NA 447.612Recalls & reported problems
Synthes GmbH · 2.0MM TI Y-PLATE/3 HOLES HEAD 8 HOLES SHAFT/48MM · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982224002), received between May 15, 2018 and May 15, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Failure of Implant1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 15, 2018Break; Break
IT WAS REPORTED THAT A CLOSED WEDGE OSTEOTOMY OF PROXIMAL FIRST METATARSAL OF THE RIGHT FOOT WAS PERFORMED ON (B)(6) 2018. DURING THE PROCEDURE, A FIXATION OF BONE FRAGMENTS WAS ACHIEVED WITH 2.0MM 3X8 HOLE Y PLATE-TITANIUM AND 6 CORTEX SCREWS FROM 2.0 MODULAR HANDSET. THE SURGEON REPORTED NO IN-OPERATIVE EVENTS OR MALFUNCTION OF THE IMPLANTS OR INSTRUMENTATION USED. PRIOR TO IMPLANTATION THE STRAIGHT PORTION OF THE PLATE HAD BEEN CUT, AND THE THREE HOLES ON THE END OF THE PLATE WERE REMOVED BEFORE BEING USED ON THE PATIENT. REPORTEDLY, THE PATIENT WAS SEEN IN A POST-OPERATIVE VISIT TWO DAYS LATER AND IMAGES OF THE IMPLANTS SHOWED THE PLATE WAS BROKEN. IT WAS FURTHER EXPLAINED THAT THE JUNCTION BETWEEN THE 3-HOLED PRONGED END OF THE PLATE AND THE STRAIGHT 5-HOLED PORTION WAS BROKEN. ALL SIX IMPLANTED SCREWS WERE INTACT AND IN THE SAME SITE VISUALIZED IN POST-OPERATIVE IMAGING. WHEN ASKED IF ANYTHING HAD OCCURRED BETWEEN THE INITIAL SURGERY AND THE POST-OPERATIVE VISIT, THE PATIENT REPORTED THAT SHE HAD TRIPPED AND FELL ONE HOUR AFTER THE INITIAL SURGERY. THE PATIENT WILL BE REVISED ON (B)(6) 2018, WHEN THE SAME IMPLANT TYPE/SIZE AND NEW SCREWS WILL BE PLACED AND THE TWO BROKEN PLAT
Patient problems: Failure of Implant
MAUDE report 7515162 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page