NA 352.251SRecalls & reported problems
Synthes GmbH · 12.5MM REAMER HEAD-STERILE FOR REAMER/IRRIGATOR/ASPIRATOR · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982194442), received between Nov 19, 2018 and Mar 9, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
19 other recalls cover devices cleared under the same FDA submission (K013527, K042899, K111437); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN PROCEDURE, THE RIA REAMING ROD AND REAMING DRIVE SHAFT BROKE. FRAGMENTS WERE NOT GENERATED OR LEFT IN THE PATIENT. THERE WAS A SURGICAL DELAY OF ONE (1) MINUTE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT STATUS WAS UNKNOWN. THIS REPORT IS FOR ONE (1) 12.5MM REAMER HEAD-STERILE FOR REAMER/IRRIGATOR/ASPIRATOR. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11445723 ↗InjuryNov 19, 2018Break; Break
11/21/2018: UPDATE EVENT DESCRIPTION TO CAPTURE 2 DEVICES INVOLVEMENT: IT WAS REPORTED THAT ON NOVEMBER 01, 2018, THE SURGEON WAS USING THE REAMER/IRRIGATOR/ASPIRATOR (RIA) AND WHILE REAMING THE NONUNION OF THE TIBIA, THE REAMER HEAD BROKE IN THE CANAL ALONG WITH A PIECE OF THE PLASTIC REAMER SHAFT. THE SURGEON WAS ABLE TO REMOVE THE REAMER HEAD AND THE BROKEN PIECE OF PLASTIC FROM THE CANAL. AN EXTRA INCISION WAS MADE TO REMOVE THE PIECE OF PLASTIC AT THE ENTRY POINT FOR THE TIBIAL NAIL. THE REAMER HEAD WAS REMOVED WITH REAMING ROD AND THE PIECE OF PLASTIC WAS REMOVED THROUGH A SEPARATE INCISION BELOW THE PATELLA. THERE WAS A FIVE-MINUTE SURGICAL DELAY REPORTED. THE PROCEDURE WAS COMPLETED AS PLANNED. PATIENT OUTCOME WAS UNKNOWN.
Patient problems: No Code Available; Nonunion/Delayed-union Bone Fracture
MAUDE report 8086646 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K013527, K042899, K111437. One submission often covers many models, so check the product and lot details in each record.
- Z-1237-2024 · Jan 24, 2024 · Open, ClassifiedRIA Tube Assembly, minimum 520 mm length, sterile- intended for use in adults and adolescents (12-21 years) to clear the medullary canal of bone marrow and debris and to effective
- Z-0584-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 16.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.029S
- Z-0571-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 10.0MM REAMER HEAD FOR RIA 2, STERILE-Orthopedic manual surgical instrument Part Code: 03.404.016S
- Z-0573-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 11.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.018S
- Z-0577-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 13.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.022S
- Z-0585-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 17.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.030S
- Z-0576-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 12.5MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.021S
- Z-0587-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 18.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.032S
- Z-0586-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 17.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.031S
- Z-0583-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 16.0MM REAMER HEAD FOR RIA 2 STERILE--Orthopedic manual surgical instrument Part Code: 03.404.028S
- Z-0572-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 10.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.017S
- Z-0580-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 14.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.025S
- Z-0581-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 15.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.026S
- Z-0582-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 15.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.027S
- Z-0575-2021 · Nov 19, 2020 · TerminatedDePuy Synthes 12.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.020S
…and 4 more.
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page