Skip to content

NA 456.416SRecalls & reported problems

Synthes GmbH · 11MM/130 DEG TI CANN TROCH FIXATION NAIL 360MM/RIGHT-STER · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982227829), received between May 23, 2019 and May 23, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

7 other recalls cover devices cleared under the same FDA submission (K011857); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Bone Fracture(s)1
  • Impaired Healing1
  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 23, 2019Break; Break

    IT WAS REPORTED ON (B)(6) 2019, PATIENT UNDERWENT REMOVAL OF TITANIUM CANNULATED TROCHANTERIC FIXATION NAIL DUE TO BROKEN NAIL AT THE LEVEL OF HELICAL BLADE AND FRACTURE DOES NOT HEAL. THE NAIL WAS REMOVED AND HIP REPLACED. THERE WAS NO SURGICAL DELAY. PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICES REPORTED: UNKNOWN HELICAL BLADE (PART #: UNKNOWN, LOT #: UNKNOWN, QUANTITY #: 1), UNKNOWN SCREW (PART #: UNKNOWN, LOT #: UNKNOWN, QUANTITY #: UNKNOWN). THIS REPORT IS FOR ONE (1) 11MM/130 DEG TI CANN TROCH FIXATION NAIL 360MM/RIGHT-STER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Bone Fracture(s); Impaired Healing; No Code Available

    MAUDE report 8637503 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K011857. One submission often covers many models, so check the product and lot details in each record.

  • Z-0604-2018 · Jul 21, 2017 · Terminated
    11MM/130 Degree Titanium Trochanteric Fixation Nail 380MM/Right, Sterile
  • Z-0375-2016 · Nov 2, 2015 · Terminated
    Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System) The TFN-ADVANCED Proximal Femoral Naili
  • Z-0219-2015 · Oct 29, 2014 · Terminated
    10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM - Sterile: intended to treat stable and unstable fractures of the proximal femur
  • Z-0387-2014 · Jul 31, 2013 · Terminated
    Large Buttress Compression Nut For Blade Guide Sleeve TFN The Synthes Large Buttress Compression Nut For Blade Guide Sleeve TFN is intended to treat stable and unstable pertrocha
  • Z-0452-2014 · Jul 30, 2013 · Terminated
    Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the proximal femur.
  • Z-0474-2014 · Jul 30, 2013 · Terminated
    Synthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFN Nail is intended to treat stable and unstable fractures of the proximal
  • Z-2058-2013 · Nov 2, 2012 · Terminated
    Synthes Trauma Nail System. The devices are indication for bone fixation.

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates