NA 201.772Recalls & reported problems

Synthes GmbH · 2.4MM CORTEX SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 22MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982142795), received between May 3, 2019 and Aug 18, 2021 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

5 other recalls cover devices cleared under the same FDA submission (K002271, K063049, K102694, K112583); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility4
  • Fitting Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT A SURGICAL PROCEDURE. DURING THE SURGERY, SURGEON ATTEMPTED TO PUT CORTEX SCREW THROUGH THE GUIDE BLOCK ON A TWO COLUMN DISTAL RADIUS PLATE. BUT THE STARDRIVE CORTEX SCREW HEADS DO NOT FIT THROUGH THE GUIDE BLOCK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 1-2 MINUTES. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN 2.4MM TWO COLUMN DISTAL RADIUS PLATE (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1) THIS COMPLAINT INVOLVES THREE(3) DEVICES. THIS REPORT IS FOR (1) 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 22 THIS REPORT IS 3 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12337818 ↗
  • MalfunctionAug 18, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT A SURGICAL PROCEDURE. DURING THE SURGERY, SURGEON ATTEMPTED TO PUT CORTEX SCREW THROUGH THE GUIDE BLOCK ON A TWO COLUMN DISTAL RADIUS PLATE. BUT THE STARDRIVE CORTEX SCREW HEADS DO NOT FIT THROUGH THE GUIDE BLOCK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 1-2 MINUTES. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN 2.4MM TWO COLUMN DISTAL RADIUS PLATE (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1) THIS COMPLAINT INVOLVES THREE(3) DEVICES. THIS REPORT IS FOR (1) 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 22 THIS REPORT IS 1 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12337811 ↗
  • MalfunctionOct 28, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, SOME OF THE 2.4MM CORTEX SCREWS WERE NOT FITTING THROUGH THE PLATE GUIDES FOR THE TWO COLUMN VARIABLE ANGLE DISTAL RADIUS PLATES. THE SET WAS OPENED AND TESTED THE THREE (3) 2.4MM CORTEX SCREWS 20MM, TWO (2) 2.4MM CORTEX 22MM, AND ONE (1) 2.4MM CORTEX 24MM. THE 2.4MM CORTEX SCREWS WOULD NOT FIT THROUGH THE GUIDE BLOCK FOR TWO-COLUMN PLATE 6 HOLE HEAD, GUIDE BLOCK FOR TWO-COLUMN PLATE 7 HOLE HEAD, NOR THE GUIDE BLOCK FOR TWO-COLUMN PLATE NARROW 6 HOLE. THE REST OF THE 2.4MM CORTEX SCREWS DID. THE ISSUE WAS FOUND IN STERILE PROCESSING DEPARTMENT (SPD). THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICES REPORTED: UNKNOWN CORTEX SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# UNKNOWN). THIS COMPLAINT INVOLVES NINE (9) DEVICES. THIS IS 5 OF 9 FOR REPORT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9244343 ↗
  • MalfunctionOct 28, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, SOME OF THE 2.4MM CORTEX SCREWS WERE NOT FITTING THROUGH THE PLATE GUIDES FOR THE TWO COLUMN VARIABLE ANGLE DISTAL RADIUS PLATES. THE SET WAS OPENED AND TESTED THE THREE (3) 2.4MM CORTEX SCREWS 20MM, TWO (2) 2.4MM CORTEX 22MM, AND ONE (1) 2.4MM CORTEX 24MM. THE 2.4MM CORTEX SCREWS WOULD NOT FIT THROUGH THE GUIDE BLOCK FOR TWO-COLUMN PLATE 6 HOLE HEAD, GUIDE BLOCK FOR TWO-COLUMN PLATE 7 HOLE HEAD, NOR THE GUIDE BLOCK FOR TWO-COLUMN PLATE NARROW 6 HOLE. THE REST OF THE 2.4MM CORTEX SCREWS DID. THE ISSUE WAS FOUND IN STERILE PROCESSING DEPARTMENT (SPD). THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICES REPORTED: UNKNOWN CORTEX SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# UNKNOWN). THIS COMPLAINT INVOLVES NINE (9) DEVICES. THIS IS 9 OF 9 FOR REPORT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9244317 ↗
  • MalfunctionMay 3, 2019Fitting Problem; Fitting Problem

    IT WAS REPORTED THAT ON (B)(6) 2019, THE SCRUB TRIED ON THE BACK TABLE AND ONE (1) CORTEX SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS 18MM BECAME INCARCERATED IN THE GUIDE BLOCK. ADDITIONALLY, THREE(3) CORTEX SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS 12MM, THREE(3) CORTEX SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS 14MM, THREE(3) CORTEX SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS 18MM, ONE (1) CORTEX SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS 22MM, TWO (2) CORTEX SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS  24MM, ONE (1) CORTEX SCREW SELF-TAPPING WITH T8 STARDRIVE RECESS 28MM WOULD NOT FIT THROUGH THE GUIDE BLOCK. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES SIXTEEN (16) DEVICES. THIS COMPLAINT INVOLVES 16 DEVICES DUE TO THE SYSTEM LIMITATION, THESE INCIDENTS WILL BE CAPTURED UNDER THIS (B)(4) FOR (10 DEVICES) AND LINKED (B)(4) FOR (6 DEVICES). THIS REPORT IF FOR SIX (6) DEVICES. THIS REPORT IS 2 OF 6 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8579772 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K002271, K063049, K102694, K112583. One submission often covers many models, so check the product and lot details in each record.

  • Z-2389-2016 · Jun 28, 2016 · Terminated
    Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone,
  • Z-0323-2016 · Nov 2, 2015 · Terminated
    2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone frag
  • Z-0167-2016 · Oct 6, 2015 · Terminated
    2.7 mm Locking Screw SLF-TPNG with T8 Stardrive Recess 12mm. The Synthes modular Mini Fragment LCP System is intended for fixation of fractures, osteotomies, nonunions, replantati
  • Z-0380-2014 · Jul 31, 2013 · Terminated
    2.4 and 2.7 MM Locking Screw SLF-TPNG with T8 Stardrive Recess 12 MM Fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragmen
  • Z-0956-2013 · Jan 11, 2013 · Terminated
    Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Cor

Other PLATE, FIXATION, BONE devices