NA 206.040Recalls & reported problems
Synthes GmbH · 4.0MM CANCELLOUS BONE SCREW FULLY THREADED/40MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982147349), received between Nov 5, 2018 and Nov 16, 2022 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break2
Patient problems
- Hypersensitivity/Allergic reaction1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- Pain1
- Rash1
- Skin Irritation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 16, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, THE DRILL BITS AND SCREWS BROKE WHILE DRILLING AROUND A FEMORAL PROSTHESIS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED; IT IS UNKNOWN WHETHER THEY WERE EASILY REMOVED WITHOUT FURTHER INTERVENTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE 4.0MM CANCELLOUS BONE SCREW FULLY THREADED/40MM. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15805004 ↗InjuryNov 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT A REMOVAL SURGERY OF PROXIMAL TIBIA HARDWARE DUE TO PAIN. THE PROCEDURE SUCCESSFULLY COMPLETED. THERE WAS NO KNOWN PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES FIVE (5) DEVICES. THIS REPORT IS FOR ONE (1) 4.0 CANCELLOUS BONE SCREW FULLY THRD/40. THIS IS REPORT 4 OF 5 FOR (B)(4).
Patient problems: Pain
MAUDE report 12778763 ↗MalfunctionOct 14, 2019Break; Break
IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE PROXIMAL HUMERUS ON (B)(6) 2019, THE 4.0 CANCELLOUS SCREW SNAPPED HALFWAY INTO THE INTRAMEDULLARY. THE PROCEDURE WAS COMPLETED BY RETRIEVING THE BROKEN PORTION WITH NARROW RONGEURS, INSIDE THE CANAL, AND THEN PROCEEDING AS PLANNED, WITH PLATES AND SCREWS. THE SURGEON THEN ELECTED TO USE A CORTICAL SCREW, IN PLACE OF THE CANCELLOUS. THE SCREW WAS BEING RUN THREW A PLATE. THE ISSUE HAPPENED WHILE THE SURGEON WAS IMPLANTING THE REPORTED SCREW. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH 5 MINUTES SURGICAL DELAY. THE SCREW FRAGMENT WAS EASILY RETRIEVED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN PLATE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). THIS IS REPORT 1 FOR 1 (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9186352 ↗InjuryNov 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS IMPLANTED WITH A LOCKING COMPRESSION PLATE (LCP) 8 HOLE PLATE, FIVE (5) CORTEX SCREWS, TWO (2) CANNULATED SCREWS, AND ONE (1) CANCELLOUS BONE SCREW ON (B)(6) 2017 FOR REPAIR OF A BROKEN RIGHT TIBIA AND FIBULA. SURGERY WAS COMPLETED SUCCESSFULLY. IN (B)(6) 2017, PATIENT REPORTS SKIN ISSUES. IN (B)(6) 2018, PATIENT REPORTS BOTH LEGS FROM KNEE TO FOOT, FOOT AND HANDS, HAD A RASH AND SKIN IRRITATION. PATIENT WAS REFERRED TO A DERMATOLOGIST IN (B)(6) 2018. DERMATOLOGIST DETERMINED PATIENT IS ALLERGIC TO THE IMPLANTED DEVICES. THIS REPORT IS FOR ONE (1) 4.0MM CANCELLOUS BONE SCREW. THIS IS REPORT 7 OF 9 FOR (B)(4).
Patient problems: Hypersensitivity/Allergic reaction; Rash; Skin Irritation
MAUDE report 8038372 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page