NA 394.43Recalls & reported problems

Synthes GmbH · SLIDING CARRIAGE · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982199713), received between Sep 4, 2018 and Nov 20, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED DURING A PROCEDURE ON AN (B)(6) 2025, THE COMPONENTS OF THE LARGE DISTRACTOR NO LONGER HELD THE DISTRACTIONS; THEY KEPT SLIPPING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A FIVE (5) MINUTE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23605668 ↗
  • MalfunctionSep 4, 2018Device-Device Incompatibility; Device-Device Incompatibility

    (B)(6) REPORTS FOLLOWING: IT WAS REPORTED THAT DURING STERILE PROCESSING DEPARTMENT (SPD) INSPECTION ON AN UNKNOWN DATE, THE SLIDING CARRIAGE WAS NOTED TO BE CROSS THREADED AND SEIZED ON TWO (2) HOLDING SLEEVES AND A SCHANZ SCREW BLUNTED TROCAR. THERE WAS NO PATIENT INVOLVEMENT. DURING MANUFACTURING INVESTIGATION, THE CONDITION OF TWO DEVICES BEING STUCK TOGETHER WAS CONFIRMED, WHICH AGREES WITH THE REPORTED COMPLAINT CONDITION. THIS REPORT IS FOR ONE (1) SLIDING CARRIAGE. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7841673 ↗

Other APPARATUS, TRACTION, NON-POWERED devices

All APPARATUS, TRACTION, NON-POWERED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.