NA 356.543Recalls & reported problems
Synthes GmbH · EXTRACTION SCREW FOR TITANIUM TIBIAL NAIL · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982195500), received between Dec 15, 2020 and Jul 25, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent2
- Break1
- Device-Device Incompatibility1
- Entrapment of Device1
- Material Deformation1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE EXTRACTION SCREW WAS FOUND TO BE BENT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE EXTRACTION SCREW FOR TITANIUM TIBIAL NAIL. THIS IS REPORT 1 OF 1 FOR (B)(4) .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19829293 ↗MalfunctionAug 7, 2023Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT DURING SURGERY ON (B)(6) 2023, THE EXTRACTOR THREADS WERE DAMAGED. THE DEVICE COULD NO LONGER BE ENGAGED PROPERLY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY, AND THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE EXTRACTION SCREW FOR TITANIUM TIBIAL NAIL. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17471092 ↗MalfunctionMar 17, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE ITEMS HAD STRIPPED THREADS ON THEM. IT WAS DISCOVERED DURING ROUTINE SET EVALUATION POSTOPERATIVELY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE EXTRACTION SCREW FOR TITANIUM TIBIAL NAIL. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16564144 ↗MalfunctionMar 30, 2021Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 DURING IMPLANT REMOVAL EXTRACTION SCREW DID NOT THREAD PROPERLY. SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY DELAY, BUT THE EXTRACTION SCREW NEEDS TO BE REPLACED. THIS REPORT IS FOR ONE (1) EXTRACTION SCREW FOR TITANIUM TIBIAL NAIL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11589751 ↗InjuryJan 6, 2021Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT ON (B)(6) 2020, DURING THE TITANIUM TIBIA NAIL REMOVAL PROCEDURE, THE THREADS OF EXTRACTION SCREW FOR TIBIAL NAIL BROKE. THE SURGEON THREADED THE EXTRACTION SCREW INTO CANNULATED TIBIA NAIL AND AFTER THE REPEATED BACK SLAPPING BLOWS, THE THREADS OF THE EXTRACTION SCREW FOR TIBIAL NAIL BROKE OFF. THE NAIL WAS RETAINED IN THE PATIENT ALONG WITH THE BROKEN THREADS. THE REMAINING SHAFT OF EXTRACTION SCREW WAS DISCARDED. THERE WAS A FIFTEEN (15) MINUTES OF SURGICAL DELAY. THE PROCEDURE WAS NOT SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE: UNKNOWN TITANIUM TIBIA NAIL (PART# UNKNOWN, LOT # UNKNOWN, QTY 1). THIS REPORT IS FOR ONE (1) EXTRACTION SCREW FOR TITANIUM TIBIAL NAIL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 11127510 ↗MalfunctionDec 15, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2020 THAT THE EXTRACTION SCREW SEIZED WITHIN THE RAM GUIDE. THE ISSUE WAS DISCOVERED DURING LOAN KIT INSPECTION. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) EXTRACTION SCREW FOR TITANIUM TIBIAL NAIL. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11007877 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.