Skip to content

DHS 338.26Recalls & reported problems

Synthes GmbH · TIP FOR DHS®/DCS® IMPACTOR (338.28) · Class 1

Quick summary

This Synthes DHS/DCS Impactor tip is a surgical instrument piece used to hammer bone screws and plates into place during hip fracture surgery. The 35 reports describe the tip breaking, chipping or splitting, sometimes during impaction of the plate, and the instrument then becoming unusable. No deaths and no injuries were reported, and most cases mention no effect on the patient or delay to surgery. No recalls name this device, and reports have slowed sharply, from 2 in the prior 12 months to none in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 10886982193292), received between Apr 13, 2017 and Aug 12, 2025 — classified as 35 malfunctions.

The most frequently reported specific device problem is “Break” (35 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break35
  • Dull, Blunt1
  • Material Integrity Problem1
  • Material Twisted/Bent1

Patient problems

  • No Consequences Or Impact To Patient15
  • No Clinical Signs, Symptoms or Conditions12
  • No Patient Involvement5
  • Insufficient Information1
  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2025Break; Break

    IT WAS REPORTED THAT THAT ON AN UNKNOWN DATE THE TIP OF THE DHS IMPACTOR IS DAMAGED AND NEEDS REPLACED. THE TIP IS CHIPPING AND PARTS BROKEN OFF. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. IT WAS FINISHED IMPACTING TO GET IMPLANT DOWN. THERE WAS NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22788452 ↗
  • MalfunctionApr 2, 2025Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2025, DURING THE PROCEDURE, WHICH WAS A STANDARD DHS, UPON FOLLOWING THE STEPS PER THE SURGICAL TECHNIQUE, THE DHS/DCS IMPACTOR BROKE WHILE TRYING TO SEAT THE PLATE. SPECIFICALLY, THE TIP OF THE IMPACTOR BROKE. THERE WAS NO EXCESSIVE FORCE APPLIED TO THE IMPACTOR. THE BROKEN TIP WAS RECOVERED, HOWEVER THE IMPACTOR AND TIP AS A WHOLE WERE UNUSABLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAYS, NEGATIVE OUTCOMES, OR INTERVENTION REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 21756606 ↗
  • MalfunctionJun 12, 2024Break; Break

    IT WAS REPORTED THAT ON UNKNOWN DATE WHILE WORKING ON SETS, IT WAS DISCOVERED THAT THE IMPACTOR TIP WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) TIP FOR DHS/DCS IMPACTOR. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19520146 ↗
  • MalfunctionJan 16, 2024Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2023, DURING A DHS CASE OVER THE PAST HOLIDAY WEEKEND, AFTER THE LAG SCREW A CORRECTLY IMPLANTED, THE PLATE WAS THEN INSERTED OVER THE LAG SCREW. THE PLATE WAS THEN MALLETED OVER THE TOP TO SIT FLUSH WITH THE LATERAL CORTEX, UPON MALLETING WITH THE MALLET AND IMPACTOR, THE PLASTIC END OF THE IMPACTOR BROKE INTO TWO PIECES. ONCE THE PIECE (338.26) BROKE, THE SURGEON AND RESIDENT AGREED THAT THE IMPLANT WAS PERFECTLY PLACED AND NO FURTHER MALLETING WAS NECESSARY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY. NO PATIENT CONSEQUENCE WAS REPORTED. NO FRAGMENTS WERE GENERATED. THIS REPORT IS FOR ONE (1) TIP FOR DHS/DCS IMPACTOR (338.28) THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18519488 ↗
  • MalfunctionDec 6, 2023Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED DURING A PROCEDURE ON (B)(6) 2023 THE TIP OF AN IMPACTOR BROKE OFF INTRAOPERATIVELY DURING IMPACTION. NO SURGICAL DELAY REPORTED. PROCEDURE COMPLETELY SUCCESSFULLY. THIS REPORT IS FOR TIP FOR DHS/DCS IMPACTOR (338.28) FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18275625 ↗
  • MalfunctionJul 7, 2023Break; Break

    IT WAS REPORTED THAT DURING A HIP SURGERY ON (B)(6) 2023, THE DHS IMPACTOR PIECE BROKE OFF WHILE IN USE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO DELAY. NO FRAGMENTS WERE GENERATED. THIS REPORT INVOLVES ONE TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17277672 ↗
  • MalfunctionMar 16, 2023Break; Break

    DEVICE REPORT FROM DEPUY SYNTHES, REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT THE IMPACTOR TIP WAS SPLIT OPEN DURING INSERTION. THERE WAS NO SURGICAL DELAY AND NO PATIENT CONSEQUENCES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) TIP FOR DHS/DCS IMPACTOR. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16558394 ↗
  • MalfunctionJul 11, 2022Break; Material Twisted/Bent; Break; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THE TIP OF IMPACTOR WAS WORN-OUT. NO KNOWN PATIENT OR SURGICAL INVOLVEMENT. DURING INVESTIGATION IT WAS DETERMINED THAT THE DEVICE TIP BROKE. IT IS UNKNOWN WHEN MALFUNCTION OCCURRED. THIS IS REPORT 1 OF 1 FOR (B)(4). THIS REPORT IS FOR A DHS/DCS IMPACTOR TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14977136 ↗
  • MalfunctionMay 23, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DHS IMPACTOR TIP CHIPPED OFF. THIS COMPLAIN INVOLVES ONE DEVICE INS F/DHS/DCS IMPACTOR NO. 338.280 SINGL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14468867 ↗
  • MalfunctionApr 21, 2022Break; Material Integrity Problem; Break; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED A DYNAMIC HIP SCREW (DHS) SYSTEM IMPACTOR WAS NOTED AS DAMAGED IN A DHS SET. THE DEVICE HAS BEEN DAMAGED THROUGHOUT YEARS OF USE. NO PATIENT INVOLVEMENT. THIS IS REPORT 1 OF 1 FOR (B)(4). THIS REPORT IS FOR A DHS IMPACTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14166977 ↗
  • MalfunctionApr 8, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT PLASTIC PORTION OF DHS IMPACTOR BROKE OFF DURING INSERTION. PROCEDURE WAS ABLE TO BE COMPLETED WITHOUT ANY COMPLICATIONS. THERE WAS NO SURGICAL DELAY AND PATIENT CONSEQUENCES. THIS COMPLAIN INVOLVES ONE(1) DEVICE DHS/DCS IMPACTOR. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14053623 ↗
  • MalfunctionJan 20, 2022Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2021, THE TIP FOR DHS IMPACTOR WAS BROKEN. A FRAGMENT WAS GENERATED AND WAS EASILY REMOVED WITHOUT ADDITIONAL INTERVENTION. THERE WAS A SURGICAL DELAY OF TWO (2) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13308337 ↗
  • MalfunctionFeb 1, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, WHILE INSERTING A DYNAMIC HIP SYSTEM (DHS) PLATE WITH THE IMPACTOR, THE PLASTIC PORTION OF THE IMPACTOR BROKE. THE PIECES WERE RETRIEVED SUCCESSFULLY, AND THE SURGERY WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT HARM OR SURGERY DELAY. THIS REPORT INVOLVES ONE (1) TI TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11258588 ↗
  • MalfunctionSep 21, 2020Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2020, DURING THE APPLICATION OF A DYNAMIC HIP SYSTEM (DHS) FOR AN INTERTROCHANTERIC FRACTURE THE DHS COUPLING SCREW BROKE ALONG WITH THE TWO (2) TIP FOR DHS IMPACTOR. IT WAS UNKNOWN IF THERE WERE FRAGMENTS GENERATED. ANOTHER SET WAS OPENED, AND THE ITEMS WERE PULLED FROM THE SECOND SET. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICES REPORTED: TIP FOR DHS®/DCS® IMPACTOR (338.28) (PART NUMBER 338.26, LOT 6043333, QUANTITY 1), DHS®/DCS® COUPLING SCREW (PART NUMBER 338.20, LOT 1165694, QUANTITY 1). THIS REPORT INVOLVES ONE (1) TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 2 OF 3 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10560202 ↗
  • MalfunctionSep 21, 2020Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2020, DURING THE APPLICATION OF A DYNAMIC HIP SYSTEM (DHS) FOR AN INTERTROCHANTERIC FRACTURE THE DHS COUPLING SCREW BROKE ALONG WITH THE TWO (2) TIP FOR DHS IMPACTOR. IT WAS UNKNOWN IF THERE WERE FRAGMENTS GENERATED. ANOTHER SET WAS OPENED, AND THE ITEMS WERE PULLED FROM THE SECOND SET. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICES REPORTED: TIP FOR DHS®/DCS® IMPACTOR (338.28) (PART NUMBER 338.26, LOT H479753, QUANTITY 1), DHS®/DCS® COUPLING SCREW (PART NUMBER 338.20, LOT 1165694, QUANTITY 1). THIS REPORT INVOLVES ONE (1) TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 1 OF 3 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10560155 ↗
  • MalfunctionSep 15, 2020Break; Break

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2020, A DYNAMIC HIP SCREW (DHS) IMPACTOR BROKE DURING IMPACTION OF THE DHS PLATE. ANOTHER SET WAS OPENED TO COMPLETE THE CASE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A TWO (2) MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE. AN ADDITIONAL DEVICE WAS RECEIVED BY THE MANUFACTURER ON AUGUST 24, 2020. CONCOMITANT DEVICES: DHS/DCS PLATE (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A TIP FOR DHS®/DCS® IMPACTOR. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10535195 ↗
  • MalfunctionAug 27, 2020Break; Break

    IT WAS REPORTED ON (B)(6) 2020, A PIECE BROKE OFF THE END OF THE IMPACTOR FOR DYNAMIC HIP SCREW/DYNAMIC CONDYLAR SCREW (DHS/DCS). THIS WAS NOTICED DURING CLEANING. THERE WAS NO PATIENT AND SURGICAL INVOLVEMENT. THIS REPORT IS FOR A TIP FOR DHS®/DCS® IMPACTOR. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10459330 ↗
  • MalfunctionJul 7, 2020Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2020, DURING IMPACTION OF THE UNKNOWN DHS PLATE. THE IMPACTOR TIP BROKE OFF. ALL PIECES WERE REMOVED. FRAGMENTS WERE GENERATED, THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THERE WAS NO SURGERY DELAY. PROCEDURE SUCCESSFULLY COMPLETED. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNK - PLATES: DHS/DCS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10239635 ↗
  • MalfunctionJun 4, 2020Break; Break

    CONCOMITANT DEVICES REPORTED: UNK - PLATES: DHS/DCS (PART#UNKNOWN, LOT# UNKNOWN, QUANTITY# 1).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10118596 ↗
  • MalfunctionMar 25, 2020Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE DHS IMPACTOR WAS BROKEN DURING IMPACTION OF THE DHS PLATE. ANOTHER SET WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS A SURGICAL DELAY OF TWO (2) MINUTES. NO FURTHER INFORMATION PROVIDED. CONCOMITANT DEVICE REPORTED: UNK - PLATES: DHS/DCS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1). THIS REPORT IS FOR ONE (1) TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9879851 ↗
  • MalfunctionDec 3, 2019Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIP FOR DYNAMIC HIP AND CONDYLAR SCREW SYSTEM (DHS/DCS) IMPACTOR BROKE. THE ISSUE WAS FOUND OUT IN THE STERILE PROCESSING DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS 1 OF 1 FOR REPORT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9401873 ↗
  • MalfunctionAug 29, 2019Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE TIPS OF THE TWO (2) DYNAMIC HIP SYSTEM (DHS) IMPACTORS BROKE. NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) TIP FOR DHS/DCS IMPACTOR (338.28). THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8946118 ↗
  • MalfunctionAug 29, 2019Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE TIPS OF THE TWO (2) DYNAMIC HIP SYSTEM (DHS) IMPACTORS BROKE. NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) TIP FOR DHS/DCS IMPACTOR (338.28). THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8946094 ↗
  • MalfunctionMay 10, 2019Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2019, THERE WERE NOTCHES ON THE PLASTIC HANDLE OF THE INSERT FOR DYNAMIC HIP SCREW (DHS)/DYNAMIC CONDYLAR SCREW (DCS) IMPACTOR. THE DHS/DCS IMPACTOR ALSO WOULD DISCONNECT. IT IS UNKNOWN IF WHEN WAS THE ISSUE WAS DISCOVERED. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) TIP FOR DHS®/DCS® IMPACTOR (338.28). THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8599944 ↗
  • MalfunctionJun 19, 2018Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2018, A DYNAMIC HIP SCREW/ DYNAMIC CONDYLAR SCREW (DHS/DCS) IMPACTOR WAS FOUND BROKEN AND UNUSABLE DURING RESTOCKING OF SET. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A TIP FOR DHS®/DCS® IMPACTOR (338.28) THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 7613809 ↗

Other IMPACTOR devices

All IMPACTOR devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.