Skip to content

NA 214.542Recalls & reported problems

Synthes GmbH · 4.5MM CANNULATED SCREW PARTIALLY THREADED/42MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982154170), received between Nov 5, 2019 and May 11, 2021 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT REVISION OF HARDWARE DUE TO FALL. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY. PATIENT CONSEQUENCE IS UNKNOWN. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR (1) 4.5 CANNULATED SCREW PARTIALLY THRD/40 THIS REPORT IS 2 OF 2 (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11809111 ↗
  • MalfunctionNov 5, 2019Crack; Crack

    IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE TIBIAL TUBERCLE ON (B)(6) 2019, THE 4.5 CANNULATED SCREW WAS PLACED INTO THE PATIENT AND THE THREAD CAME OFF. THE SURGEON STATED THAT HE BELIEVED THE THREAD CAME OFF DUE TO EXTREMELY HARD BONE AND THAT HE DID NOT PRE DRILL. DURING SCREW INSERTION THE THREAD OF THE SCREW UNWOUND OFF THE CORE OF THE SCREW. NO ATTEMPTS WERE MADE TO RETRIEVE THE FRAGMENT. DEVICE WAS LEFT IN THE PATIENT. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. PATIENT OUTCOME WAS GOOD. CONCOMITANT DEVICES: PLATE (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A 4.5MM CANNULATED SCREW. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 9280538 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page