TOMOFIX 440.874Recalls & reported problems
Synthes GmbH · TI TOMOFIX(TM) LATERAL DISTAL FEMUR PLATE-4 HOLES/LEFT · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982220776), received between Jul 9, 2018 and Jul 9, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K023941, K141796); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Mechanical Problem1
Patient problems
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 9, 2018Mechanical Problem; Mechanical Problem
(B)(6) REPORTS THAT DURING REMOVAL OF TOMOFIX LATERAL DISTAL FEMUR PLATE ON AN UNKNOWN DATE, THE SURGEON DESCRIBED THE SCREW AND PLATE JUNCTION STATUS AS ¿COLD WELDING¿. THE SURGEON HAD TROUBLE REMOVING THE TITANIUM TOMOFIX PLATE BECAUSE THE UNKNOWN SCREWS STRONGLY ADHERED TO THE PLATE. THE SURGEON ALWAYS FINISHED THE LAST SCREW INSERTION THREADS USING A TORQUE LIMITING SCREWDRIVER AS DESCRIBE AND PROPOSED IN THE TECHNIQUE GUIDE. DURING THE REMOVAL, THE SURGEON HAD TO USE A BROKEN SCREW REMOVAL SET AND DIAMOND BURR TO SHEAR OFF THE SCREW HEAD AT THE PLATE JUNCTION TO ENABLE REMOVAL OF THE PLATE AND SCREW CONSTRUCT. THE SURGEON ADDED THAT THE PLATE ALWAYS ACHIEVED CLINICAL RESULT GOAL BUT THE CONSTRUCT IS A NIGHTMARE TO REMOVE AFTER BONE HEALING. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY RELATED TO THE EVENT. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR A TI TOMOFIX(TM) LATERAL DISTAL FEMUR PLATE-4 HOLES/LEFT. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7669722 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K023941, K141796. One submission often covers many models, so check the product and lot details in each record.
- Z-0473-2014 · Oct 9, 2013 · TerminatedSynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial and
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page