NA 207.630Recalls & reported problems

Synthes GmbH · 4.0MM CANNULATED SCREW SHORT THREAD/30MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982148193), received between Feb 4, 2025 and Feb 4, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Entrapment of Device1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 4, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2025, PATIENT WAS TREATED FOR BILATERAL PARS FRACTURE. SURGEON WAS USING 4.0 CANNULATED SCREWS (STAINLESS STEEL, SHORT THREAD) FOR THE PROCEDURE. BOTH WIRES WERE PLACED AND ONE SCREW WAS DRILLED AND PUT IN AT A TIME. THE FIRST SCREW WAS PLACED AND WAS LEFT IN POSITION. WHEN PUTTING IN THE SECOND SCREW, THE PRE-DRILL WAS DONE AND WHEN THE SCREW WAS GOING IN, THE THREADS OF THE SCREW CAME UNRAVELED. THE SCREW WAS THEN BACKED OUT BUT THE REST OF THE SCREW (THREADS) WERE STUCK IN THE PATIENT. THE WIRE WAS NEVER MOVED OUT OF POSITION AND REMAINED THE ENTIRE TIME AS THE SCREW WAS PLACED. DUE TO THIS COMPLICATION, THE PATIENT NOW NEEDED TO UNDERGO AN EMERGENCY LAMINECTOMY TO REMOVE THE REST OF THE SCREW FROM THE SPINAL CANAL.

    Patient problems: Foreign Body In Patient

    MAUDE report 21307399 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →