HEADS BROKE. THE SURGEON REMOVED ONE LOCKING SCREW, BUT THE OTHER LOCKING SCREW WAS INSERTED IN A GOOD POSITION FOR FIXATION, SO THE SURGEON LEFT IT IN PLACE AND REMOVED THE IMPLANT. THE SURGEON THEN CHANGED TO 2.0MM PLATE AND SCREW AND FIXED IT, COMPLETING THE SURGERY. ALL THAT REMAINS IN THE PATIENT¿S BODY IS A SINGLE SCREW WITH A BROKEN SCREW HEAD. THE SURGEON THOUGHT THE CAUSE OF THE FRACTURE WAS THAT THE PATIENT WAS 19 YEARS OLD AND HAD GOOD BONE QUALITY. THE SURGERY WAS COMPLETED SUCCESSFULLY MORE T

    Patient problems: Insufficient Information

    MAUDE report 17819753 ↗
  • MalfunctionNov 7, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN SURGICAL PROCEDURE, THE TWO (2) SCREW HEADS BROKE WHILE INSERTING INTO THE BONE AFTER DRILLING. THERE WERE NO FRAGMENTS GENERATED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) 1.5MM TI VA-LCKNG SCR SLF-TPNG WITH T4 STARDRIVE RECESS 10MM. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12760196 ↗
  • MalfunctionNov 7, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN SURGICAL PROCEDURE, THE TWO (2) SCREW HEADS BROKE WHILE INSERTING INTO THE BONE AFTER DRILLING. THERE WERE NO FRAGMENTS GENERATED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) 1.5MM TI VA-LCKNG SCR SLF-TPNG WITH T4 STARDRIVE RECESS 10MM. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12760169 ↗
  • InjuryFeb 12, 2018Failure To Adhere Or Bond; Failure To Adhere Or Bond

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DEVICE WAS USED IN THE SURGERY FOR THE FRACTURE OF THE LEFT SECOND METACARPAL BONE ON (B)(6) 2017. THE VARIABLE ANGLE (VA) LOCKING Y-SHAPE PLATE WAS USED FOR THE FIXATION. AFTER THE FIXATION WITH THE PROXIMAL THREE SCREWS AND DISTAL TWO SCREWS BY USING THE VA SLEEVE, RE-REDUCTION WAS NEEDED SO THE DISTAL SCREWS WERE REMOVED ONCE. WHEN THE VA DRILLING WAS DONE AGAIN AFTER RE-REDUCTION, THE LOCKING SCREW WAS INSERTED AND IT WAS FOUND THAT THE PLATE WAS NOT LOCKED BY THE SCREW. WHEN THE FIX-MODE DRILLING WAS DONE TO ANOTHER HOLE, AND THE LOCKING SCREW WAS INSERTED, IT WAS ALSO FOUND THAT THE PLATE WAS NOT LOCKED BY THE SCREW. BECAUSE THE FIXATION WAS NOT POSSIBLE AT THIS MOMENT, THE SURGEON DECIDED TO REPLACE THE PLATE IN QUESTION. WHEN THE SURGEON TRIED TO REMOVE THE Y-SHAPE PLATE IN QUESTION, THE PLATE COULD NOT BE REMOVED SINCE THE PROXIMAL TWO SCREWS WERE SPINNING AROUND. EVENTUALLY, THE PLATE IN QUESTION WAS CUT AND REMOVED FROM THE PATIENT. THE SURGEON REPLACED AND FIXED WITH THE T-SHAPE PLATE (1.5 MM). THE SURGEON COMMENTED THAT SINCE THE VA SLEEVE WAS DRILLED BEYOND THE RANGE OF T

    Patient problems: No Code Available

    MAUDE report 7264340 ↗

Other Screw, fixation, bone devices