Skip to content

NA 202.248Recalls & reported problems

Synthes GmbH · 2.7MM LOCKING SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 48MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982144157), received between Aug 10, 2018 and Nov 7, 2019 — classified as 2 injuries.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

5 other recalls cover devices cleared under the same FDA submission (K033995, K050110, K063049); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device-Device Incompatibility1

Patient problems

  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2019, DURING PLATING OF A PILON FRACTURE, THE SURGEON USED 3.5 MM LOCKING COMPRESSION PLATE (LCP) ANTEROLATERAL DISTAL TIBIA PLATE 9 HOLES RIGHT. THE PLATE ACCEPTS 3.5 MM LOCKING AND CORTICAL SCREWS PROXIMALLY AND 2.7 MM LOCKING DISTALLY. THERE WERE NO ISSUES WITH ANY OF THE 3.5 MM SCREWS PROXIMALLY, BUT NONE OF THE 2.7 MM LOCKING SCREWS HE DRILLED AND INSERTED DISTALLY ACTUALLY LOCKED INTO THE PLATE, THEY JUST SPUN. THE SURGEON DRILLED AND ATTEMPTED TO INSERT FOUR (4) OR FIVE (5) 2.7 MM LOCKING SCREWS DISTALLY, ALL OF WHICH FAILED, BEFORE RESORTING TO USING 2.7 MM CORTICAL SCREWS INSTEAD. HE USED THE APPROPRIATE LOCKING DRILL GUIDE, DRILL, AND SCREWS. THERE WAS NO SURGICAL DELAY. PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT OUTCOME WAS GOOD. CONCOMITANT DEVICE REPORTED: UNKNOWN LOCKING DRILL GUIDE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1), UNKNOWN 2.0 DRILL BIT (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1). THIS COMPLAINT INVOLVES SIX (6) DEVICES. THIS IS 6 OF 6 FOR REPORT (B)(4).

    Patient problems: No Code Available

    MAUDE report 9291773 ↗
  • InjuryAug 10, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT PREVIOUSLY SUSTAINED AN INJURY TO HIS LEFT ARM AND UNDERWENT A FIXATION OF THE SUPRACONDYLAR HUMERUS FRACTURE ON AN UNSPECIFIED DATE. THE PATIENT REPORTS SUBSEQUENT LOOSENING OF THIS HARDWARE. PATIENT IS NOW SEVERAL YEARS OUT FROM THIS PROCEDURE AND HAS PAIN AND DYSFUNCTION WITH DIFFICULTY LIFTING ON THAT SIDE. X-RAYS REVEALED EVIDENCE OF A NON-UNION WITH GROSS MOTION TO THE FRACTURE SITE, SURGICAL INTERVENTION FOR THE CORRECTION OF THE NON-UNION WAS PERFORMED ON (B)(6) 2015. PATIENT WAS RETURNED TO SURGERY ON (B)(6) 2016 FOR REPAIR OF A NONUNION LEFT SUPRACONDYLAR HUMERAL FRACTURE WAS PERFORMED USING PRODUCTS MANUFACTURED BY DEPUY AND MEDTRONIC. IT IS UNKNOWN IF THE DEVICES USED IN ORIGINAL PROCEDURE WERE SYNTHES OR NON SYNTHES DEVICES. DURING THIS PROCEDURE, THE FAILED HARDWARE WAS REMOVED WHICH INCLUDED A PLATE THAT WAS FOUND TO BE BROKEN, ALONG WITH AN UNSPECIFIED AMOUNT OF BROKEN AND FAILED SCREWS. PATIENT WAS AGAIN RETURNED TO SURGERY ON (B)(6) 2016 WHERE SURGEON PERFORMED A CORRECTION OF A NON-UNION OF LEFT SUPRACONDYLAR HUMERUS FRACTURE. DEPUY SYNTHES HARDWARE WAS EXPLANTED AND BONE GRAFT WAS IMPLANTED. SURGEON ALSO PERFORMED NEUROLYSIS OF THE LEFT U

    Patient problems: Not Applicable

    MAUDE report 7774888 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033995, K050110, K063049. One submission often covers many models, so check the product and lot details in each record.

  • Z-2389-2016 · Jun 28, 2016 · Terminated
    Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone,
  • Z-0323-2016 · Nov 2, 2015 · Terminated
    2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone frag
  • Z-0167-2016 · Oct 6, 2015 · Terminated
    2.7 mm Locking Screw SLF-TPNG with T8 Stardrive Recess 12mm. The Synthes modular Mini Fragment LCP System is intended for fixation of fractures, osteotomies, nonunions, replantati
  • Z-0380-2014 · Jul 31, 2013 · Terminated
    2.4 and 2.7 MM Locking Screw SLF-TPNG with T8 Stardrive Recess 12 MM Fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragmen
  • Z-2278-2013 · Mar 8, 2013 · Terminated
    Synthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for fixation of fractures of the distal humerus, olecranon, and ulna in adults

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →