The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982200945), received between Aug 13, 2018 and Mar 18, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation4
- Device-Device Incompatibility2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THEY WERE DOING EXTRA-LARGE EXTERNAL FIXATOR CASE WITH HA COAT UPON INSERTION HA COAT THROUGH THE SLEEVE THEY GOT CUT INSIDE. THE PATIENT WAS NOT AFFECTED, AND SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) 6.0MM/5.0MM THREADED DRILL SLEEVE-LONG THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18924612 ↗MalfunctionSep 28, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE SLEEVE IS BENT AND WEDGED SHANZ PIN. IT IS UNABLE TO SPECTATE. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) 6.0MM/5.0MM THREADED DRILL SLEEVE-LONG THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17836495 ↗MalfunctionSep 28, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE SLEEVE IS BENT AND WEDGED SHANZ PIN. IT IS UNABLE TO SPECTATE. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) 6.0MM/5.0MM THREADED DRILL SLEEVE-LONG. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17836224 ↗MalfunctionMay 25, 2023Material Deformation; Material Deformation
IT WAS REPORTED, THAT ON AN UNKNOWN DATE. THE ITEMS IN QUESTION WERE DISCOVERED, TO BE BENT AND DAMAGED. THIS WAS OBSERVED, POST-OP WHILE THE ITEMS WERE BEING CLEANED IN STERILE PROCESSING. THERE WAS NO REPORTED PATIENT INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 6.0MM/5.0MM THREADED DRILL SLEEVE-LONG. THIS IS REPORT 3 OF 6 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17007187 ↗MalfunctionMay 25, 2023Material Deformation; Material Deformation
IT WAS REPORTED, THAT ON AN UNKNOWN DATE. THE ITEMS IN QUESTION WERE DISCOVERED TO BE BENT AND DAMAGED. THIS WAS OBSERVED, POST-OP WHILE THE ITEMS WERE BEING CLEANED IN STERILE PROCESSING. THERE WAS NO REPORTED PATIENT INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 6.0MM/5.0MM THREADED DRILL SLEEVE-LONG. THIS IS REPORT 6 OF 6 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17007186 ↗MalfunctionAug 13, 2018Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT A SCREW WAS STUCK IN A LONG-THREADED DRILL SLEEVE. IT WAS FOUND OUT AT THE STERILE PROCESSING DEPARTMENT ON (B)(6) 2018. THERE WAS NO PATIENT AND PROCEDURE INVOLVED. THIS REPORT IS FOR ONE (1) 6.0MM/5.0MM THREADED DRILL SLEEVE-LONG. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 7778610 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page