MAXFRAME 03.312.001Recalls & reported problems
Synthes GmbH · WIRE TENSIONER · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982261106), received between May 15, 2020 and Jan 3, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Device Slipped” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Slipped4
- Mechanical Jam3
- Mechanical Problem2
- Crack1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 3, 2024Device Slipped; Device Slipped
IT WAS REPORTED THAT ON AN (B)(6) 2023 THE MAXFRAME TENSIONER SLIPPED DURING TENSIONING. THE TENSIONER WAS REMOVED FROM THE PROCEDURE AND SET ASIDE FOR REVIEW. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE. THERE WAS NO GENERATION OF FRAGMENTS. THIS REPORT INVOLVES ONE (1) WIRE TENSIONER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18433995 ↗MalfunctionNov 2, 2022Device Slipped; Mechanical Jam; Device Slipped; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2022 THEY WERE DOING A MAXFRAME WITH SURGEON. DURING THE CASE HAVING ISSUES WITH THE TENSIONERS IN THE SET. THEY KEPT PISTONING ON THE WIRE AND LOSING THEIR TENSION. THEY REACHED OUT AND THE ENGINEERS FOR MAXFRAME CALLED IN DURING THE CASE TO DISCUSS THE ISSUES. SURGEON HAD BEEN DOING FRAMES FOR YEARS. THEY WERE ABLE TO FINISH THE CASE AND GET OUT OF THE OR. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH SURGICAL DELAY. THIS REPORT IS FOR A WIRE TENSIONER. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15716967 ↗MalfunctionNov 2, 2022Device Slipped; Mechanical Jam; Device Slipped; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6)2022 THEY WERE DOING A MAXFRAME WITH SURGEON. DURING THE CASE HAVING ISSUES WITH THE TENSIONERS IN THE SET. THEY KEPT PISTONING ON THE WIRE AND LOSING THEIR TENSION. THEY REACHED OUT AND THE ENGINEERS FOR MAXFRAME CALLED IN DURING THE CASE TO DISCUSS THE ISSUES. SURGEON HAD BEEN DOING FRAMES FOR YEARS. THEY WERE ABLE TO FINISH THE CASE AND GET OUT OF THE OR. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH SURGICAL DELAY. THIS REPORT IS FOR A WIRE TENSIONER. THIS IS REPORT 3 OF 3 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15716954 ↗MalfunctionNov 2, 2022Device Slipped; Mechanical Jam; Device Slipped; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6)2022 THEY WERE DOING A MAXFRAME WITH SURGEON. DURING THE CASE HAVING ISSUES WITH THE TENSIONERS IN THE SET. THEY KEPT PISTONING ON THE WIRE AND LOSING THEIR TENSION. THEY REACHED OUT AND THE ENGINEERS FOR MAXFRAME CALLED IN DURING THE CASE TO DISCUSS THE ISSUES. SURGEON HAD BEEN DOING FRAMES FOR YEARS. THEY WERE ABLE TO FINISH THE CASE AND GET OUT OF THE OR. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH SURGICAL DELAY. THIS REPORT IS FOR A WIRE TENSIONER. THIS IS REPORT 2 OF 3 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15716950 ↗MalfunctionJun 28, 2022Crack; Scratched Material; Crack; Scratched Material
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT THE WIRE TENSIONER WAS UNABLE TO STERILIZE/ASSEMBLE. CHIPPING AND CRACKING. UNABLE TO USE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. THIS COMPLAINT INVOLVES ONE (1) DEVICE WIRE TENSIONER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14845167 ↗MalfunctionMay 15, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE. DURING THE PROCEDURE, 2 WIRE TENSIONERS WERE NOT TENSIONING, THEY WERE NOT GRIPPING THE WIRE. THE FRAME CONSTRUCT HAD 8 WIRES TOTAL, AND 6 OF THEM WERE SUCCESSFULLY TENSIONED BUT THE REMAINING TWO WERE NOT. THE DOCTOR HAD GOTTEN THE WIRE TENSIONERS TO WORK AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PATIENT'S STATUS IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN WIRES (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# 2). THIS REPORT IS FOR 1 WIRE TENSIONER. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10063589 ↗MalfunctionMay 15, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE. DURING THE PROCEDURE, 2 WIRE TENSIONERS WERE NOT TENSIONING, THEY WERE NOT GRIPPING THE WIRE. THE FRAME CONSTRUCT HAD 8 WIRES TOTAL, AND 6 OF THEM WERE SUCCESSFULLY TENSIONED BUT THE REMAINING TWO WERE NOT. THE DOCTOR HAD GOTTEN THE WIRE TENSIONERS TO WORK AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PATIENT'S STATUS IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN WIRE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# 2). THIS REPORT IS FOR 1 WIRE TENSIONER. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10063578 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page