Skip to content

NA 395.911Recalls & reported problems

Synthes GmbH · DRILL SLEEVE HANDLE · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982200891), received between Sep 17, 2021 and Oct 10, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility1
  • Material Deformation1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON SEPTEMBER 21, 2023, THE PATIENT UNDERWENT A TEMPORARY EXTERNAL FIXATION FOR AN ANKLE FRACTURE. WHILE DRILLING THE TIBIAL USING THE SELDRILL WITH THE SLEEVE, THE SELDRILL STOPPED ADVANCING. SINCE THE INSERTION DEPTH WAS SHALLOW, THE SELDRILL AND SLEEVE WERE ABLE TO BE REMOVED, AND THE SURGEON CONTINUED THE OPERATION WITHOUT USING A SLEEVE. THE SELDRILL AND SLEEVE GOT STUCK AND COULD NOT BE DETACHED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THIS REPORT INVOLVES ONE DRILL SLEEVE HANDLE. THIS IS REPORT 3 OF 3 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 17903963 ↗
  • MalfunctionSep 17, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE DRILL SLEEVE HANDLE WAS OVERTORQUED AND WAS BENT WHEN INSERTING THE EX FIX SLEEVES. WHILE INSERTING THE TWO STRAIGHT OUTRIGGERS ON THE COMBI CLAMP, IT ALSO WAS OVERTIGHTENED AND WAS STRIPPED. THE MULTI PIN GUIDE FOR THE TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) WAS STICKING DURING THE CONTRALATERAL FEMUR WAS BEING IMPLANTED. THE HANDLE OF 395.911 WAS USE TO PULL TRACTION AND IT ENDED UP SCREWING UP THE THREADS OF THE HANDLE. A TRIPLE SLEEVE WAS THREADED IN AND WE WERE UNABLE TO THREAD IT IN. THE STRAIGHT OUTRIGGERS WERE BOTH UNABLE TO BE TIGHTENED. THERE WAS A (3) MINUTE SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING DIFFERENT CLAMPS. CONCOMITANT DEVICE REPORTED: UNKOWN - EXFIX: (PART# UNKNOWN; LOT# UNKNOWN; QUALITY:1). THIS REPORT IS FOR (1) DRILL SLEEVE HANDLE. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12488874 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page