Skip to content

NA 205.040Recalls & reported problems

Synthes GmbH · 3.5MM CANNULATED SCREW PARTIALLY THREADED/40MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982146892), received between Apr 19, 2018 and Dec 8, 2020 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Bent1
  • Loose or Intermittent Connection1
  • Material Twisted/Bent1

Patient problems

  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 8, 2020Material Twisted/Bent; Material Twisted/Bent

    UPDATE: CONCOMITANT DEVICES REPORTED: CANNSCR Ø3.5 L40/13 SST - PART #205.040, LOT #2L84635.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10970854 ↗
  • MalfunctionNov 13, 2018Break; Break

    IT WAS REPORTED ON AN UNKNOWN DATE, A CANNULATED SCREW PARTIALLY THREADED BROKE UPON INSERTION INTO THE BONE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. IT IS UNKNOWN IF THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8066283 ↗
  • MalfunctionJun 7, 2018Break; Break

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, PATIENT UNDERWENT ELBOW REDUCTION PROCEDURE. DURING THE PROCEDURE, THE PARTIALLY THREADED CANNULATED SCREW BEGAN TO DISINTEGRATE AND LOSES THE THREAD UPON INSERTION. THERE WAS A SURGICAL DELAY OF APPROXIMATELY AN HOUR AND THIRTY MINUTES. THERE WAS NO PATIENT HARM REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) 3.5MM CANNULATED SCREW PARTIALLY THREADED/40MM. (B)(4).

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 7577954 ↗
  • InjuryApr 19, 2018Bent; Loose or Intermittent Connection; Bent; Loose or Intermittent Connection

    IT WAS REPORTED BY ARGENTINA THAT A PATIENT UNDERWENT AN ANTECEDENT OSTEOSYNTHESIS OF TIBIA (KNEE JOINT) BY LCA AVULSION IN 2012 WITH TWO (2) 3.5MM CANNULATED SCREWS. PATIENT HAD REVISION SURGERY ON (B)(6) 2017 DUE TO LOOSENING OF PAINFUL OSTEOSYNTHESIS. THE PATIENT HAD SOME PAIN IN HIS KNEE. AN X-RAY TAKEN ON AN UNKNOWN DATE SHOWED ONE OF THE SCREWS TURNED INTO THE JOINT (LOOSE SCREW) AND THE OTHER SCREW WAS BENT. MOVEMENT OF THE LOOSE SCREW CAUSED DAMAGE TO THE FEMORAL CARTILAGE AND SURGEON DECIDED TO REMOVE BOTH SCREWS. THE LOOSE SCREW WAS COMPLETELY REMOVED. THE LOOSE SCREW DID NOT BREAK DURING REMOVAL AND THERE WERE NO FRAGMENTS LEFT IN THE PATIENT. THE SECOND SCREW WAS BENT PRIOR TO REMOVAL. INITIALLY THE SURGEON TRIED USING THE SCREWDRIVER BUT WAS UNABLE TO REMOVE THE SCREW IN ITS BENT STATE AND A FORCE CLAMP WAS USED FOR REMOVAL AND IN THAT MOMENT THE HEAD OF THE SCREW BROKE. THERE WERE NO FRAGMENTS FROM THE SECOND SCREW BECAUSE A FORCE CLAMP WAS USED. BOTH SCREWS WERE COMPLETED REMOVED. PATIENT WAS NOT REVISED TO ANY ADDITIONAL HARDWARE, AS THE FRACTURE HAD CONSOLIDATED. BOTH SCREWS ARE AVAILABLE FOR RETURN. IT IS UNKNOWN AS TO WHICH SCREW WAS LOOSE AND WHICH SCREW WAS BEN

    Patient problems: Failure of Implant

    MAUDE report 7444069 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates