NA 422.402Recalls & reported problems
Synthes GmbH · 5.0MM TI LOCKING SCREW SELF-TAPPING 14MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982215925), received between Dec 15, 2017 and Dec 15, 2017 — classified as 4 injuries.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Mechanical Jam2
Patient problems
- No Code Available4
- Bone Fracture(s)2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 15, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION OF A TITANIUM DISTAL FEMUR LESS INVASIVE STABILIZATION SYSTEM (LISS) PLATE AND THIRTEEN (13) SCREWS TO AN UNKNOWN ZIMMER PLATE CONSTRUCT DUE TO A FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE ON (B)(6) 2017. ONE (1) LISS PLATE ALONG WITH APPROXIMATELY THIRTEEN (13) UNKNOWN SCREWS WERE INITIALLY IMPLANTED ON AN UNKNOWN DATE A FEW YEARS AGO. BASED ON A PREVIOUSLY TAKEN X-RAY (UNAVAILABLE) TAKEN BEFORE THE REVISION SURGERY, THE PLATE AND SCREWS ALL SEEMED INTACT AND WERE FOUND INTACT DURING SURGERY. THE REVISION PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN DELAY IN SURGERY DUE TO THREE (3) SYNTHES SCREWDRIVERS BREAKING WHILE REMOVING THE THREE (3) COLD WELDED / CROSS THREADED SCREWS THAT WERE STUCK TO THE PLATE. AFTER THE SCREWDRIVERS BROKE, THE SCREWS HAD TO BE CUT OUT WITH A BIG BURR. IT IS UNKNOWN IF THERE WERE ANY FRAGMENTS LEFT INSIDE THE PATIENT. THIS REPORT ADDRESSES THE POSTOPERATIVE REVISION DUE TO PATIENT¿S FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE THE INTRAOPERATIVE ISSUE OF COLD WELDED / CROSS THREADED SCREWS THAT WERE STUCK TO THE PLATE. THE INTRAOPERATIVE ISSUE OF TH
Patient problems: Bone Fracture(s); No Code Available
MAUDE report 7122010 ↗InjuryDec 15, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION OF A TITANIUM DISTAL FEMUR LESS INVASIVE STABILIZATION SYSTEM (LISS) PLATE AND THIRTEEN (13) SCREWS TO AN UNKNOWN ZIMMER PLATE CONSTRUCT DUE TO A FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE ON (B)(6) 2017. ONE (1) LISS PLATE ALONG WITH APPROXIMATELY THIRTEEN (13) UNKNOWN SCREWS WERE INITIALLY IMPLANTED ON AN UNKNOWN DATE A FEW YEARS AGO. BASED ON A PREVIOUSLY TAKEN X-RAY (UNAVAILABLE) TAKEN BEFORE THE REVISION SURGERY, THE PLATE AND SCREWS ALL SEEMED INTACT AND WERE FOUND INTACT DURING SURGERY. THE REVISION PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN DELAY IN SURGERY DUE TO THREE (3) SYNTHES SCREWDRIVERS BREAKING WHILE REMOVING THE THREE (3) COLD WELDED / CROSS THREADED SCREWS THAT WERE STUCK TO THE PLATE. AFTER THE SCREWDRIVERS BROKE, THE SCREWS HAD TO BE CUT OUT WITH A BIG BURR. IT IS UNKNOWN IF THERE WERE ANY FRAGMENTS LEFT INSIDE THE PATIENT. THIS REPORT ADDRESSES THE POSTOPERATIVE REVISION DUE TO PATIENT¿S FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE THE INTRAOPERATIVE ISSUE OF COLD WELDED / CROSS THREADED SCREWS THAT WERE STUCK TO THE PLATE. THE INTRAOPERATIVE ISSUE OF TH
Patient problems: Bone Fracture(s); No Code Available
MAUDE report 7121863 ↗InjuryDec 15, 2017Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING A REVISION SURGERY ON (B)(6) 2017, TO REMOVE A SYNTHES TITANIUM DISTAL FEMUR LESS INVASIVE STABILIZATION SYSTEM (LISS) PLATE AND THIRTEEN (13) SCREWS, THREE (3) SYNTHES SCREWDRIVERS BROKE WHEN THE SURGEON ATTEMPTED TO REMOVE THREE (3) COLD WELDED / CROSS THREADED SCREWS FROM THE PLATE. ONE (1) LISS PLATE ALONG WITH APPROXIMATELY THIRTEEN (13) UNKNOWN SCREWS WERE INITIALLY IMPLANTED ON AN UNKNOWN DATE A FEW YEARS AGO. THE PATIENT HAD A FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE AND THE PLAN WAS TO REVISE THE PATIENT TO ANOTHER ZIMMER PLATE. HENCE THE REVISION SURGERY WAS CONDUCTED. BASED ON A PREVIOUSLY TAKEN X-RAY (UNAVAILABLE) TAKEN BEFORE THE REVISION SURGERY, THE PLATE AND SCREWS ALL SEEMED INTACT AND WERE FOUND INTACT DURING SURGERY. THE SCREWS THAT WERE COLD WELDED WERE THE ONLY ONES THAT BROKE / GOT DAMAGED TO REMOVE THEM DURING THE SURGERY. THE REVISION PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN DELAY IN SURGERY DUE TO THREE (3) SYNTHES SCREWDRIVERS BREAKING WHILE REMOVING THE THREE (3) COLD WELDED / CROSS THREADED SCREWS THAT WERE STUCK TO THE PLATE. AFTER THE SCREWDRIVERS BROKE, THE SCREWS HAD TO BE CUT OUT WITH
Patient problems: No Code Available
MAUDE report 7120701 ↗InjuryDec 15, 2017Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING A REVISION SURGERY ON (B)(6) 2017, TO REMOVE A SYNTHES TITANIUM DISTAL FEMUR LESS INVASIVE STABILIZATION SYSTEM (LISS) PLATE AND THIRTEEN (13) SCREWS, THREE (3) SYNTHES SCREWDRIVERS BROKE WHEN THE SURGEON ATTEMPTED TO REMOVE THREE (3) COLD WELDED / CROSS THREADED SCREWS FROM THE PLATE. ONE (1) LISS PLATE ALONG WITH APPROXIMATELY THIRTEEN (13) UNKNOWN SCREWS WERE INITIALLY IMPLANTED ON AN UNKNOWN DATE A FEW YEARS AGO. THE PATIENT HAD A FRACTURE WHICH WAS SLIGHTLY DISTAL TO THE SYNTHES LISS FEMUR PLATE AND THE PLAN WAS TO REVISE THE PATIENT TO ANOTHER ZIMMER PLATE. HENCE THE REVISION SURGERY WAS CONDUCTED. BASED ON A PREVIOUSLY TAKEN X-RAY (UNAVAILABLE) TAKEN BEFORE THE REVISION SURGERY, THE PLATE AND SCREWS ALL SEEMED INTACT AND WERE FOUND INTACT DURING SURGERY. THE SCREWS THAT WERE COLD WELDED WERE THE ONLY ONES THAT BROKE / GOT DAMAGED TO REMOVE THEM DURING THE SURGERY. THE REVISION PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN DELAY IN SURGERY DUE TO THREE (3) SYNTHES SCREWDRIVERS BREAKING WHILE REMOVING THE THREE (3) COLD WELDED / CROSS THREADED SCREWS THAT WERE STUCK TO THE PLATE. AFTER THE SCREWDRIVERS BROKE, THE SCREWS HAD TO BE CUT OUT WITH
Patient problems: No Code Available
MAUDE report 7120699 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page