Skip to content

NA 202.878Recalls & reported problems

Synthes GmbH · 2.7MM CORTEX SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 18MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982145086), received between Dec 26, 2017 and Jul 10, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

6 other recalls cover devices cleared under the same FDA submission (K010321, K050110, K063049, K102694, K112583, K113364); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Entrapment of Device1
  • Material Twisted/Bent1

Patient problems

  • No Consequences Or Impact To Patient2
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 10, 2025Break; Break

    IT WAS REPORTED THAT THE PATIENT HAD A PROCEDURE WITH THE CORTEX SCREW SLF-TPNG T8 SD REC 18. POSTOPERATIVELY, THE IMPLANTED ROD BROKE, AND PATIENT UNDERWENT ANOTHER PROCEDURE. PATIENT SUFFERED PAIN AS A RESULT OF THE EVENT. THE PATIENT WAS REPORTED TO BE DOING WELL POSTOPERATIVELY.

    Patient problems: Pain

    MAUDE report 22467531 ↗
  • MalfunctionAug 19, 2022Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2022, THE DRILL BIT BROKE IN THE MIDDLE OF THE DRILL. SEVERAL SCREW HEADS APPEARED TO HAVE A BAD HEAD AND WERE NOT USED IN SURGERY. SURGERY WAS SUCCESSFULLY COMPLETED. PATIENT STATUS WAS NOTED AS OK. THIS IS REPORT 8 OF 11 FOR (B)(4). THIS REPORT IS FOR A CORTEX SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15261319 ↗
  • InjuryJun 3, 2022Break; Entrapment of Device; Break; Entrapment of Device

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION (ORIF) PROCEDURE FOR THE CLAVICLE, THE CORTEX SCREW HEAD SHEARED OFF ON TIGHTENING IN THE MOST MEDIAL HOLE OF THE LATERAL CLAVICLE PLATE. THE SCREW HEAD WAS RETRIEVED AND THE SCREW SHAFT WAS RETAINED IN THE CLAVICLE. THE PLATE IS PLANNED TO BE REMOVED IN THREE TO NINE (3 ¿ 9) MONTHS AND THE SCREW SHAFT WILL BE REMOVED ALONG WITH IT. NO SURGICAL DELAY WAS ENCOUNTERED AS THE SCREW WAS LEFT AS IS AND THE OTHER HOLES OF THE PLATE WERE FILLED. THIS REPORT IS FOR A CORTEX SCREW. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 14598000 ↗
  • MalfunctionJul 3, 2018Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT A SURGICAL PROCEDURE OF A BROKEN PALM FRACTURE TO FIX ULNA WITH A LOCKING COMPRESSION PLATE (LCP). UPON INSERTION OF THE FINAL CORTEX SCREW, THE HEAD OF THE SCREW CAME OFF. THE SURGEON DECIDED TO REMOVE THE ENTIRE CONSTRUCT TO REMOVE THE SCREW FRAGMENT AND UPON AN ATTEMPT TO REMOVE THE NEXT SCREW IN THE PLATE, THE SCREW CAME OFF. THE SURGEON REMOVED ALL THE SCREWS AND THE PLATE. A BROKEN SCREW REMOVAL SET WAS USED TO TAKE OUT THE SCREW FRAGMENT. THERE WAS A SURGICAL DELAY OF AN HOUR REPORTED TO AN ATTEMPT TO REMOVE THE SCREW. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICES REPORTED: SCREWS (PART 202.873, LOT UNKNOWN, QUANTITY 3), 2.7MM CORTEX SCREW, SELF-TAPPING 16MM (PART 202.876, LOT UNKNOWN, QUANTITY 1), 2.7MM CORTEX SCREW, SELF-TAPPING 20MM (PART 202.880, LOT UNKNOWN, QUANTITY 1), 2.0MM LOCKING COMPRESSION PLATE 7 HOLES (PART 247.347, LOT NUMBER UNKNOWN, QUANTITY 1). THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7659002 ↗
  • MalfunctionDec 26, 2017Break; Break

    IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE ULNA PROCEDURE ON (B)(6) 2017, THE HEAD OF A 2.7MM 18MM CORTEX SCREW BROKE OFF INTO THE PATIENT AS THE SURGEON WAS SCREWING IT INTO A 2.7MM LOCKING COMPRESSION PLATE (LCP) PLATE. FRAGMENTS WERE GENERATED IN THE PATIENT AND PART OF THE SCREWS THREADING WAS REPORTED TO BE LEFT IN THE PATIENT¿S ARM. ALL OTHER FRAGMENTS WERE REPORTED AS EASILY REMOVED. DURING THE TIME OF THE BREAK, TWO (2) OTHER 2.7MM 18MM CORTEX SCREWS AND THE 2.7MM LCP PLATE WERE IMPLANTED IN THE PATIENT. THE PLATE AND 2 SCREWS REMAINED IN THE PATIENT AND THE SURGERY WAS COMPLETED USING ANOTHER 2.7MM 18MM CORTEX SCREW. THE PATIENT WAS REPORTED TO BE IN GOOD CONDITION FOLLOWING THE PROCEDURE. THERE WAS A REPORTED 10-MINUTE SURGICAL DELAY DUE TO THE BREAK. CONCOMITANT DEVICE REPORTED: 2.7MM LCP PLATE 10 HOLES / 94MM (247.374, LOT NUMBER UNKNOWN, QUANTITY 1), SCREWDRIVER (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1). (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7146262 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K010321, K050110, K063049, K102694, K112583, K113364. One submission often covers many models, so check the product and lot details in each record.

  • Z-2389-2016 · Jun 28, 2016 · Terminated
    Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone,
  • Z-0323-2016 · Nov 2, 2015 · Terminated
    2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone frag
  • Z-0167-2016 · Oct 6, 2015 · Terminated
    2.7 mm Locking Screw SLF-TPNG with T8 Stardrive Recess 12mm. The Synthes modular Mini Fragment LCP System is intended for fixation of fractures, osteotomies, nonunions, replantati
  • Z-1274-2014 · Jan 7, 2014 · Terminated
    Drill Template and the Saw Guide for the Ulna Osteotomy System. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bon
  • Z-0380-2014 · Jul 31, 2013 · Terminated
    2.4 and 2.7 MM Locking Screw SLF-TPNG with T8 Stardrive Recess 12 MM Fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragmen
  • Z-0956-2013 · Jan 11, 2013 · Terminated
    Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Cor

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page